Contents Safety General Information Getting Started Operation of PCMT Calibration Specifications Guarantee and service Thermo-shaker PCMT Version 1.01 - January 2013 Page 1 Operating Instructions...
Do not impede the platform motion. Do not operate the unit in environments with aggressive or explosive chemical mixtures. Do not operate the unit if it is faulty or been incorrectly installed. Thermo-shaker PCMT Version 1.01 - January 2013 Page 2...
Page 4
Do not check the temperature by touch. Use a thermometer. BIOLOGICAL SAFETY It is the user's responsibility to carry out appropriate decontamination if hazardous material is spilt on or inside the equipment. Thermo-shaker PCMT Version 1.01 - January 2013 Page 3 Operating Instructions...
2. General Information Thermo-Shaker PCMT is an ideal instrument for intensive mixing of samples with heating or cooling. Heating (up to +100°C), cooling (up to 15°C below room temperature - RT) and mixing modes can be used both simultaneously and independently i.e. the unit is three products in one: 1) a shaker, 2) a thermostat and 3) a thermo-shaker.
Automatic balancing system in this product produces light metal-like noise when moving the unit which is likely to be heard during unpacking. It is normal occurrence and does not indicate a fault. The Thermo-shaker PCMT set includes: • Thermo-shaker PCMT ..................1 piece • Microtube block ....................1 piece •...
Page 7
Disconnect the external power supply unit from the device. Remove the four knurled screws (fig.2/1); disconnect the plug (fig.1/1). Select the new block and install it as point 3.5 Fig.2 Thermo-block set up Thermo-shaker PCMT Version 1.01 - January 2013 Page 6 Operating Instructions...
4. Operation of PCMT Recommendations during operation • Please check that the tubes/microwell plates are thermo resistant. Don't heat the tubes/microwell plates over the melting point of the material they are made of (use thermo resistant polypropylene tubes). Remember that thin-walled tubes have a higher thermo conducting factor.
Page 9
25°C (The display will show OFF – T, °C – set point) Program execution After the PCMT has reached a stable temperature (when the set and current temperature readings become the same): Insert tubes into the block or place the microwell plate on the block and close the lid.
Hold down the TIME-RUN/STOP key (Fig. 3/5) for more than 8 sec. to activate calibration mode. The following parameters will be shown on the display (Fig. 5) Fig.5 Control panel in calibration mode Thermo-shaker PCMT Version 1.01 - January 2013 Page 9...
Page 11
5.12. The display will show calibrated temperature as shown on fig.6B/3 and the unit will continue thermal stabilization according to the previously set temperature. Thermo-shaker PCMT Version 1.01 - January 2013 Page 10 Operating Instructions...
Input current/power consumption ..................12V, 4.9 A / 60W External power supply unit ........... input AC 100-240 V 50/60Hz, output DC 12V Weight with block and external power supply, not more ..............3.7 kg Thermo-shaker PCMT Version 1.01 - January 2013 Page 11...
Page 13
HC18 HC24 block for 24x2 ml microtubes block for 96-well PCR microplate (semi-skirted, unskirted) HC96 PCMT blocks and PHMT blocks are not interchangeable, i.e. PCMT blocks cannot be used on the PHMT and vice versa. Replacement parts Description Rubber belt 117x5x0.6 mm...
Page 14
1. Disconnect the external power supply from the mains. 2. Remove 4 fixation screws on the shaker bottom and remove the bottom plate. 3. Replace the rubber belt. 4. Re-assemble the unit. Thermo-shaker PCMT Version 1.01 - January 2013 Page 13 Operating Instructions...
Page 15
Thermo-shaker PCMT Version 1.01 - January 2013 Page 14 Operating Instructions...