Page 1
Dignity Hydraulic Height Adjustable Reclining Shower Chair Operator’s Instructions EZ Way, Inc. PO Box 89 Clarinda, IA 51632 1-800-627-8940 www.ezlifts.com Form 2-293 Rev. 07/30/18 WARNING: Cancer and Reproductive Harm - www.P65Warnings.ca.gov.
Product Description Table of Contents General advisories Introduction Symbols used Intended Proper Use Use and storage Equipment Warranty Nameplate, ordering spare parts Re-use, product lifetime Disposal Safety advisories for use Product specifications Delivery scope Technical specifications Spare parts, accessories Setting up the chair for use Position of the footrests Adjusting the foot plate height Installing, adjusting the headrest...
1. General advisories 1.4 Use and storage • The product’s temperature range for use must be between +41°F and +104°F. Store the product 1.1 Introduction in a dry room with a relative humidity of In this booklet you will find information and 30-50%.
3. Nameplate, ordering spare parts 5. Disposal The nameplate contains the following information: Disposal and recycling of used products as well as the packaging material must be done according to the applicable rules and legal regulations. Please contact your local disposal company or environmental authorities for further information.
When the chair is to be used by patients Danger of pinching! • To prevent body parts, etc. from being pinched, with insensitive skin (persons who cannot never reach into the hinge area when folding the feel temperature rises and / or have skin armrests up or down.
• Schematic: 1 safety belt (Art. No.: 322.30.10) • 1 User Instructions booklet 7.2. Technical specifications For the purpose of continuous development, we reserve the right to make modifications with respect to the chair depicted in the User Instructions booklet. Dimension (ca.) PU-H* 1-total height...
9. Setting up the chair for use 9.3 Installing, adjusting the headrest The product is delivered completely assembled in a Installation: carton. Before using the first time, the chair first must be checked over for transport damage. Settings and Flip the levers (A) upwards adjustments may be done only by a qualified specialist.
9.5 Setting the headrest angle By rotating knob (A), the headrest angle can be adjusted. Warning! Danger of pinching (B) between the headrest and backrest! 9.6 Adjusting the seat height The chair’s seat height is done in steps via the hydraulic unit, by activating the foot pedal.
10. Using the chair 10.2 Standing up Before standing up, it must be certain that the chair Clean the chair before first-time use! is stable and secured against slipping away or tipping over. The four wheels must be locked, both armrests must be latched in position, the It must be ensured that the chair is in a proper and footrests swiveled to the side or folded up, and the safe condition before use.
10.4 Sitting and laying while lifting 10.6 Removing the seat upholstery The seat with (I) and without (O) hygiene opening, the leg pad (K) and the backrest upholstery (H) can be removed from their mountings in the chair frame for cleaning purposes or for replacing (see description on page 2).
10.9 Removing the footrests Lift the footrests approx. 3.94" and rotate them outward 90°. Now you are able to push the footrests all the way up and to pull them out. 10.10 Using the commode pan 11. Troubleshooting When using the chair (PU) with a commode pan, it Chair malfunctions/faults must be used only in the sitting position (1) and not Fault...
12. Care Malfunctions/faults – hydraulic unit Fault Possible Remedy 12.1 Caring for the chair cause Piston Overflow Max. stroke reached. • moves in exceeded, From pressure buildup, In general the wheelchair can be cleaned using and out maximum overflow is a mild conventional cleaning and disinfecting only a stroke...
First push it to the side. on your own. Contact EZ Way at (800) 627-8940. Proceed with venting as follows: 1. Move the piston rod completely out by pressing 13.2 Maintenance on the hydraulic unit...
– no water spraying! 13.4 Upkeep • If you have questions concerning maintenance and repairs to the chair, then contact EZ Way at (800) 627-8940. • In the event of damage and/or needed repairs, contact EZ Way without delay.
Page 16
EZ Way Dignity Hydraulic Height Adjustable Reclining Shower Chair Warranty: EZ Way warrants that its products are free from defects in materials and workmanship for a period of 5 years from the date of delivery on the frame & 2 years on components. This warranty is in lieu of any other warranty or representation, expressed or implied.
Need help?
Do you have a question about the EZ352 and is the answer not in the manual?