Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
Page 2
Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 3
Confidential Information The material contained herein consists of information that is the property of LUXTRON Corporation and intended solely for use by the purchaser of the equipment described in this manual. All specifications are subject to change without notice. Changes are made periodically to the information in this publication, and these changes will be incorporated in new editions.
1.0 Introduction This chapter provides an overview of the Model 710, 712, and 790 LUXTRON Fluoroptic thermometers. It also describes the major components and default settings of the instruments. Figure 1.1 shows a typical rack mount installation of the instrument.
LUXTRON Corporation. Each probe consists of a fiber-optic cable with a connector at one end and temperature sensor mounted at the other end. Contact the LUXTRON Sales Department for more information. Model 710, 712, and 790 Fluoroptic Thermometer User Manual...
Firmware The instrument is set up, operated, and maintained using its built-in, menu-driven firmware. It can be controlled locally or remotely: • Locally, from the front of the instrument. The front panel of the instrument front panel has a single-line display of 24 characters and a set of six keys for running the firmware.
Default Setup Parameters Information describing how to change parameters for particular applications are provided in Chapter 3, “Using the Instrument.” Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
How to Use this Manual This manual explains how the LUXTRON Model 710, 712, and 790 ® Fluoroptic thermometers work and provides information on how to set up, operate, and maintain these thermometers. These Fluoroptic Thermometer models, despite their advanced technology, are simple to use.
Used when information is critical to avoid personal injury. Warning Used when information is critical to avoid personal injury when an electrical hazard is present. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
• One shielded RS-232 cable Contact the LUXTRON Customer Service Department immediately if there is any visible damage to the instrument or probes or if any items are missing. If everything is in order, save the packing material for possible future use.
Page 15
Do not uncoil the probe until instrument power-on testing is complete. Screw probe metal connector into a front panel probe connector. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 16
Figure 2.1 shows the front view of the stand alone installation of the instrument (Model 790 only). Figure 2.2 shows the front view of the rack mount installation of the instrument (all models). Figure 2.3 shows the rear view of the instrument. An overall view of each model is provided in Appendix A, “Specifications.”...
Page 17
Figure 2.2 Front Panel—Rack Mount, All Models Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
To test the instrument Power-on the instrument, using the power switch on the front panel. The display scrolls through the following single-line messages, in the order listed: LUXTRON CORPORATION (MODEL NUMBER) (C) COPYRIGHT (PART NUMBER) (FIRMWARE VERSION) SETUP Unpacking and Installing the Instrument Artisan Technology Group - Quality Instrumentation ...
Replacing this cap when the probe is not in use prevents dirt from accumulating on the open fiber end. Occasionally clean the connector with spectrograde isopropyl alcohol or methanol. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
3.0 Using the Instrument This chapter describes how to use the front panel buttons, as well as how to use Setup, Calibration, and Run modes. Setup Mode Setup mode allows displaying and changing setup parameter values. Use Setup mode for initial setup or for when the application changes and different types of probes or system devices are required.
Page 21
CHANGE DIGITAL PORT? MENU PREV NEXT SELECT PORT: RS232 MENU PREV NEXT Figure 3.1 Setup Mode Menu Tree Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 22
CHANGE BAUD RATE? MENU PREV NEXT 9600 RS-232 BAUD RATE: MENU PREV NEXT DIGITAL FORMAT: FULL MENU PREV NEXT ENABLE TIME STAMPS: MENU PREV NEXT CHANGE IEEE ADDRESS? MENU PREV NEXT IEEE DEVICE ADDRESS: MENU PREV NEXT CHANGE ANALOG OUTPUT? MENU PREV NEXT...
Page 23
DISPLAY CURRENT TIME? MENU PREV NEXT XX:XX:XX XX:XX:XX MENU PREV NEXT Figure 3.1 Setup Mode Menu Tree (continued) Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 24
RESTORE DEFAULTS? MENU PREV NEXT REINITIALIZE? MENU PREV NEXT ARE YOU SURE? MENU PREV NEXT SETUP MENU PREV NEXT Legend Active buttons Note Inactive buttons Default is NO. Return to the Main Menu MENU Press + to toggle to YES. Return to the previous menu level PREV Toggle channel on or off...
Page 25
“ ” – l l o . ) t Table 3.2 Setup Menu First-Level Messages and Definitions (continued) Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 26
i f i i c l ° ( “ :” F ° – °C ° o i t n i t l l o “ ” , y r t t u i r t . s t a l f i g i l a t o i t...
