DEVICE FOR DIAMAGNETIC THERAPY

Advertisement

Table of Contents
loading

Summary of Contents for Periso CTU MEGA 20

  • Page 1 DEVICE FOR DIAMAGNETIC THERAPY...
  • Page 2 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 PERISO SA...
  • Page 3: Table Of Contents

    CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Index 1. INTRODUCTION ........................4 Introduction to the CTU Mega 20 device ................4 Introduction to the manual ...................... 4 Safety precautions and warnings ..................5 Shipping damage ........................7 Contents of the package ......................7 Warranty Conditions........................
  • Page 4: Introduction

    All information may be subject, without notice, to change by the Manufacturer for purposes and reasons related to technical or business improvements. The use of the CTU Mega 20 Diamagnetic Pump implies knowledge of this manual in all its parts. Usage not conform to the Manual automatically invalidates the warranty and liability by the Manufacturer.
  • Page 5: Safety Precautions And Warnings

    CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Safety precautions and warnings Before connecting the CTU Mega 20 Diamagnetic Pump device to the power supply you must ensure that it has the following characteristics: Mains voltage 230 Volt (+/- 10%)
  • Page 6 If the power cable is damaged, it must be replaced only by the manufacturer Periso or by personnel appropriately qualified by Periso. Any repair work on the CTU Mega 20 product must be carried out by the manufacturer Periso or by personnel appropriately qualified by Periso.
  • Page 7: Shipping Damage

    Contents of the package CTU Mega 20 device will be shipped carefully packed in a casing that also contains all the accessories and the manual, provided with the basic equipment. Below is a list of items included in the casing of delivery.
  • Page 8: Warranty Conditions

    Warranty Conditions Refer to the Service and Warranty Booklet provided in the shipping wrap. Service life expected for CTU Mega 20 device is 10 years. Area of Jurisdiction For any dispute regarding these conditions the exclusive jurisdiction is the one in Lugano (CH), in Italian Language.
  • Page 9 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Service messages that appear at the top of the screen. Label symbols Temperature Limitation. CE mark. Indicates the Name Humidity Limitation. and Address of the Manufacturer. Atmospheric pressure Consult the Instruction limitation.
  • Page 10: Writing Conventions

    3. INSTALLATION 3.1 Installation Instructions The CTU Mega 20 device must be extracted from the front of the box after removing the safety locks. Then remove all accessories and check its integrity. The device comes partially assembled, the only operations to be performed are:...
  • Page 11 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Assemble secure shelf with appropriate and supplied screws. Fix the control module on the trolley with four supplied screws. Button to move up the arm. Button to move down the arm.
  • Page 12 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 1) Connect the female DB9 pin connector to its male connector on the back of the power module. 2) Connect the plug of the control module to the socket. 3) Connect the motor plug of the arm to the connector.
  • Page 13 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 7) Plug the connectors of the water pipes of the handpiece to the respective sockets, applying slight pressure. Removable cap for topping up with dematerialized water. 8) Connect the device to the power network.
  • Page 14: Operating Handpiece

    Periso SA recommends to carry out treatments lasting 20/30 minutes and leave the CTU MEGA 20 device on standby without performing any treatment for at least 10 minutes, to avoid overheating of the handpiece. Replacing of the operating handpiece...
  • Page 15: Operating Instructions

    MOD07-2-5 Rev.13 of 28/08/2019 4. OPERATING INSTRUCTIONS 4.1 Required skills The CTU Mega 20 Diamagnetic Pump is a device intended to be used by medical staff and/or physiotherapist. It can be used in: Physiotherapy Clinics and Centers Hospital Rehabilitation Clinic Aesthetic Medicine 4.2 Side effects...
  • Page 16 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 select from three possible applications: Movement of liquids; Endogenous biostimulation; Pain Control. Select one of the operating modes among those displayed. Depending on the operating mode required, select the working parameters related to the section Magnetic Field and relative to the section Radio Frequencies.
  • Page 17 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Endogenous Biostimulation Biostimulation It corresponds to the amplitude of electrical stimulation induced within the tissues. You can operate selectively, for each specific tissue, choosing the parameter corresponding to the activity that is intended to get on at cellular and sub-cellular level.
  • Page 18: Radiofrequency For Capacitive And Resistive Diathermy

    CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Pain Control In this operating mode you can treat acute or chronic pain. The parameters on which to interact are Joule and Hz. The Hz represent the repetition frequency of the pulses generated by the magnetic field.
  • Page 19 It is important to remember that with the CTU Mega 20 there is the automatic recognition of the capacitive and/or resistive electrode, so it must not be preselected on the display.
  • Page 20: Indications, Contraindications And Precautions

    1 session/day for a maximum period of 3 weeks. EXCLUSION CRITERIA Before performing the treatments with PEMF CTU Medical Device – PERISO SA, all the patients received a clinical evaluation to exclude patients with: open physis, inflammatory joint disease. Acromegaly, Charcot’s arthropathy, Haemochromatosis, Wilson’s disease, Ochronosis, terminal illnesses/malignancies, pregnancy or lack of...
  • Page 21 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 EXCLUSION CRITERIA Before performing the treatments with PEMF CTU Medical Device – PERISO SA, all the patients received a clinical evaluation to detect: unsuitable physiological states (pregnancy or lack of contraception use in women of childbearing age), open physis, terminal illnesses/malignancies, , use of pacemaker or any implanted electrical device, ferromagnetic implants in the area of the treatment.
  • Page 22: Contraindications

