Page 1
Ultrasonic Piezo Scaler Place Disk Owner’s Guide...
Page 2
Catalog Number 60014237 BioSonic Suvi Premier (tap water) USA Plug Type B 60014238 BioSonic Suvi Premier Plus (Medicament feeder) USA Plug Type B 60014241 BioSonic Suvi Premier (tap water) Japan Plug Type B 60014242 BioSonic Suvi Premier Plus (Medicament feeder) Japan Plug Type B...
Premier / Premier Plus Safety Safety Indications for Use The BioSonic Suvi Premier/Premier Plus Ultrasonic Scaler is intended for use in dental applications such as supra and subgingival scaling, periodontal therapy, endodontic procedures, cavity preparation and restorative dentistry. General Requirements Service of the product is only to be performed by authorized service personnel.
Page 7
Premier / Premier Plus Safety Safety notices in this manual Warning indicates a potentially dangerous situation. Non-observance may lead to death or injury. Caution indicates a potentially harmful situation. Non-observance may damage the equipment. Note indicates a situation where special notice should be observed.
Non-observ- ance may damage the equipment. 1 Unpacking Caution Carefully unpack your BioSonic Suvi Premier / Premier Plus unit and verify that all accessories and components are included according to Note indicates a situation where the content lists below: special notice should be ob- served.
BioSonic Suvi Premier / Premier Plus is highly adaptable to any procedure or user approach. It is not only an outstanding scaling and cleaning device, but it also brings power and versatility to endodontics, implantology, restorative treatments, minimally invasive treatments and apical procedures.
Premier / Premier Plus Equipment Description 1. AC power input 2. Fuse holder 3. Water hose coupling (optional) 4. Foot control pedal connection 5. Decal serial label 2.2 Control Panel 1. Cleaning key 2. POWER ON key 3. POWER OFF key 4.
Premier / Premier Plus Equipment Description 2.3 Foot Control Pedal 11. Connection for foot control pedal cable 12. OFF position 13. ON position 14. Diagnostic function, turn LED light ON or OFF by a single-click on the pedal 15. Irrigation position 16.
Equipment Description 2.4 Medicament Bottle (Premier Plus) BioSonic Suvi Premier Plus has a medicament dispenser system, making the device independent of a fixed water supply connection. The medicament bottle can be used for either medicament solutions or ordinary clean water.
Premier / Premier Plus Equipment Description 2.5 Symbols on the equipment Working mode 1 = 0 to 40% power Working mode 2 = 0 to 70% power Working mode 3 = 0 to 100% power Irrigation mode Dry mode Automatic cleaning function Power ON Power OFF Example of decal serial label.
Page 14
Premier / Premier Plus Equipment description Manufacturer Date of Manufacture European Union Authorized Representative IPX1 Ingress Protection Rating, Device has dripping water protection Autoclave at 132 Caution: Federal (United States) laws restricts this device to sale by or on the order of a physician or dentist.
Premier / Premier Plus Installation 3 Installation 3.1 General Installation Instructions Checklist • Position the unit horizontally with the handpiece in the holder and the hose hanging free. • Position the unit where air is free to circulate on all sides and be- Caution neath it.
Premier / Premier Plus Installation General Installation Instructions, continued • Verify that the voltage rating on the rear side matches the voltage WARNING of the AC power outlet. The unit must be connected • Verify that the AC power outlet is provided with a protective to an AC power outlet ground.
Premier / Premier Plus Operating Instructions 4 Operating Instructions Preparations (Premier Plus) 1. Fill the medicament bottle with water or medicament solution according to the Medicaments that can be used, listed on page 26. WARNING 2. Screw the bottle connector onto the bottle and push it onto the con- A tip that is bent, altered, or nector.
Page 18
Premier / Premier Plus Operating Instructions 4. Turn on the Premier / Premier Plus scaler by pressing the ON key on the control panel. The blue indicator light illuminates and the scaler is activated. 5. The indicator lights next to the working modes are flashing to For scaler tips without a fluid remind that a working mode has to be selected.
Premier / Premier Plus Operating Instructions How to use the scaler The side of the tip-end shall be applied to How to use the scaler 4.2 How to use the Premier / Premier Plus Scaler the tooth so that the movement of the tip The side of the tip-end shall be applied to is parallel to the surface of the tooth.
