Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions IEC
61000-3-2
Voltage
fluctuations/ flicker
emissions IEC
61000-3-3
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic environment specified
below. The user of the Philips Avent breast pump should assure that it is used in such an environment.
IMMUNITY test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
SurgeIEC 61000-4-5
Compliance
Electromagnetic environment – guidance
Group 1
The Philips Avent breast pump uses RF energy only
for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The Philips Avent breast pump is suitable for
Class A
use in all establishments, including domestic
establishments and those directly connected to
the public low voltage power supply network that
supplies buildings used for domestic purposes.
Complies
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for
power supply
lines
± 1 kV line(s) to
line(s)
Compliance
Electromagnetic
level
environment – guidance
± 6 kV contact
Floors should be wood, concrete
± 8 kV air
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
± 2 kV for
Mains power quality should be
power supply
that of a typical residential or
lines
hospital environment.
± 1 kV line(s)
Mains power quality should be
to line(s)
that of a typical residential or
hospital environment.