Langer FLS 106 IC User Manual

Ic scanner 4-axis positioning system / pcb scanner 3-axis positioning system
Hide thumbs Also See for FLS 106 IC:
Table of Contents

Advertisement

Quick Links

User manual
FLS 106 IC set / FLS 106 PCB set
IC scanner 4-axis positioning system /
PCB scanner 3-axis positioning system
Copyright © Februar 2019
LANGER EMV-Technik GmbH
- Translation of the original German user manual into English -
2019.08.15 User Manual FLS 106 eng

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FLS 106 IC and is the answer not in the manual?

Questions and answers

Summary of Contents for Langer FLS 106 IC

  • Page 1 User manual FLS 106 IC set / FLS 106 PCB set IC scanner 4-axis positioning system / PCB scanner 3-axis positioning system Copyright © Februar 2019 LANGER EMV-Technik GmbH - Translation of the original German user manual into English -...
  • Page 2: Table Of Contents

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Table of contents: Page Declaration of Conformity.................... 7 General Information ..................... 8 Storage of the User Manual ....................8 Reading and Understanding the User Manual ..............8 Local Safety and Accident Prevention Regulations ..............
  • Page 3 LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 6.1.1 Operation with the Protective Enclosure ............... 20 6.1.2 Operation with a Spatial Separation ................20 Switching the Scanner on in Normal Operation and after a Power Failure ......20 EMERGENCY STOP Switch .....................
  • Page 4 Operation with Spatial Separation and External NA 5 EMERGENCY STOP Switch ..43 14.2 Checking the Cables of the Rotary Unit ................44 14.3 Installing the Digital DM-CAM Microscope Camera on the FLS 106 IC ......44 14.3.1 Installing the DM-CAM Holder.3 for the Microscope Camera........44 14.3.2 Installing the DM-CAM Digital Microscope Camera ............
  • Page 5 Installing a Near-Field Probe with the SH 01 Probe Holder ........67 18.1 Installing a Near-Field Probe on the FLS 106 IC ..............67 18.1.1 Connecting a Near-Field Probe Directly to the FLS 106 IC Scanner ......68 18.1.2 Connecting a Near-Field Probe with a Preamplifier ............69 18.2 Installing a Near-Field Probe on the FLS 106 PCB ............
  • Page 6 LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 21.9 Removing the GND 25....................... 78 21.10 Removing the GND 25 Holder ..................78 21.11 Removing the DM-CAM Microscope Camera ..............78 21.12 Removing the DM-CAM Holder.3 Camera Holder ............78 Maintenance ........................
  • Page 7: Declaration Of Conformity

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Declaration of Conformity Manufacturer: Langer EMV-Technik GmbH Nöthnitzer Hang 31 01728 Bannewitz Germany Langer EMV-Technik GmbH herewith declares that the FLS 106 IC / FLS 106 PCB positioning system...
  • Page 8: General Information

    Figures have been included in this user manual to assist the reader's understanding but may differ from the device's actual version. Limitation of Liability In the following cases, Langer EMV-Technik GmbH can assume no liability for damage to property and personal injury if: The information given in this user manual has not been observed.
  • Page 9: Description Of Symbols

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Description of Symbols Warning signs General warning sign Warning, Electricity Warning, Crushing of hands - 9 / 83 -...
  • Page 10: Scope Of Delivery

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Scope of Delivery FLS 106 IC set Scope of Delivery* Item Name of article Short name Qty. 4-axis positioning system FLS 106 IC 2 ChipScan-Scanner software...
  • Page 11: Technical Parameters

    20 µm 20 µm 20 µm 1° distance Positioning speed 20 mm/s 25 mm/s 10 mm/s 90°/s Table 2: Technical parameters of the x-, y- and z-axis and the rotation axis (only FLS 106 IC) - 11 / 83 -...
  • Page 12: Technical Parameters Of The Subassemblies

