  • Page 1 PERMA PURE LLC Zero-Air™ Generator User’s Manual PERMA PURE LLC 1001 NEW HAMPSHIRE AVE LAKEWOOD NJ 08701 (732) 244-0010 ▪ (800) 337-3762 ▪ fax (732) 244-8140 ▪ www.permapure.com ▪ info@permapure.com...
  • Page 2: Table Of Contents

    Zero-Air™ Generator User’s Manual PERMA PURE LLC Doc. ZA-MAN-001 REV 01 Page 2 of 10 Contents Unpacking ........................... 3 Introduction ........................3 General description ......................4 Specifications and features ....................5 Important safety warnings ....................5 Start-up procedure ....................... 6 Maintenance ........................
  • Page 3: Unpacking

    Zero-Air™ Generator User’s Manual PERMA PURE LLC Doc. ZA-MAN-001 REV 01 Page 3 of 10 Unpacking Perma Pure has made every effort to ship you a high quality product that has been thoroughly inspected and tested. It has been carefully packed to ensure that it arrives at your facility in good condition.
  • Page 4: General Description

    Zero-Air™ Generator User’s Manual PERMA PURE LLC Doc. ZA-MAN-001 REV 01 Page 4 of 10 General description Perma Pure’s portable Zero-Air generator (ZAG) is a light-weight, self-contained system capable of producing 18 lpm of high purity, particle free dry air. The dual head diaphragm pump draws ambient air of into the ZAG where it is compressed then purified using a combination of membranes, adsorbents, and filters as shown in figure 1.
  • Page 5: Specifications And Features

    Zero-Air™ Generator User’s Manual PERMA PURE LLC Doc. ZA-MAN-001 REV 01 Page 5 of 10 Specifications and features • Pump: Dual head diaphragm, up to 30 psig, 18 slpm • Inlet filter: Sintered SST, 10 micron porosity ® • Dryer: PD-100T-12MZA Nafion gas dryer •...
  • Page 6: Start-Up Procedure

    Zero-Air™ Generator User’s Manual PERMA PURE LLC Doc. ZA-MAN-001 REV 01 Page 6 of 10 Start-up procedure 1. Place ZAG on level, dry surface. 2. Check to make sure power switch is in “OFF” position. 3. Plug unit into standard 115VAC electrical outlet. 4.
  • Page 7: Maintenance

    Zero-Air™ Generator User’s Manual PERMA PURE LLC Doc. ZA-MAN-001 REV 01 Page 7 of 10 Maintenance WARNING !! De-pressurize and unplug the ZAG before attempting any maintenance or removing the top or bottom panels! Routine maintenance consists of periodic carbon replacement and outlet filter replacement. Under normal, intermittent use, desiccant should last up to six months.
  • Page 8: Final Filter Replacement

    Zero-Air™ Generator User’s Manual PERMA PURE LLC Doc. ZA-MAN-001 REV 01 Page 8 of 10 Final filter replacement The final particulate filter should be replaced each time the activated carbon canister is changed. The filter is located immediately after outlet of CO catalyst canister. 1.
  • Page 9: Appendix A: Warranty And Disclaimers

    Zero-Air™ Generator User’s Manual PERMA PURE LLC Doc. ZA-MAN-001 REV 01 Page 9 of 10 Appendix A: Warranty and disclaimers Perma Pure LLC Perma Pure (Seller) warrants that product supplied hereunder shall, at the time of delivery to Buyer, conform to the published specifications of Seller and be free from defects in material and workmanship under normal use and service.
  • Page 10: Appendix B: Wiring Diagrams

    Zero-Air™ Generator User’s Manual PERMA PURE LLC Doc. ZA-MAN-001 REV 01 Page 10 of 10 Appendix B: Wiring diagrams Figure 4 -Wiring diagram 115 VAC...