Page 27
6 9 , o i t n i t Table 3.2 Setup Menu First-Level Messages and Definitions (continued) Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 28
“ ” i f i f l l – r o + b i t h t i i l r o i t • n i t • y l t “ ” • – r o + o i t ’...
Page 29
Table 3.2 Setup Menu First-Level Messages and Definitions (continued) 3.10 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 30
i l b t s r . t n o i t “ ” . s r – r o + i f i o i t i t n “ ” . s r i f i – i t n .
Page 31
– b i t c i t n Table 3.2 Setup Menu First-Level Messages and Definitions (continued) 3.12 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 32
l l e “ ” – r o + “ ” – r o + y l f “ , ” Table 3.2 Setup Menu First-Level Messages and Definitions (continued) Using the Instrument 3.13 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Figure 3.2 illustrates the Calibration mode menu tree. Table 3.3 defines each menu selection and describes how to change the parameters of each. 3.14 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 34
SETUP PREV NEXT MENU CAL CORRECTION IS: CAL CORRECTION IS: MENU PREV NEXT MENU PREV NEXT SETUP PROBE: PREV NEXT PREV MENU NEXT MENU Notes Press – or + to turn CAL CORRECTION off or on. INPUT REF TEMP: 25°C Follow appropriate branch of menu tree, based on whether CAL CORRECTION is on or off.
Page 35
Table 3.3 Calibration Menu Display Messages and Definitions 3.16 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
r i f i l i i l i o l l i l i i l a i l a i l a s l i i l a c i t y l l i l a t e l Table 3.3 Calibration Menu Display Messages and Definitions (continued) Run Mode...
Page 37
Legend Active buttons Inactive buttons Figure 3.3 Run Mode Menu Tree 3.18 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
4.0 Theory of Operation This chapter discusses the temperature sensor and how the instrument calculates temperature measurements. Temperature Sensor The temperature sensor is the key element in the system. It is mounted on the end of the probe. The sensor consists of a small amount of a temperature sensitive phosphor, which is manganese-activated magnesium fluorogermanate.
Page 39
(RS-232 port or optional IEEE-488 interface). -200 -100 Temperature (°C) Figure 4.3 Time/Temperature Plot of Magnesium Fluorogermanate Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Exciting Light Pulse Fluorescent Signal Data Points Time Figure 4.4 Plot Representation of Method for Extracting Fluorescent Decay Time How Temperature Measurements are Calculated The instrument calculates temperatures based on the SET SAMPLES PER MEAS Setup parameters. Figures 4.5 through 4.7 SET MEAS UPDATE TIME illustrate three examples of temperature measurement.
Page 41
SAMPLES/MEASUREMENT One measurement per sample 1 second Continuous Measurement Update Time Figure 4.7 Example of Fast Temperature Measurement Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
LUXTRON designates as repairable. Information for contacting LUXTRON is provided in Appendix F, “Warranty and Service.” Maintenance and Troubleshooting Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
, e l b i l Table 5.3 Temperature Related System Error Messages Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Return the unit to LUXTRON for flashlamp bulb replacement. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Appendix A: Specifications This appendix defines the specifications for the accuracy, precision, and performance of the instrument. Table A.1 lists the basic performance specifications. Supplemental information provided in sections to follow. b i l ± ° ± ( ° ± °...
For minimum warm-up time and maximum instrument life, keep the system powered on at all times, even when not in use. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Precision or Repeatability Root mean square (RMS) and Standard Deviation are equivalent terms that describe a statistical method for quantifying the spread of the distribution curve for measurement data. The following briefly defines these terms, relative to the specification of the instrument. In practical terms: •...
Operation of this instrument in a residential area is likely to cause harmful interference, in which case users must correct the interference at their own expense. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Overall View Figures A.2 through A.5 provide an overall view of each instrument model. Figure A.2 Overall View—Model 710 (Available Only as Rack Mount) Specifications Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 53
Figure A.3 Overall View—Model 712 (Available Only as Rack Mount) Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
LUXTRON Corporation. Each probe consists of a fiber-optic cable with a connector at one end and temperature sensor mounted at the other end. Contact the LUXTRON Sales Department for more information. The instrument uses both immersion and surface measurement (industrial) probes, depending on the application.
Probe Length Signal level decreases as the fiber length increases. The losses experienced vary by fiber type. The use of multiple connectors and extensions also reduces signal level. A good rule of thumb is that the signal level is reduced by half for each additional connection. The lower the signal, the worse the signal-to-noise ratio.