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Technical Parameters of the Subassemblies 4.2.1 UH DUT Figure 2: Dimensions of the UH DUT universal holder 4.2.2 Claw 01 and Claw 02 Figure 3: Dimensions of the claws for the UH DUT (claw 01 and claw 02)
  • Page 13: Gnd 25 Holder

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Figure 4: Claws from above with the long hole and M3 screw 4.2.3 GND 25 Holder Figure 5: GND 25 holder with anti-rotation element - 13 / 83 -...
  • Page 14: Gnd 25 Ground Plane

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 4.2.4 GND 25 Ground Plane Figure 6: GND 25 ground plane Diameter 218 mm Depth of the recess 1.7 mm Recess dimensions (L x W)
  • Page 15: Suh 106 Safety Enclosure

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 4.2.6 SUH 106 Safety Enclosure Length 1260 mm Width 890 mm Height 1000 mm Connection 25-pin D-Sub Weight 50 kg Table 5: Technical parameters of the SUH 106 safety enclosure 4.2.7...
  • Page 16: Requirements For The Chipscan-Scanner Software

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Requirements for the ChipScan-Scanner Software Operating system Windows 7 64 bit (latest service packs) Monitor resolution (1280 x 1024) px Hard-disk storage 1 GB minimum / 10 GB recommended...
  • Page 17: Intended Use

    The probes can be moved across the DUT surface on the X-, Y- and Z-axis. In addition, the FLS 106 IC allows the probe's rotation around the Z-axis. The probe tip position can be checked visually with the DM-CAM digital microscope camera at any time.
  • Page 18: Risks If Not Used For Its Intended Purpose

    Carry out a visual check before performing a measurement with a product from Langer EMV-Technik GmbH. Replace any damaged connecting cables, attachments and probes. Please contact Langer EMV-Technik GmbH if damaged parts need to be replaced or repaired. Wear tight-fitting clothes when using the scanner.
  • Page 19: Hazard Due To Electrical Voltage

    Never use any damaged or defective devices. Do not connect or disconnect any cables whilst the FLS 106 scanner is in operation. Only staff from Langer EMV-Technik GmbH may open the scanner and work on the electrical components and electrical lines.
  • Page 20: Safety Devices

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Safety Devices Prerequisites for Safe Operation The FLS 106 scanner may only be operated if either a protective enclosure or a spatial separation in connection with the external NA 5 EMERGENCY STOP switch is used.
  • Page 21: Environmental Protection

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Environmental Protection Dispose of any packaging material paying due regard to the local environmental protection regulations. Do not allow lubricants such as grease and oil based on mineral oils to escape into the environment.
  • Page 22: Overview Of The Fls 106 Scanner

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Overview of the FLS 106 Scanner FLS 106 with Axis Designation Figure 9: Overview FLS 106 Description of the Subassemblies 9.2.1 Guide Rails along the X-Axis...
  • Page 23: Guide Rails Along The Y-Axis (Long Side)

    Figure 12: Z-axis tower with rotary unit A rotary unit is located at the bottom of the Z-axis tower on the FLS 106 IC. The SH 01 probe holder and the KA 220 camera arm are located at the bottom of the Z-axis tower on the FLS 106 PCB.
  • Page 24: Gantry

    The gantry transports the probe that is attached to the rotary unit and the digital DM-CAM microscope camera along the Y-axis. 9.2.5 Rotary Unit Figure 14: Rotary unit with SMA-SMA coupling and fastening screws The FLS 106 IC scanner comes with a rotary unit as a standard feature. - 24 / 83 -...
  • Page 25: Aluminium Mounting Bracket

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de The rotary unit is used as a fourth axis that allows the installed probe to rotate around the Z-axis by ±180°. The DM-CAM holder.3 for the digital microscope camera is attached to the rotary unit. Either an ICR probe or the SH 01 probe holder can be mounted on the rotary unit's rotary ring.
  • Page 26: T-Slot Table

