Invacare Stratos Portable IRC 1720 Operator's Manual page 36

Compressor nebulizer system
Hide thumbs Also See for Stratos Portable IRC 1720:
Table of Contents

Advertisement

Available languages

Available languages

SECTION 6—TROUBLESHOOTING
PROBLEM
Nebulizer does not work,
and the light indicator is red.
Optional battery is not
working optimally.
Optional battery is not
charging.
Stratos™Portable
All manuals and user guides at all-guides.com
The thermal protector has been activated due
to:
a. The device operating outside the conditions
outlined in this manual.
b. The device has been operating near heat
sources or in hot environments.
Allow the nebulizer to cool down for a mini-
mum of 40 minutes. Wait until the RED light
turns off before turning the nebulizer on again.
Battery was not properly conditioned (charged
for at least 16 hours before first use).
If it has been in use for approximately four
months, it may be time to recondition the bat-
tery.
Battery will not charge if connected to the nebu-
lizer.
Check all power cord connections and cord
conditions.
Check power outlet.
36
SOLUTION
Part No 1130202