    (for a total of 20 applications). EXCLUSION CRITERIA Before performing the treatments with PEMF CTU Medical Device – PERISO SA, all the patients received a clinical evaluation to detect: unsuitable physiological states (...
  • Page 23: Precautions

    In case of absence of personnel trained in the use of the machine, turn off the device by the general switch on the system. Handle with care the CTU Mega 20 handpiece, making sure not to cause damage (scratches, drops, bumps, etc.).
  • Page 24 The device will have to be disposed of in accordance with standards for the disposal of WEEE. All devices from PERISO SA are marked with the following symbol: According to art. 13, paragraph 4 of the Legislative Decree no. 151/2005, all electrical and...
  • Page 25: Maintenance And Trouble Shooting

    Make sure that the device is disconnected from the power supply when cleaning Before cleaning the neutral plates, disconnect them from the device. Only use the cream produced by Periso SA for treatments with the CTU Mega 20 Diamagnetic Pump.
  • Page 26 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 If the red light is on, only after having disconnected the device from the mains, open the red cap of the tank (indicated in the image above) and fill the tank using only demineralised water and the supplied bottle.
  • Page 27: Troubleshooting

    CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Connection Type of Verification Frequency to the mains supply Power on Daily Integrity of neutral plates cables Daily Integrity of the handpiece cables Daily Integrity of the power supply cable Daily Monthly Integrity of power connection cable –...
  • Page 28 The plug of the control shelf is regularly pushed forward, on the back of the power module? The CTU Mega 20 is connected to a 230V 50 / 60 Hz power source? The key has been turned to the ON...
  • Page 29 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Make sure there are no devices that emit strong electromagnetic fields nearby. Turn off the device, disconnect all connection cables, reinsert them, switch the device back on. On the display of the horizontal or vertical...
  • Page 30 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 operate the device for about 10-15 minutes and then verify that the light is green/blue again; again off the device; reposition the handpiece on the arm; reboot the device and try that the flow of the magnetic field is correct.
  • Page 31 If the problem persists, please contact PERISO SA Service and support to send the device to the service center. It is obligatory to make a careful description of the problem by sending a written communication to the service center to the following e-mail...
  • Page 32: Technical Specifications

    CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 7. TECHNICAL SPECIFICATIONS Model CTU Mega 20 Commercial Denomination Diamagnetic Pump General power supply 50/60 Hz 230 Vac Current Consumption 9.5 A Power Absorption 2200 VA (800 W) Conversion factor 0.36 Working frequency 1 Hz –...
  • Page 33: Information Provided By The Manufacturer

    CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 8. INFORMATION PROVIDED BY THE MANUFACTURER Label of the control module Label of the power module PERISO SA...
  • Page 34 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Label of the mobile arm Label of the handpiece Serial number label Label on the connector of the handpiece PERISO SA...
  • Page 35 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Label on the trolley Applied parts connections labels (neutral plates (M) – handpiece (N)) PERISO SA...
  • Page 36 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Packaging label Model: CTU MEGA 20 30.00.001 ID PERISO: HS Code: 9018.9084 Via Senago 42 d 6912 Pazzallo Switzerland Dimensions: 80x60x158cm 144 kg +/- 5% Weight: PERISO CODE DESCRIPTION A388.019 1 User and Maintenance Manual A388.021...
  • Page 37 CTU MEGA 20 deve garantire che esso venga usato in tale ambiente. The Equipment CTU MEGA 20 is intended for use in the electromagnetic environment specified below. The customer or the user of the Equipment CTU MEGA 20 should assure that it is used in such an environment.
  • Page 38 CTU MEGA 20 deve garantire che esso venga usato in tale ambiente. The equipment CTU MEGA 20 is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment CTU MEGA 20 should assure that it is used in such an environment.
  • Page 39 (30% buco in_dip in U (30% buco in_dip in U interruptions, it is recommended to power the CTU MEGA 20 device with an per_for 25 cicli_cycles per_for 25 cicli_cycles uninterruptible power supply or a battery.
  • Page 40 Portable and mobile RF communications equipment should be used no closet to any part of the equipment CTU Mega 20 , including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
  • Page 41 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 da 80 MHz a 800 MHz IEC 61000-4-3 da 80 MHz a 2,5 GHz 80 MHz to 800 MHz 80MHz to 2,5 GHz da 800 MHz a 2,5 GHz 800 MHz to 2,5 GHz Dove P è...
  • Page 42 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 Le intensità di campo per trasmettitori fissi come le stazioni base per radiotelefoni (cellulari e cordless) e radiomobili terrestri, apparecchi di radioamatori, trasmettitori radio in AM e FM e trasmettitori TV non possono essere previste teoricamente e con precisione.
  • Page 43 RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE EQUIPMENT CTU MEGA 20 L’apparecchio CTU MEGA 20 è previsto per funzionare in un ambiente elettromagnetico in cui sono sotto controllo i disturbi irradiati RF. Il cliente o l’operatore dell’apparecchio CTU MEGA 20 possono contribuire a prevenire interferenze elettromagnetiche assicurando una distanza minima fra gli apparecchi di comunicazione mobili e portatili a RF (trasmettitori) e l’apparecchio CTU MEGA 20 come sotto raccomandato, in relazione alla potenza di uscita massima degli...
  • Page 44 CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019 PERISO SA Via Senago 42 d 6912 Pazzallo CH Tel. +41(0)919359400 info@periso.ch www.periso.ch N. IVA CHE-101.514.101 IVA PERISO SA...