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 9.2.7 T-Slot Table Figure 16: T-slot table T-slots with openings for the T-nuts The T-slot table (600 mm x 540 mm x 19 mm) has several hollow sections. T-nuts (Figure 17) can be inserted into the individual hollow sections (T-slots) to fasten the clamping pieces (Section 10.4)
  • Page 27: Control Panel With Emergency Stop Switch And Control Leds

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 9.2.8 Control Panel with EMERGENCY STOP Switch and Control LEDs Figure 18: Control panel of the FLS 106 scanner The EMERGENCY STOP switch that is integrated in the control panel is used to stop the FLS 106 scanner immediately.
  • Page 28: On / Off Switch And Connectors

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de ON / OFF Switch and Connectors 9.3.1 ON / OFF Switch Figure 20: ON / OFF switch An ON / OFF toggle switch is located on the rear of the FLS 106 scanner.
  • Page 29: Connectors On The Z-Axis Tower

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 9.3.4 Connectors on the Z-Axis Tower Figure 23: Connectors on the Z-axis tower - HR10 connectors:  12-pin connector marked Camera to connect the DM-CAM digital microscope camera ...
  • Page 30: Overview Of Attachments

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Overview of Attachments 10.1 KA 220 Camera Arm The KA 220 camera arm of the FLS 106 PCB scanner holds the DM-CAM digital microscope camera. It is attached to the aluminum mounting bracket.
  • Page 31: Dm-Cam With Camera Screw

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 10.3 DM-CAM with Camera Screw The digital DM-CAM microscope camera is used to observe the probe tip position and the distance between the probe tip and DUT.
  • Page 32: Clamping Pieces

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 10.4 Clamping Pieces The clamping pieces are used as fasteners to individually position and mount the UH DUT universal holder or a DUT on the FLS 106 scanner's T-slot table.
  • Page 33: Gnd 25 Holder

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 10.6 GND 25 Holder The GND 25 holder holds the GND 25 ground plane. The GND 25 holder can be mounted directly on the T-slot table or on the UH DUT universal holder.
  • Page 34: Sh 01 Probe Holder

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 10.8 SH 01 Probe Holder The SH 01 probe holder is designed to hold a near-field probe. The SH 01 comprises a mounting bracket, the magnetic holder and the probe sleeve.
  • Page 35: Delivery

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Delivery 11.1 Transport The FLS 106 scanner is delivered in a customized transport box. The scanner itself weighs 75 kg. The transport box's dimensions and weight may vary depending on the equipment and place of destination.
  • Page 36: Opening The Transport Box

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 11.4 Opening the Transport Box Ensure that the transport box is placed on level ground. Use appropriate tools to open the transport box and prevent damage to the equipment. Do not break open the box by brute force.
  • Page 37: Set-Up And Preparations For Putting The Fls 106 Scanner Into Service

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Set-Up and Preparations for Putting the FLS 106 Scanner into Service 12.1 Preparations for the Safe Operation of the FLS 106 Scanner There are two options to ensure the system's safe operation: use the SUH 106 safety enclosure or provide for a spatial separation (see Section 6.1).
  • Page 38: Setting Up The Fls 106

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de The FLS 106 scanner may only be set up and used in closed rooms under appropriate light conditions. Make sure that the free working area around the scanner is large enough for the scanner to be handled properly.
  • Page 39: Removing The Protective Film

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 12.5 Removing the Protective Film After setting up the scanner at its place of operation, remove the protective film. Danger due to use of sharp-edged, metallic tools!
  • Page 40: Removing The Transport Protection From The Guide Rails

    12.7 Removing the Transport Protection from the Guide Rails Langer EMV-Technik GmbH applies a thin layer of grease to all sliding rails so as to ensure appropriate corrosion protection during transport and storage. Remove this grease before putting the system into service.
  • Page 41: Checking / Releasing The Emergency Stop Switch

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 12.8 Checking / Releasing the EMERGENCY STOP Switch The FLS 106 is transported with the EMERGENY STOP switch engaged. If the EMERGENCY STOP switch is engaged, the scanner cannot be put into operation. Therefore, it should be checked whether the EMERGENCY STOP switch is engaged before switching on.
  • Page 42: Putting The Fls 106 Scanner Into Service For The First Time

