The Manufacturer Thought Technology Ltd. 5250 Ferrier, Suite 812 Montreal, Quebec Canada H4P 1L3 MyoTrac™ EMG Biofeedback System System Name System No. T4000P/T4001P MyoTrac™ Biofeedback Unit Device Name Device No. SA4000P/SA4001P EMERGO EUROPE Prinsessegracht 20 The Hague, 2514 AP The Netherlands ...
Page 3
Type BF Equipment Internally powered equipment Continuous operation Read Instruction CAUTION RxOnly US Federal Law restricts this device to sale by or on order of licensed health care practitioners. WARNING Do not operate MyoScan sensor within 10 feet of an operating cellular phone, similar radio transmitting device, other powerful radio interference producing sources such as arc...
Page 4
Make sure to remove electrodes from sensor snaps immediately after use. Do not plug third party sensors directly into instrument inputs. Plug only Thought Technology Active Sensor cable connectors into instrument inputs. All EMG electrodes and third party sensors must be connected to MyoScan sensors, either directly or through an adapter.
Model No. MyoTrac (4000P 60Hz) Model No. MyoTrac (4001P 50Hz) Table of Contents Introduction ................1 Installing and Testing Your Battery ........2 Trying Your MyoTrac for the First Time ........4 Use of the MyoScan™ Sensor ..........4 Front Panel Controls .............. 8 Battery Compartment Controls ..........
Introduction Biofeedback is a rapidly developing scientific field that has grown out of advances in physiology, psychology and electronics. Ordinarily, we are unaware of the subtle internal body activities that are part of our everyday lives. Biofeedback uses sensitive electronics to detect and amplify theses activities in order to bring them to awareness.
Work with EMG and relaxation and tension began during the 1920's and 1930's through the pioneering efforts of Edmund Jacobson. Jacobson found that the more relaxed a muscle was, the lower its EMG activity level was. He also noted that this lowered EMG activity level was associated with a more general subjective sense of relaxation.
Page 9
Fig. 1 Opening the battery compartment Note: Remove old batteries promptly to prevent corrosion. Remove the battery if the device is not going to be used for periods longer than one month. Testing the Battery: To check the battery, turn the unit on, place the threshold potentiometer to 5, and the scale-reading switch to x1.
Trying Your MyoTrac for the First Time Plug the MyoScan sensor into the INPUT jack. Set the switches inside the battery compartment to: OFF, OFF, ABV, NAR (fig. 7). Set the gain switch to the x1 position. Turn the volume up. Set the threshold potentiometer to 10 (fig.
Page 11
Extender Cable (T8720M): Thought Technology's 450mm extender cable can be inserted directly into the sensor head (figure 3), being careful to align the alignment dot on the cable connector with the groove on the sensor head.
Page 12
The protected pins are designed to prevent any possibility of direct skin contact with the metallic elements that conduct the amplified signals. They also help prevent accidental damage from attempting to plug Thought Technology sensors and encoders to equipment for which they were not designed. ...
Page 13
Replacement Sensor Cable (Model T9385M) Thought Technology sensors use a highly flexible and lightweight cable. If, for any reason, a break should occur in the cable, a replacement cable is available. Model No.SA9385M To change cables, tightly grip the connector plugged into the MyoScan with your thumb and index, and pull gently on the connector.
Front Panel Controls The front panel (see figure 6) includes controls for sound level (VOLUME), threshold level (THR), three gain settings, x1, x10 and x100 for ranges of .08-20, .08-200 and .08-2000 microvolt (V) RMS, respectively. A power switch lets you select OFF and either continuous proportional tone feedback (CONT) or threshold feedback (THR).
Page 15
ABV/BLW: This switch determines if the tone feedback is turned on when the signal exceeds (ABV) or falls below (BLW) the threshold potentiometer setting. CUE TONE/OFF: When either the 4 or 60 second switch is set, and the threshold is exceeded for at least 4 or 60 seconds, a continuous (not proportional) cue tone will sound.
The three threshold switches, LOCK/OFF, CUE TONE/OFF and BLW/ABV are only functional when the front panel power switch is set to THR. Possible combinations are: LOCK/OFF CUE/OFF ABV/BLW Proportional tone feedback above threshold. Proportional tone feedback below threshold. Cue tone when EMG above threshold for > CUE TONE 4 or 60 seconds.
Fig. 8 Side Panel Jacks WARNING: Never connect your MyoTrac to an A.C. powered device! Setting the Threshold Potentiometer The threshold control’s main function is to set a goal value for the EMG signal. When that value is reached, the tone is heard, which provides immediate feedback to reward or inhibit a particular behavior.
Page 18
The Gain setting determines the reading range: 0-20 V at x1, 0- 200 V at x10 and 0-2000 V at x100. For each gain setting, the possible threshold values are: .5-10 V (x1), 5-100 V (x10), and 50-1000 V (x100) respectively. Example of Threshold Setting: If you want to do a feedback session for the purpose of encouraging an increased level of activity in a specific muscle, initially set the...
Connecting to a Voltage Monitoring Device Connect the 0-2V output jack (see figure 8) to any instrument capable of reading from 0-2 Volts (digital voltmeter, analog voltmeter, polygraph, integrator, etc.). Make certain the instrument is battery operated or complies with IEC 601-1. The output voltage will always be in the 0-2 Volt range;...
Page 20
Connect and disconnect with care: When unplugging the sensor from the MyoTrac or the sensor cable from the sensor head, be careful to avoid pulling on the cable. Instead, grip the connector firmly and pull straight out. Always make sure to align the notch on the cable connectors (indicated by a dot) with the guiding groove.
90 days. During this time, in the unlikely event that repair is necessary, Thought Technology Ltd. will repair or replace your MyoTrac or MyoScan, free of charge. This warranty does not apply to damage caused by static electricity,...
Extended Warranty Program An Extended Warranty program is available directly through Thought Technology Ltd. and its authorized service dealers/distributors. The program provides for an additional two years of warranty coverage, extending the one year warranty to three years. All Extended Warranty Terms and Conditions apply to the second and third year of ownership and take effect following the regular one year warranty period.
Repair Return Form (Remove and include with the unit) Name __________________________________________________ Company _______________________________________________ Address ________________________________________________ Phone No. ______________________________________________ Fax No. ________________________________________________ E-Mail _________________________________________________ Date Purchased __________________________________________ Retailer _________________________________________________ Model No. ______________________________________________ Serial No. MyoTrac ______________________________________ Serial No. MyoScan _______________________________________ Return for credit (less than 30 days) YES ________ NO __________ Return for repair (clearly state problem) _______________________________________________________ _______________________________________________________...
8 channels simultaneously including EMG, EEG, skin conductance, temperature, heart rate, blood volume pulse, respiration and voltage inputs from other instruments. Thought Technology's active sensors provide unmatched signal fidelity while our fiber optic cable provides complete safety and signal accuracy.