If you have any questions, please contact Type BF Applied Part Novis Healthcare on 1300 738 885. Alternating Current Novis Healthcare has a policy of continuous product Manufacturer improvement and reserves the right to amend Manufacturing Date specifications presented in this guide. Information Serial Number correct at time of production (February 2019) .
System Overview The ProCair Quad is an alternating mattress The system consists of the following components: replacement system for the prevention and ⬡ Mattress replacement with umbilical air hoses treatment of skin breakdown and pressure injuries and CPR release in patients of high to very high risk. It is designed ⬡...
Patient conditions for which the application of bases. pressure therapy on the ProCair Quad mattress replacement systems contraindications include: The ProCair Quad control unit can be fitted to the foot ⬡ Instable spinal cord injury board of most hospital or aged care beds.
Only the control unit and mattress disconnect the power cord from the control combination as indicated by Novis unit. Novis Healthcare recommends using Healthcare should be used, otherwise the the cord retention loops on either side of the...
Page 7
Ensure there is no moisture in or near the heated surfaces. Should the power cord be found power inlet, power switch and power cord to be worn or damaged, contact Novis Healthcare before reconnecting the power supply. for a replacement.
Carefully unpack the system and inspect each item for any damage that may have occurred during transit and handling. Any damage or missing components should be reported to Novis Healthcare as soon as possible. Confirm there are no sharp objects in the immediate area which may risk damage to the mattress replacement.
Page 9
Ensure the umbilical cord is not trapped between the mattress and bed. Failure to do so could result in an under inflated mattress leading to patient injury. ProCair Quad User Guide...
Page 10
The system automatically defaults to Alternating Mode at start up, with AutoCair feature (automatic weight detection) always in operation to select an appropriate pressure for patient support according to mattress load, unless overridden manually. User Guide ProCair Quad...
You may need to unzip the cover to feel the cells Place the patient on to the ProCair Quad for inflation. mattress. With the patient lying supine, unzip one side...
SENSOR DISCONNECTED - umbilical connector not FLASHING plugged in properly or not detected NORMAL FLASHING: CHARGING FLASHING BATT. NORMAL FLASHING: CHARGING BATT. SERVICE DUE SERVICE DUE IN USE IN USE <30% FLASHING: LOW BATT <30% FLASHING: LOW BATT User Guide ProCair Quad...
Page 13
When when the alarm is turned off. periodical servicing is required, the amber Standby light will flash. Please contact Novis Healthcare for system maintenance. This equipment must only be serviced by a qualified service agent.
3 years and should be replaced by patients to remain in a seated position while an authorised Novis repair agent. minimising the risk of bottoming out. Once the head of the mattress is brought back to a CPR Release reclined position (or below a 5°...
To deactivate the alarm, press BED EXIT button clicks are heard. Air will remain in the system for up to again, the ON light will extinguish to indicate the 24 hours, depending on patient and environmental alarm is off. circumstances. ProCair Quad User Guide...
Once air has been released from the system, detach the mattress from the bed by unfastening the straps, then fold and roll the mattress from head end to foot end for storage. Return all items to the custom carry bag for safe keeping. User Guide ProCair Quad...
This could result in damage to the equipment and may result in damage to the polyurethane coating, or negate the biocompatibility properties of the fabric. Cleaning and Infection Control It is recommended that the ProCair Quad system is cleaned every two weeks if in constant use. Top Cover Cleaning Control Unit Cleaning...
Page 18
Clean the dust from the filter or discard and replace with a new filter. Refit the air filter cover to the control unit before use. It is recommended that the air filter be replaced each year. Replacement air filters are available. User Guide ProCair Quad...
Air cells have failed to resumed. alternate ALTERNATION FAILURE If issue persists, contact Novis Healthcare – a service may be required. Control unit has failed to detect connection to Ensure air hose connector is securely fastened and reconnect if loose.
Page 20
For faster mattress reinflation, press MAX INFLATE and wait until the LP alarm code stops flashing. Press the ALT button to resume alternation. If the problem persists, move patient to an alternate product and contact Novis Healthcare. ProCair Plus User Guide...
Technical Specifications MODEL ProCair Quad ProCair Quad King Single SYSTEM CODE APMPC-R04 APMPC-R04K CAPACITY 40 - 250 kg SYSTEM 20, including 3 static head cells and 7 disconnectable heel cells. NO OF CELLS All alternating cells include a static lower chamber (cell-in-cell)
Page 22
(landfills) that are not appropriate according the ProCair Quad system or any of its components. legislation. Please be environmentally responsible and contact your local authority on available options to recycle this product at its end of life.
Page 23
Guidance and manufacturer’s declaration – electromagnetic emissions The ProCair Quad control unit is intended for use in the electromagnetic environment specified below. The customer or the user of the ProCair Quad control unit should ensure that it is used in such an environment. Emissions test Compliance Electromagnetic environment –...
Page 24
If the measured field strength in the location in which the ProCair Quad control unit is used exceeds the applicable RF compliance level above, the ProCair Quad control unit should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the ProCair Quad control unit.
Page 25
The carrier shall be modulated using a 50 % duty cycle square wave signal. As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case. ProCair Quad User Guide...
To Novis will undertake at its option, to repair or replace, obtain compensation, you will need to provided free of charge, each product or part thereof on the...