    Standard procedure: a) Install and connect the digital microscope camera: - refer to section 14.3.2 for FLS 106 IC or - refer to section 14.4 for FLS 106 PCB b) Check that the cable of the rotary unit is connected correctly (connect if necessary) (Figure 40).
  • Page 43: Installation

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Installation 14.1 Ensuring the System's Safe Operation 14.1.1 Operation with the SUH 106 Safety Enclosure The safety enclosure protects the user and prevents any work on the scanner whilst it is in operation.
  • Page 44: Checking The Cables Of The Rotary Unit

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 14.2 Checking the Cables of the Rotary Unit Check that the rotary unit is connected to rot. Axis on the Z-axis tower. Figure 40: Cable of the rotary unit on the "rot. Axis“ connector 14.3...
  • Page 45: Installing The Dm-Cam Digital Microscope Camera

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 14.3.2 Installing the DM-CAM Digital Microscope Camera Use the large, silver-colored knurled screw and the black, knurled plastic nut to mount the camera screw on the DM-CAM holder.3. Insert the digital microscope camera into the opening of the camera screw and fix it in the desired place with the small, silver-colored knurled screw.
  • Page 46: Installing The Dm-Cam Digital Microscope Camera On The Fls 106 Pcb

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 14.4 Installing the DM-CAM Digital Microscope Camera on the FLS 106 PCB Figure 45: Digital DM-CAM microscope camera Figure 46: KA 220 camera arm Mount the DM-CAM on the KA 220 camera arm (Figure 47) with the FLS 106 PCB. Insert the DM-CAM into the opening of the camera screw located on the camera arm and fix with the small, silver-colored knurled screw.
  • Page 47 LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Power supply plug plugged into power supply socket - 47 / 83 -...
  • Page 48: Connecting A Computer

    Figure 51: Connecting the measuring cable with SMA plug to the scanner's "RF out" port Please refer to the annex of the ChipScan-Scanner software user manual or the Langer EMV-Technik website for a list of measuring devices supported by the ChipScan-Scanner software www.langer-emv.com/fileadmin/ChipScan-ESA%20Supported%20Spectrum%20Analyzers.pdf...
  • Page 49: Switching The Fls 106 Scanner On

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 14.8 Switching the FLS 106 Scanner on To switch on the scanner, the device lead must be properly connected and the toggle switch on the rear of the scanner pushed into the “I”...
  • Page 50: Software Installation

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Software Installation 15.1 Note on Software Installation Put the device into service as described in chapter 13 before installing the software. 15.2 Installing the scanner driver As the driver is not installed automatically by Windows when the IC scanner is connected to your computer, the driver installation needs to be done manually.
  • Page 51 LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de In the appearing properties dialog select Driver tab and click on Update Driver. Figure 54: Updating the driver 4) In the following dialog choose the option Browse my computer for driver software.
  • Page 52: Installing The Chipscan-Scanner Software

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 5) Click the Browse button that appears, select the USB memory stick, select the Driver sub- folder and click the OK button. Figure 56: Browsing the enclosed USB memory stick for the driver 6) The driver will now be installed.
  • Page 53: Putting The Fls 106 Scanner Into Service With The Chipscan-Scanner Software

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 15.4 Putting the FLS 106 Scanner into Service with the ChipScan-Scanner Software Ensure that the system has been set up as described in chapter 13 (item a – k).
  • Page 54 Select the COM interface to which the scanner is connected in the RS232 selection box in the Scanner section. o Select the Langer FLS 106 entry in the selection box below RS232. o Then click the Detect Devices button.
  • Page 55 LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 4. Calibrate the FLS 106 scanner. - Click Calibrate (lower right corner) to move the scanner to its home position. Note: This has to be done each time the ChipScan-Scanner software is started, each time the scanner is switched on and each time the scanner is stopped with the EMERGENCY STOP switch.
  • Page 56: Fastening The Device Under Test