2. All exterior surafaces may be soaked in surface disinfectant to remove gross soil. Examine the Piezo tip for damage and discard if excessive wear, broken tip or distortion is noted. 3. Clean thoroughly using an ultrasonice cleaner such as the Coltène/Whaledent BioSonic Ultrasonic Cleaning System with BioSonic UC32 Solution ® ® Concentrate.
WARNING fully open (dots in line, see BioSonic Suvi Handpiece picture). Caution indicates a potentially 5.2 Routine Cleaning Procedures harmful situation. Non-observ- Beginning of the day ance may damage the equipment.
Premier / Premier Plus Cleaning and Maintenance 5.3 Maintenance Power cord Inspect the power cord, cables and the handpiece hose daily to insure that the equipment is in good condition without mechanical damage. O-rings (bottle connector) Lubricate the O-rings regularly with a glycerine based, water soluble A petroleum based lubricant lubricant.
Premier / Premier Plus Troubleshooting 6 Troubleshooting Premier / Premier Plus Type of problem Please see A. Unit is not responding and no lights are lit on the control panel Page 23 B. Lights are flashing on the control panel Page 24 C.
Page 24
2. Check that the tip is not worn out. Use a tip check card to evaluate wearing or try an unused tip. Use original BioSonic Suvi tips for reliable performance. 3. If you have an additional handpiece, try replacing the handpiece.
Page 25
Premier / Premier Plus Troubleshooting E. I nsufficient or no water flow when foot control pedal is activated 1. If the unit is equipped with a medicament bottle: Check that the medicament bottle is properly connected (i.e. the cap is tightened and pushed in all the way onto the connector). Check that the o-rings on cap and connector are in good condition.
Premier / Premier Plus Technical Data 7 Technical Data Coltène/Whaledent Inc. Manufactured for: 235 Ascot Parkway Cuyahoga Falls, OH 44223 / USA Model Premier / Premier Plus Classification EN60601-1: Class 1, Type B 93/42 EU: Medical products, class IIa L x W x H 270 x 110 x 165 mm (without bottle) Weight 2900 g...
Page 27
Premier / Premier Plus Technical Data Water supply pressure 1 - 10 bar (0.1–1.0 MPa, 14.5–145 PSI) (version conn. to tap water) Water comsumption 10 - 50 ml/min Bottle volume 500 ml (bottle version) Medicaments that can be used - Premier Plus Unit •...
Premier / Premier Plus Warranty and Declaration of Conformity 8 Warranty and Declaration of Conformity 8.1 Warranty Terms All items are manufactured to impeccable standards and have been developed specifically for use in dentistry and are intended to be operated only by dental professionals in accordance with the instructions contained in this guide.
Premier / Premier Plus Warranty and Declaration of Conformity 8.2 Certificate of Limited Warranty CERTIFICATE OF LIMITED WARRANTY Coltène/Whaledent (C/W) warrants this product to be free from defects in workmanship and materials for a period of twelve (24) months from the date of original purchase. During such warranty period, C/W undertakes to repair or replace this product, at its option, if such product does not function properly under normal use and service, which malfunction is due solely to a defect in workmanship or material.
Guidance and manufacturer's declaration - electromagnetic emissions The BioSonic Suvi is intended for use in the electromagnetic environment specified below. The customer or the user of the BioSonic Suvi should assure that it is used in such an environment. Emission test...
Page 31
Guidance and manufacturer's declaration - electromagnetic immunity The BioSonic Suvi is intended for use in the electromagnetic environment specified below. The customer or the user of the BioSonic Suvi should assure that it is used in such an environment. Compliance...
Page 32
RF communications equipment and the BioSonic Suvi The BioSonic Suvi is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the equipment can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the BioSonic Suvi as recommended below, according to the maximum output power of the communications equipment.
Catalog No. Unit Description Quantity 60014239 Biosonic Suvi Elite (tap water) USA Plug Type B 600014240 BioSonic Suvi Elite Plus (Medicament Feeder) USA PLug Type B 60014259 Medicament Feeder Bottle woth connector, 500ml 60014260 Polisher Powder Contaner, Transparent 60014261 Handpiece - Piezo...
Page 34
Premier / Premier Plus Order Information Biosonic SUVI Tips Catalog No. Description Category 60014265 PE-31 Heavy Calculus Removal 60014266 PE-32 Heavy Calculus Removal 60014267 PE-39 Heavy Calculus Removal 60014268 PE-41A Heavy Calculus Removal 60014269 PE-37 Universal tip for Scaling 60014270...
Need help?
Do you have a question about the Suvi Premier and is the answer not in the manual?
Questions and answers