Medxing ESM101 User Manual

Wrist type blood pressure monitor

Advertisement

Quick Links

Shenzhen Med-link Electronics Tech Co., Ltd
4th and 5th Floor, Building Two, Hualian Industrial
Zone, Xinshi Community, Dalang Street, Longhua
District, 518109 Shenzhen, CHINA
Fax:+86 755 61120055
E-mail:kefu@medxing.com
Website:www.medxing.com
Shanghai International Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537 Hamburg, Germany
Tel: +49-40-2513175
Fax: +49-40-255726
Digital Blood Pressure Monitor
User Manual
ESM101
R
E
V
Please read the user manual carefully before using this device.
A
.
*
The picture is for reference only, the actual item is the standard.
0
(Wrist Type)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ESM101 and is the answer not in the manual?

Questions and answers

Summary of Contents for Medxing ESM101

  • Page 1 Eiffestrasse 80, 20537 Hamburg, Germany Tel: +49-40-2513175 Fax: +49-40-255726 Digital Blood Pressure Monitor (Wrist Type) User Manual ESM101 Please read the user manual carefully before using this device. The picture is for reference only, the actual item is the standard.
  • Page 3: Table Of Contents

    Table of Contents 1. Important Safety Instructions ● Never diagnose or treat by yourself only based on the 1. Important Safety Instructions blood pressure measurement result since self-diagnose or 1.1 Warning self-treat may make the illness get worse. Please follow 1.2 Precaution the doctor’s directions.
  • Page 4: Precaution

    ● Do not open, alter or repair this device by yourself. ● Do not pull the wrist cuff with excessive force. Do not twist it. ● Only pump up the cuff once fitted. ● Do not impact, drop, trample and shake the device. ●...
  • Page 5: Preparation Before Use

    3. Preparation before Use 4. Take a Blood Pressure Measurement 3.1 Install or Replace Batteries 4.1 Important Notes Open battery compartment cover ⑦ and insert batteries to ensure the To ensure the measurement accuracy please follows the below suggestions: polarity (<+> and <->) of the batteries are correct and then close it. If the ●...
  • Page 6: Correct Body Posture During Measurement

    4.3 Correct Body Posture during Measurement ● If you feel any uncomfortable such as the cuff is inflated for a long time it may be a risk. You must stop the ● Support your arm in a relaxed position and ensure that the device is at measurement by pressing ON/OFF button ①.
  • Page 7: Memory

    4.5 Memory 5. Error Indication 4.5.1 Memory Recall The below table explains the error code that may display on the LCD: When the device is power off or stand-by press MEMORY button ⑥ it will display the average value of the last three measurement results. You Error code Description Remedy...
  • Page 8: Troubleshooting

    6. Troubleshooting 7. Maintenance and Storage When the device encounters malfunctions during the use, refer to table below: 7.1 Maintenance Abnormal situation How to correct ● Please often clean the device ● Use a soft, dry cloth to clean the surface of this device when it is dirty. (1) Check batteries’...
  • Page 9: Storage

    7.2 Storage 9. Technical Specifications ● Do not wet the wrist cuff. Product name: Digital Blood Pressure Monitor ● Protect the device from: ‐Water and moisture Display: Liquid Crystal Display (LCD) ‐Extreme temperature (heat and cold) Measurement method: Oscillometric ‐contamination and dust 1) Internally powered equipment;...
  • Page 10: Device Symbols

    10. Device Symbols 11. Electromagnetic Environment The following symbols may appear on the product labeling: Electromagnetic Interference Caution This device has been tested and found to comply with the limits for Refer to instruction Type BF applied part medical devices to the IEC/EN 60601-1-2. These limits are designed to manual/ booklet provide reasonable protection against harmful interference in a typical medical installation.
  • Page 11 The model ESM101 is intended for use in the electromagnetic environment The model ESM101 is intended for use in the electromagnetic specified below. The customer or the user of the model ESM101 should environment specified below. The customer or the user of the model assure that it is used in such an environment.
  • Page 12 The customer or the user of the model ESM101 should assure that environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the model ESM101 is used it is used in such an environment.
  • Page 13: About Blood Pressure

    12.About Blood Pressure Separation distance according to frequency of transmitter 12.1 What is Blood Pressure? Rated maximum 150 kHz to 80 80 MHz to 800 80 MHz to 800 output of Blood pressure (BP) is the pressure exerted by circulating blood upon the walls of transmitter blood vessels, and is one of the principal vital signs.
  • Page 14: What Is Morning Hypertension

    Category Diastolic Systolic Optimal Normal 80-84 120-129 and/or High normal 85-89 130-139 and/or Grade 1 hypertension and/or 90-99 140-159 Grade 2 hypertension 100-109 160-179 and/or Grade 3 hypertension ≥110 ≥180 and/or Isolated systolic hypertension ≥140 The blood pressure (BP) category is defined by the highest level of BP, whether systolic or diastolic.

Table of Contents