Page 2
EHD-2350SWIR (1000nm to 2350nm) E H D -2500SWIR (1000nm to 2500nm) All contents in this manual must not be copied without permission by EHD imaging GmbH. Attention to this manual (1) Before you use this product, read this manual completely in order to fully understand the contents.
IMPORTANT SAFETY PRECAUTION In this product and manual, many symbols are used to highlight warnings and cautions to read so that this product can be used properly to avoid users or their surrounding persons from any injuries and any losses of their property. ...
Page 5
This product with power supply other than a designated voltage could result in fire or electric shock. Do not touch any cables of this product after it begins to thunder. Touching the cable in thunder could result in electric shock. ...
Page 6
Do not put anything on this product. It could result in malfunction. Do not touch this product with its power on for a long time. Because its surface temperature tends to be high, it could result in cold burn of skin. ...
Page 7
EMC Directive (2004/108/EC) EHD-2350SWIR / EHD-2500SWIR complies with EMC Directive (2004/108/EC). AC limitation. For DC supply to this product, use an indoor DC supply system with lightening countermeasure or a DC supply instrument such as a stabilized power source other than the indoor supply system.
Table 1 shows the main components of EHD-SWIR series Table 1 Main components of EHD-SWIR series Item Product Number Remark EHD-2350SWIR NIR area camera EHD-2500SWIR Power cable LF10WBP-4S ASSY 5m length OKI Electric Cable CL-S-MS-050 CameraLink cable (for grabber board)
2 Preparation before use 2-1 Appearance Figure 2 shows the appearance drawing of EHD-SWIR series. Power connector CameraLink connector (158) Trigger/alarm connector Filter Figure 2 Appearance drawing of EHD-SWIR series 2-2 Setup 2-2-1 Installation When the camera body of EHD-SWIR series is installed, use a tap on the bottom of the camera body in Figure 3.
2-3 Connection 2-3-1 Power cable Connect EHD-SWIR series to power source with DC +24V by using the attached power cable. Pay careful attention to the polarity of the power cable. Camera side: one 4-pin round receptacle connector Power source side: four pieces of cut-off bear cables Power connector pin# Voltage...
2-3-3 Lens Prepare a lens compatible for objective performance for magnification rate, field of view and mount the lens to the front panel of EHD -SWIR series . Mount type of lens: C-mount Caution: In mounting a lens, do not screw it by more than 10 mm in the flange of C-mount. Too much screw-in could result in damage of FPA.
3 Specifications Table 2 Specification of EHD-2350SWIR / EHD-2500SWIR Item Specs Detector InGaAs / GaAsSb type-II quantum well Wavelength range 1000nm to 2350nm (EHD-2350SWIR) 1000nm to 2500nm (EHD-2500SWIR) Pixel number 320 (H) x 256 (V) / 9.6x7.68mm / ¾" (100fps, 9ms), (100fps, 1ms), (150fps, 6ms), Frame rate, (200fps, 4.5ms), (200fps, 1ms), (250fps, 3.5ms),...
Page 13
Table 3 Requirement of power supply for EHD -SWIR series Item Specs Output voltage DC 24V Maximum current ≥ 10 A Voltage accuracy ± 10 % Voltage ripple ≤ 480mVpp Operation voltage & temperature Comparable to EHD-SWIR series Storage voltage & temperature Comparable to EHD-SWIR series Protective function Overload and overvoltage protection...
4 Operation 4-1 Start-up (1) After making sure of connections between EHD-SWIR series, cables and the frame grabber, switch on the PC and power supply. (2) Start to supply +24V to EHD-SWIR series by the power supply. (3) Wait for about 5 minutes for stabilization. (4) Execute a sample software for image capture of an attached CD in the PC.
For disposal of this product, return it to us or distributors for this product. 8 Contact Regarding information on this product, please make contact with: EHD imaging GmbH Zum Rennplatz 15, D-49401 Damme (Germany) Tel: +49-5491-2090, email: info@ehdimaging.de Web: www.ehd.de 14 / 14 ...