Page 59
Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Appendix C: RS-232 Serial Output Port This appendix discusses the serial port of the instrument, as well as complete setup and diagnostics for this feature. Overview The serial port on the instrument meets Electronics Industry Association (EIA) standard RS-232. This port is used for communication with another digital instrument (such as a serial printer, terminal, or computer).
(default is CONTINUOUS) DATASAVE Feature Although this option is available for selection, it is currently nonfunctional. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Hardware Configurations for the RS-232 Port To use the RS-232 port of the instrument, use a shielded RS-232 cable that is wired straight through (for example, pin 1 to pin 1, pin 2 to pin 2, pin 3 to pin 3, and so forth). It is recommended that a fully shielded RS-232 cable be used in electrically noisy environments.
A carriage return and line feed, <CR><LF>, are normally transmitted at the end of each line. FULL is the default format for Model 790. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 64
(2Dh) in positions 6 through 12 and the two digit error code appears in positions 13 and 14. Model 710 is a one-channel instrument and only shows data for FIELD 1 and FIELD T in ABBR format. Each temperature measurement is sent in Fields 1 through 2, each of which has 14 characters.
Page 65
Table C.3 Temperature Message Format Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Command Set The RS-232 serial port can be used to control or program the instrument by sending the proper control character. Each control character or command is echoed back from the instrument. If the command is not allowed or the system is currently in that mode, the instrument returns a “?”...
Main menu with the front panel buttons locked, awaiting further commands. Table C.7 lists the Remote mode control commands. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 68
Table C.7 Remote Mode Control Commands While the instrument is in Remote Run mode, the Initiate Reading command starts a measurement cycle and disables the transmission of digital output. For example, to take a single measurement using the remote enable command, follow the sequence described below.
If it becomes necessary to abort the calibration sequence, send CTRL- A (01h) during the sequence. C.10 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Upon completion, the instrument returns to the Main menu and displays one of the following messages on the front panel and as output on the serial port: • means the temperature is unstable. The temperature readings were unstable during the calibration period and prevented the instrument from calibrating.
Page 71
< > — o i t < > Table C.10 Remote Setup—Query System Status Command Field Descriptions C.12 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 72
Figure C.5 shows the response provided by the instrument to a query. Table C.11 describes each field. command = configuration parameters<CR><LF> Figure C.5 Remote Setup—Query System Status Response — — f n i < > < > Table C.11 Remote Setup—Query System Status Response Table C.12 lists the remote setup commands and the required syntax.
Page 73
. ) t i l a ° F ° r o D Table C.12 Remote Setup Commands C.14 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Sample Remote Setup Commands The RS-232 port echoes each command it receives, with a carriage return and a line feed. Embedded spaces are ignored. Table C.13 lists six sample setup commands. < > < > < > < > < >...
Page 75
C.16 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Appendix D: Optional IEEE-488 Interface This appendix discusses the optional IEEE-488 interface, as well as complete setup and diagnostics for this option. The IEEE-488 option is used for communicating with other computers. It can be used as an output port for logging measurements, and/or as an input port for remote setup or control.
The instrument must use a unique address on the IEEE bus or it conflicts with other devices using the same address. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 78
Table D.2 lists the digital output settings that LUXTRON sets prior to shipping an instrument with the optional IEEE-488 interface. Table D.2 IEEE-488 Option Digital Output Settings If default settings have been restored (as described in the section, “Reinitializing the Instrument,” in Chapter 5, “Maintenance and Troubleshooting”), reprogram Setup parameters as follows.
There are two ways to control the instrument over the digital port: • Run the instrument continuously • Take measurements on the command of the controller Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 80
To run the instrument in Continuous mode Send CTRL-T (14h) to confirm instrument is in Standby mode. This assures the instrument is in Standby mode at the Main menu and not running. If the instrument is in Standby mode, it echoes “...
Mode,” in Chapter 3, “Using the Instrument,” for calibration instructions. • means that calibration completed successfully without any error. The instrument is ready for use. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 82
Figure D.1 shows a typical calibration result string. DC 1, 25.00,"CU"<CR><LF> One-Channel Instrument DC 1, 25.00,"CU"; 2, 25.00,"CC"<CR><LF> Two-Channel Instrument DC 1, 25.00,"CU"; 2, 25.00,"CC"; 3, 25.00,"CC"; 4, 25.00,"CC"<CR><LF> Four-Channel Instrument Figure D.1 Remote Calibration Result String The strings in Figure D.1 show the calibration temperature to be 25 °C (77 °F).
