  • Page 1 USER MANUAL Micromotor for oral surgery 01_26/04/2018 USER MANUAL DSI MICROMOTOR FOR ORAL SURGERY 0476 Page 1 of 15...
  • Page 2: Intended Use

    Micromotor for oral surgery 01_26/04/2018 INTENDED USE The DSI Mini Surgery Motor product are brushless micromotor for oral surgery. The use of DSI Mini Surgery Motor is allowed only to specialized personnel (licensed medical specialists) in hospitals, nursing homes, clinics and private practices.
  • Page 3 USER MANUAL Micromotor for oral surgery 01_26/04/2018 ATTENTION 1. Conformity to the Electromagnetic Compatibility Directive DOES NOT GUARANTEE the total absence of interferences with other electronic devices; therefore: The instruments MUST NEVER be used on or near patients who have pace-makers or other electronic implants. •...
  • Page 4 USER MANUAL Micromotor for oral surgery 01_26/04/2018 DESCRIPTION OF VERSIONS AND COMPONENTS Pic. 1. Control unit. Micromotor with led. Variable speed pedal. Power supply cord. Micromotor support. Protective plug for sterilization. Bar for physiological solution. Pic. 2. Pic. 3. Main switch. Power supply socket.
  • Page 5: Installation

    Make sure enough air is able to circulate around the equipment to cool it down. • It is risky and severely forbidden to alter or in any case modify the characteristics of the product. DSI Dental • Solutions shall not be held responsible for any personal injury or equipment damage caused by failure to heed this precaution.
  • Page 6 USER MANUAL Micromotor for oral surgery 01_26/04/2018 • Put down the control unit on a stable base. • Insert the bar into the hole and assemble the support of the micromotor. Pic. 4. • Connect the micromotor connector to the control unit. •...
  • Page 7 USER MANUAL Micromotor for oral surgery 01_26/04/2018 • Connect the irrigation tube with the physiological solution bag and with the handpiece. Pic. 6. • Connect the foot-switch (or the foot-control) to the panel connector. • Place the pedal in a easy-to-use position. •...
  • Page 8 USER MANUAL Micromotor for oral surgery 01_26/04/2018 Pic. 7. Function keys: Key 1  Set in succession the parameter to modify (SPEED, TORQUE, PUMP, REDUCT): the adjustable parameter is highlighted. Key 2  Increase the highlighted parameter. Key 3  Decrease the highlighted parameter.
  • Page 9: Maintenance

    USER MANUAL Micromotor for oral surgery 01_26/04/2018 Pedal functions The handpiece can be activated only by pedal. Pic. 8. 1. Activation of the micromotor with variable speed (in proportion to the pressure on the pedal) until the maximum set value. 2.
  • Page 10 USER MANUAL Micromotor for oral surgery 01_26/04/2018  Switch off the device using the switch and unplug the power cord. Disconnect all components from the control unit and separate all accessories: Preparation before cleaning • micromotor. • micromotor support. • pedal.
  • Page 11 • Description of the electro-magnetic fields with renard to the EN 60601-1-2 Directive: The DSI Mini Surgery Motor device must function in the electro-magnetic field specified below. The customer/ • user of DSI Mini Surgery Motor must guarantee that it is used in this environment.
  • Page 12 CONFORMITY LEVEL ELECTROMAGNETIC ENVIRONMENT Portable and mobile radiofrequency communications devices must not be used near the DSI Mini Surgery Motor (including its cables) at a distance less than that recommended and calculated according to the frequency and power of the emitter.
  • Page 13 DSI Mini Surgery Motor, according to the maximum output power of the equipment, as recommended in the table below. SEPARATION DISTANCE IN METRES (M) ACCORDING TO EMITTER FREQUENCY Rated max.
  • Page 14: Technical Features

    The illegal disposal of the product by the holder implies the application of the administrative sanctions provided for by the current legislation. The Declaration of conformity to Directive 2012/19 / EU must be requested from the DSI Dental Solutions company. TECHNICAL FEATURES...
  • Page 15: Label Symbols

    It is afraid of humidity RAEE appliance, separate collection object BF type electro-medical device Fragile, load and discharge with care Made in Italy for DSI Dental Solutions Address: 59 haAvoda st., Ashdod Israel Phone: +972 8931 7235 Email: info@dsisrael.com Fax: +972 8995 9449...