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Fastening the Device Under Test 16.1 Fastening a Device Under Test With the Clamping Pieces Fasten the clamping pieces with the hexagon socket screws M6 (hexagon socket 5 mm) and the T-nuts in the selected T-slots.
  • Page 57: Fastening A Circuit Board On The Uh Dut Universal Holder

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Adjust the height and angle with the grub screws so that the device under test can be pushed under the projecting part of the clamping piece. Align the device under test and fix the clamping piece with the hexagon socket screw M6.
  • Page 58: Fastening The Circuit Board

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Place the clamping pieces above the T-nuts and secure loosely with the hexagon socket screw M6. Figure 64: Fastening the UH DUT with the clamping pieces on the T-slot table Adjust the height and angle with the grub screws as required.
  • Page 59: Installing The Gnd 25 Ground Plane For Ic Measurements

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 16.3 Installing the GND 25 Ground Plane for IC Measurements 16.3.1 Installing the GND 25 Holder First insert and position two T-nuts into the desired T-slot (Figure 66).
  • Page 60: Installing The Gnd 25 Ground Plane

    The GND 25 rests loosely on the GND 25 holder. This means that the installed GND 25 may fall off or down when the IC scanner is transported, for example. Always remove the GND 25 before transporting the FLS 106 IC scanner. - 60 / 83 -...
  • Page 61: Installing An Icr Near-Field Microprobe

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Installing an ICR Near-Field Microprobe 17.1 Safety Instructions for Installing an ICR Near-Field Microprobe The probe tip of the ICR probe is highly sensitive to mechanical...
  • Page 62 LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Screw the short SMA-SMA RA measuring cable to the SMA coupling (Figure 71) and then lay it on the rotary ring as shown in Figure 71.
  • Page 63: Connecting The Icr Near-Field Microprobe

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Figure 75: Connecting the SMA-SMA RA measuring Figure 76: Cable holder on the rotary ring cable to the SMA coupling Place the SMA-SMA cable and SSMB-SSMB cable in the open cable holder, attach the upper half of the holder and fasten with screws (Figure 76).
  • Page 64 LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Next connect the SMA-SMA RA measuring cable to the SMA connector marked RF out (Figure 78) and then the SSMB-SSMB cable to the SSMB connector on the ICR probe (Figure 79).
  • Page 65: Installing The Bt 706 Bias-Tee

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Figure 81: Near-field microprobe installed without Figure 82: Protective cap of the ICR near-field probe with protective cap lock function 17.4 Installing the BT 706 Bias-Tee The BT 706 bias-tee (Figure 83) has to be connected upstream to ensure the ICR probe's power supply.
  • Page 66: Collision Control Of The Icr Probe

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Figure 85: Connecting the SMA-SMA cable Figure 86: Connecting the power supply cable Connect the bias tee’s power supply cable to the input of the bias tee marked 12 V / DC and to the connector on the Z-axis tower marked 12 V DC (HR10 6-pin connector, blue) (Figure 86).
  • Page 67: Installing A Near-Field Probe With The Sh 01 Probe Holder

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Installing a Near-Field Probe with the SH 01 Probe Holder 18.1 Installing a Near-Field Probe on the FLS 106 IC First connect one end of the SSMB-SSMB cable to the SSMB connector of the rotary unit and lay the cable on the rotary unit's rotary ring as shown in Figure 87.
  • Page 68: Connecting A Near-Field Probe Directly To The Fls 106 Ic Scanner

    SH 01 and fastened in the correct position. 18.1.1 Connecting a Near-Field Probe Directly to the FLS 106 IC Scanner If operated without a pre-amplifier, the RF measuring cable has to be connected to the near-field probe's measurement output and directly to RF in input on the Z-axis tower (Figure 93).
  • Page 69: Connecting A Near-Field Probe With A Preamplifier