— — — Table D.4 Serial Poll Status Byte Definitions Table D.5 lists typical serial poll status responses. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
i l a e l l l l o e l l l l o s t i Table D.5 Typical Serial Poll Status Responses Remote Instrument Programming Remote programming has two basic functions: • Programing system for a measurement •...
Page 85
Remote Programming—Query String < > o i t < > < > Table D.7 Remote Programming—Query String Field Descriptions D.10 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 86
The instrument posts an SRQ when it has a response to a query. Perform a serial poll and verify that instrument responds (that is, that the bit is PROG set in the serial poll status register). The controller must then perform a read operation to obtain response.
Page 87
. ) t i l a ° F ° r o D Table D.9 Remote Programming Commands D.12 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
— , e l — , e l Table D.9 Remote Programming Commands (continued) Sample Remote Programming Commands The IEEE-488 port echoes each command it receives, with a carriage return and a line feed. Embedded spaces are ignored. Table D.10 shows six sample programming commands. <...
To order a copy of this standard (product number SH15446), see the section, “IEEE-488 Overview,” for information on contacting the IEEE. D.14 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 90
Figure D.5 shows the IEEE digital output format used by the instrument. DC 1, -50.11 <CR><LF> One-Channel Instrument DC 1, -50.11;2, 300.22 <CR><LF> Two-Channel Instrument DC 1, -50.11;2, 300.22; 3, 125.00;4, 200.00 <CR><LF> Four-Channel Instrument Figure D.5 IEEE Digital Output Format The messages in Figure D.5 are shown exactly as they appear on a monitor or printer.
Page 91
Figure D.7 shows the measurement result when only Probe 1 is active. DC 1, -50.11 <CR><LF> Figure D.7 IEEE Measurement Output Message Format with Only Probe 1 Active D.16 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 92
Each data field consists of three subfields, separated by commas (2Ch), as shown in Figure D.8. Table D.12 describes each field. Embedded spaces are ignored. pn, temp, "status"<CR><LF> Figure D.8 IEEE Measurement Output Message Data Fields r i f o l l c i l i a t °...
Page 93
The next measurement immediately follows the temperature subfield (see Figure D.11). DC 1,101.00<CR><LF> Figure D.11 IEEE Digital Output Format—No Status to Report D.18 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
FULL and ABBR Digital Output Message Formats When using the FULL or ABBR digital output formats, all remote programming commands use the equal sign (“ ”) instead of a space between the command and parameters. <ESC>MU=2S<CR><LF> Figure D.12 Remote Programming Command Format for FULL or ABBR Digital Output Formats For more information, see the section, “FULL and ABBR Digital Output Message Formats,”...
Page 95
D.20 Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Appendix E: Analog Output Options This appendix discusses the following options: • Analog Output Voltage (AOV) ±5 V Standard and ±10 V option • Analog Output Current (AOC) option Analog Output Voltage (AOV) This option, which is programmable from the Setup menu (see the section, “Setup Mode,”...
Page 97
+5 V, and a low limit trip sets the output to -5 V. These signals can be used to power relays or other control devices. Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
AOV ±10 V Relationship between Scale Factor, Temperature Range, and Resolution Note Units with the standard ±5 V range must be returned to LUXTRON for conversion to the ±10 V range. Analog Output Options Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Figure E.3 AOC Option—External Power Supply Pin Diagram Notes Model 710: Pins 3 through 8 are not used. Model 712: Pins 5 through 8 are not used. The current loop option replaces the analog voltage output and must be installed by LUXTRON.
Page 100
Figure E.4 shows the formula for calculating the range of the current, using the highest and lowest measured temperatures as input. Table E.4 lists the definitions for each variable. (mA) = SF * (T ) + 4 mA meas Figure E.4 Formula for Calculating the Current µ...
Page 101
Model 710, 712, and 790 Fluoroptic Thermometer User Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
In the event of failure of the instrument covered by this warranty, LUXTRON Corporation will repair and calibrate the instrument if it is returned to LUXTRON within one year of the original shipment, provided an examination discloses that the product is defective.
Shipments to LUXTRON for Repair All shipments of LUXTRON instruments should be made prepaid and insured by way of United Parcel Service or Best Way. For overseas customers, units should be shipped air freight, priority one. The instrument must be shipped in the original packing container or its equivalent.
Page 104
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...