NovAseptic GMP ATEX Mixer ® Installation Note The ATEX mixer is designed for operation in the classification noted on the product label. The mixer must be in good working order before operating. Before using the mixer, ensure the following: • The tank plate is installed properly on the tank and concentricity and alignment have been verified to prevent any contact between the mixing head, outer rotor and tank plate.
Installation and User Guide Operator and Equipment Safety All installers and operators of the equipment must read and understand this installation and user manual before using this equipment. Failure to follow operating and installation instructions could result in operator injury or damage to the equipment.
Installation and User Guide NovAseptic GMP ATEX Tank Plate (AX#/33) ® Mixer Components Each tank plate is marked with the serial number and a mill stamp that corresponds to its heat number. The tank plate is welded onto Catalog numbers and specifications for the the vessel and is considered an integrated part components listed here can be found on the of the vessel.
NovAseptic GMP ATEX Mixer ® Installation and Required Tools The following tools are required for installation Operating Parameters of the NovAseptic GMP ATEX Mixer. The ® tools (shown below) are not included with Please refer to the appropriate product the mixer and are available from your local specification sheet for more information.
Installation and User Guide Installation Use the Installation Checklist and the Installation Qualification Checklist (in this manual) as guides for installing the mixer and initiating use. Use the Maintenance Checklist as a guide for follow up care of the mixer. For welding of the Tank Plate use GMP ATEX Welding Instructions.
NovAseptic GMP ATEX Mixer ® Catalog Number Torque AX05/24# AX1/24# 6 Nm (4.4 Lbft) Align the mixing head with the tank plate. AX510/24# Ensure that the mixing head rotates smoothly. AX20/24# 13 Nm (9.6 Lbft) AX50200/24# 30 Nm (22.1 Lbft) Mixing Head Installation See NovAseptic GMP Mixing Head Installation...
Page 12
Installation and User Guide • Store the motor in an area which is humid and weather proof. Ensure that motor is on a stable surface to prevent falling or vibration. • Remove the transport collar before installing the motor. Mount the drive unit onto the vessel following the instructions for the appropriate mixer size: GMP 50 ATEX, GMP 100 ATEX and GMP 500 ATEX...
NovAseptic GMP ATEX Mixer ® Install the drive unit onto the tank plate. Insert the locking screws into the two oval holes in the tank plate (2). Electrical Installation The mixer must be operated with a variable frequency drive to function properly.
Page 14
Installation and User Guide Revolution Counter Observe laws, regulations and/or standards governing the use, or intended use, of this equipment. Any alteration of this equipment from factory specification may cause unsafe conditions and will void the product warranty. The tachometer must be connected to an intrinsically safe Ensure that the supply voltage and barrier.
NovAseptic GMP ATEX Mixer ® Operation NOTE Consult the parameter settings and circuit and terminal diagrams before Ensure that the installation has installing the control cabinet. been performed correctly and that the parameters are correctly Validate the current media and process conditions.
Installation and User Guide Operating Notes • For storage periods exceeding 60 days, protect all coupling surfaces (such as • Fill the vessel with liquid before adding solid flanges and shafts on the drive unit) with a material. suitable anti-oxidation product (Mobilarma 248 or equivalent).
NovAseptic GMP ATEX Mixer ® Cleaning and Sterilizing and Drive Unit Installation in this manual.) Procedures Magnetic particles are not removed Cleaning in Place (CIP) from the mixing head during CIP; they must be removed manually or by GMP mixers are designed for CIP using any alternative cleaning.
Page 18
Installation and User Guide from the mixing head during CIP; they must be removed manually or by alternative cleaning. Magnetic particles can cause corrosion and damage. They can also cause a higher torque, which may cause the mixing head to decouple and severely damage the vessel.
NovAseptic GMP ATEX Mixer ® Installation ATEX Checklist To ensure correct installation, before using the mixer make sure to complete each step in this checklist, and check the applicable box. Note: Follow all installation recommendations, cautions and warnings in this manual. Install the Male Bearing Complete Ensure the proper position of the O-ring on the male bearing.
Installation and User Guide Install the Drive Unit – GMP 1000 ATEX and GMP 2000 ATEX Complete Fit screws and tighten clockwise (view from below). q yes q no Install the drive unit into the tank plate. Install locking screws into the two oval q yes q no holes in the tank plate.
NovAseptic GMP ATEX Mixer ® Install the VFD System Complete Select the appropriate control cabinet. q yes q no Set voltage according to the AC motor. q yes q no Set frequency according to the AC motor. q yes q no Set minimum/maximum rpm.
Installation and User Guide Commissioning Valid Regulations ATEX 2014/34/EU Complete this checklist and ensure ATEX regulations are met. Commissioning Checkpoint Date Signature Tank plates are cylindrical inside the vessel. Use NovAseptic Gauge 1. ® Mixing Head freewheel. Use NovAseptic Gauge 2. ®...
NovAseptic GMP ATEX Mixer ® Installation Qualification Checklist Use this installation qualification for each vessel and mixer to ensure that the correct mixer has been installed and all important information has been gathered for future use. Installation Qualification Different applications require different operating parameters on the frequency converter depending on: •...
Page 24
Installation and User Guide Additional comments Inspected by: Print name _____________________________________________________________________ Signature _______________________________________________________________________ Date inspection completed...
NovAseptic GMP ATEX Mixer ® Maintenance Checklist To ensure safety and proper performance, perform quarterly inspection of the GMP ATEX Mixer according to the following table. Make copies of this table to use at each inspection. Refer to applicable instructions within this guide.
Page 26
Installation and User Guide Component / Maintenance Check Notes Complete Mixing Head Inspect the silicone carbide on the female q yes q no bearing for damage. Inspect the inside of the mixing head and bottom surface for damage. Check that mixing q yes q no head rotates easily.
NovAseptic GMP ATEX Mixer ® Spare Parts The number of spare parts to keep on hand depends on the number of mixers in production and their application. For further information, please contact Technical Service. The following equipment should be kept in stock: •...
Installation and User Guide Troubleshooting During troubleshooting, follow the instructions in this manual. A correct installation is crucial to proper operation. Before repair or service of the GMP ATEX Mixer, personnel must study and understand this user guide. Before performing maintenance on the GMP ATEX Mixer, make sure that the vessel is: •...
Page 29
NovAseptic GMP ATEX Mixer ® Problem Possible Causes Solutions Outer Bent shaft Contact Technical Service. drive head scratches the Bent hole-shaft in gearbox Contact Technical Service. tank plate Drive unit not properly in place Install drive unit correctly. Particles stuck to the magnets Remove particles.
Installation and User Guide Standard Warranty The applicable warranty for the products listed in this publication may be found at www.millipore. com/terms (within the “Terms and Conditions of Sale” applicable to your purchase transaction).