7
How do I copy setup?
Press the COPY or QUIT function key to copy parameters
into the memory bank you are using.
Press #2 or #8 to move from one memory bank to another.
8
How do I calibrate the pump?
Connect the distal end of the tubing for the device you are going to
use i.e., syringe, bag or bottle. Press the START button and after
dispensing approximately 70% press the STOP button. Calculate the
volume that was dispensed; press the CAL key on the right side.
The LCD Screen will display:
Enter the volume that was dispensed by using the numeric keypad
and follow it by pressing ENTER, then press QUIT key to resume.
The pump will resume from the amount you entered and stop at the
amount required.
Note: Whenever QUIT appears on the screen you may press the
function key beneath it and return to the main menu.
Foot switch may be used in lieu of start/stop button for dispensing
by plugging into round receptacle on rear of pump.
9
How do I clean the pump?
The pump can be cleaned easily by using Isopropyl Alcohol and a clean soft cloth. Apply alcohol directly
to the cloth and wipe down the outside of the housing. The area around the keypad and the motor housing
are sealed so alcohol will not seep into the inner workings of the pump.
©Acacia
M-0151 031913
Refer to the pump instruction manual for
additional manual operation details.
Call 1-800-486-6677 for assistance.
785 Challenger Street, Brea, CA 92821
Ph: 714.257.0470 Fax: 714.257.0513
Select COPY
Press the START button
Press the QUIT key
to resume