    Connecting a Near-Field Probe with a Preamplifier Note: The selection of the correct preamplifier depends on the measurement task and the near-field probe used. Please contact Langer EMV-Technik GmbH should you have any queries (Chapter 25). Connect the preamplifier's SMA output to the input marked RF in on the Z-axis tower (Figure 99).
  • Page 70 LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Figure 98: Exemplary preamplifier Figure 99: Preamplifier connected to the RF in input on the Z-axis tower Connect the near-field probe's RF measuring cable to the preamplifier's SMA input (Figure 100).
  • Page 71: Installing A Near-Field Probe On The Fls 106 Pcb

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 18.2 Installing a Near-Field Probe on the FLS 106 PCB The SH 01 probe holder is fixed to the aluminum mounting bracket with the two M3 countersunk-head screws.
  • Page 72: Connecting A Near-Field Probe Directly To The Fls 106 Pcb Scanner

    Connecting a Near-Field Probe with a Preamplifier Note: The selection of the correct preamplifier depends on the measurement task and the near-field probe used. Please contact Langer EMV-Technik GmbH should you have any queries (Chapter 25). Connect the preamplifier's SMA output directly to the input marked RF in on the Z-axis tower (Figure 107).
  • Page 73: Collision Control For The Near-Field Probe

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Figure 107: Preamplifier on the input marked RF in on the Z-axis tower Connect the near-field probe's RF measuring cable to the preamplifier's SMA input (Figure 108).
  • Page 74: Operational Notes

    A visual check of the scanner, components, attachments, measuring devices, DUT and cable guides should be performed before each measurement. Should any damage be discovered, this has to be eliminated before putting the scanner into service. Contact Langer EMV-Technik GmbH should you have any queries (Chapter 25).
  • Page 75: Estimating The Duration Of A Measurement With The Fls 106 Scanner

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 5. We recommend that the probe's further downward movement be blocked once the desired height has been reached. The positioning commands can be deactivated in the ChipScan-Scanner software under Settings ->...
  • Page 76: Memory And Storage Requirements For The Measurement

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 19.4 Memory and Storage Requirements for the Measurement Both the hard-disk storage and the RAM storage must be large enough for storing and showing a scan's measurement results.
  • Page 77: After Completing The Measurement

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de After Completing the Measurement After completion of the scan, the probe should be allowed to travel up until it has reached a safe distance to the DUT. If an ICR probe has been used, allow the probe to travel up until the complete protective cap can be put on the ICR probe tip and locked into place.
  • Page 78: Removing The Preamplifier / Bt 706 Bias Tee

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de 21.5 Removing the Preamplifier / BT 706 Bias Tee First remove the measuring cable from the signal input of the pre-amplifier / bias-tee and the power supply cable from the 12 V input on the Z-axis tower by hand.
  • Page 79: Maintenance

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Maintenance 22.1 General Information The FLS 106 scanner should be cleaned on the outside at regular intervals. We recommend that a dry, soft cloth be used for this purpose. Use a moist cloth and a neutral cleaning agent to remove heavier soiling.
  • Page 80: Dismantling

    At the end of its service life, the FLS 106 scanner has to be dismantled and disposed of in an environmentally friendly way. Unless a return and disposal agreement has been concluded with Langer EMV-Technik GmbH, the scanner has to be properly dismantled and recycled.
  • Page 81: Disposal

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Disposal 24.1 Disposal of Packaging Material Risk of environmental hazards due to improper disposal of packaging materials! Caution! Do not dispose of packaging materials in normal household waste.
  • Page 82: Customer Service

    LANGER DE-01728 Bannewitz mail@langer-emv.de FLS 106 IC set / FLS 106 PCB set EMV-Technik www.langer-emv.de Customer Service Please contact us if you have any queries, hints and suggestions. You can reach us: Monday – Friday 8:00 am to 3:00 pm...
  • Page 83: Warranty

    This document may not be copied, reproduced or electronically processed, either in its entirety or in part, without the prior written permission of Langer EMV-Technik GmbH. The management of Langer EMV-Technik GmbH assumes no liability for damage that may arise from using this printed information.