MEDMASTER
External Power Supply for MRI Luminaires
MRIPS SERIES
INSTALLATION INSTRUCTIONS
This instruction sheet applies for the Kenall MRIPS-312 External Power Supply. It describes the proper mounting
procedure and wiring instructions for singular and multiple fixture configurations. A separate instruction sheet is
provided with the MRI luminaire(s). Both sheets together are intended to provide comprehensive instructions on
system installation and electrical wiring.
The Kenall MRIPS-312 power supply and EMI filter (by others) must be installed outside the shielded MRI
environment. It contains ferrous components that can damage or interfere with MRI equipment.
www. kenall.com
P: 800-4-Kenall
INSTALLATION INSTRUCTIONS
When using electrical equipment, basic safety precautions
should always be followed, including the following:
THIS PRODUCT MUST BE INSTALLED IN ACCORDANCE WITH THE
APPLICABLE INSTALLATION CODE BY A PERSON FAMILIAR WITH
THE CONSTRUCTION AND OPERATION OF THE PRODUCT AND
THE HAZARDS INVOLVED. DISCONNECT POWER TO ALL CIRCUITS
BEFORE WIRING FIXTURE. INSTALL IN ACCORDANCE WITH
ALL NATIONAL, STATE, AND LOCAL CODES. DO NOT CONNECT
TO AN UNGROUNDED SUPPLY. READ ALL FIXTURE MARKINGS
AND LABELS TO ENSURE CORRECT INSTALLATION OF FIXTURE.
SUPPLEMENTAL INSTRUCTIONS MAY BE LOCATED ON THE
FIXTURE, IN ADDITION TO THIS INSTRUCTION SHEET, REGARDING
ORIENTATION, OR MOUNTING RESTRICTIONS.
SAVE THESE INSTRUCTIONS
F: 847-360-1781
1020 Lakeside Drive Gurnee, Illinois 60031
1
MRIPS312_F-2846_110113

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MEDMASTER MRIPS Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for kenall MEDMASTER MRIPS Series

  • Page 1 MRI luminaire(s). Both sheets together are intended to provide comprehensive instructions on system installation and electrical wiring. The Kenall MRIPS-312 power supply and EMI filter (by others) must be installed outside the shielded MRI environment. It contains ferrous components that can damage or interfere with MRI equipment.
  • Page 2 4. Secure conduit to low voltage and line voltage compartments using either the front or side KO locations. Class 1 wiring is required. 5. Make 120-240VAC, 50/60Hz supply connections on line voltage side. Make sure proper ground is connected. www. kenall.com P: 800-4-Kenall F: 847-360-1781...
  • Page 3 7. If a 0-10V dimmer is to be connected, install at this time. See Figure 2. Dimmer must be installed outside the shielded MRI environment with an intermediate MRI room EMI filter to the luminaires. Kenall recommends using a Leviton® lllumaTech™ IP710 series dimmer or Grafik Eye GRX-TVI control interface. Contact Kenall for suitability of using an alternate sink-type 0-10V dimmer.
  • Page 4: Customer Service

    Kenall’s entire liability on any claim of loss or damage resulting from a defective product is limited to the replacement price of the product. The foregoing warranty is exclusive of all other warranties and no other warranties of any kind are expressed or implied.
  • Page 5 1020 Lakeside Drive Gurnee, Illinois 60031 MRIPS312_F-2846_110113 www. kenall.com P: 800-4-Kenall F: 847-360-1781 1020 Lakeside Drive Gurnee, Illinois 60031 Content of speci cation sheets is subject to change; please consult www.kenall.com for current product details. © 2013 Kenall Mfg. Co. All rights reserved.
  • Page 6 INSTALLATION INSTRUCTIONS MRI INSTALLATION REGISTRATION FORM For warranty purposes, please fill out this form and return to Kenall by fax at (847) 360-1781. I certify that the lighting installation for the listed MRI suite location is completed per the provided installation instructions and to the best of my abilities.

This manual is also suitable for:

Medmaster mrips-312

Table of Contents