Savillex HPX-100 & HPX-200 Hotplates 1.0 Introduction This manual describes the Savillex HPX Series hotplates (referred to as HPX) and contains product specific information, current as of the date of publication, for the safe use and operation of this equipment.
Savillex HPX-100 & HPX-200 Hotplates 2.0 Safety Information This manual contains important operating and safety information. The user must carefully read and understand the contents of this manual prior to the use of this equipment. In addition, the equipment itself also has labeling and information that must be followed to ensure proper use and safe operation.
Savillex HPX-100 & HPX-200 Hotplates 2.3 Electrical Shock Hazard • Use a properly grounded electrical outlet of correct voltage and current handling capacity • Disconnect the power cord prior to maintenance and servicing • Use only the power cord supplied with the product •...
Savillex HPX-100 & HPX-200 Hotplates 3.0 About the HPX 3.1 Dimensional Specifications Model Heated Working Overall Weight Power Cord Surface Dimension Dimensions Length HPX-100 Length: 40.4 cm (15.9 in) 292 mm x 212 mm 11.61 kg 1.8 m (6.0 ft) All models (11.5 in x 8.3 in)
Should any components be missing or damaged, contact your local dealer or Savillex. We recommend retaining all packaging until it is determined that the equipment is functioning properly.
Savillex HPX-100 & HPX-200 Hotplates 3.5 Product Connections 1. Power Cord Inlet • Mating receptacle for supplied power cord 2. Heater Output • Heater output connector for HPX wiring harness 3. Fuse Holder • Location of replaceable fuse. Fuse for 120V HPX shown. Please note the fuse requirement for the voltage of your specific HPX.
Savillex HPX-100 & HPX-200 Hotplates 4.2 Operating Conditions • Indoor use • Altitude up to 2000 meters (6,500 feet) • Ambient temperatures of 17°C to 27°C • HPX installed on a flat surface at least 30.5 cm (12 in) from walls, 122 cm (48 in) from ceilings, and 30.5 cm (12 in) from other HPX if using multiple units...
Savillex HPX-100 & HPX-200 Hotplates HPX Properties • The HPX is constructed of ISO-molded carbon graphite material for quick thermal conduction and uniform surface temperature • The Teflon PFA surface coating is corrosion proof, temperature resistant and resists aggressive chemicals •...
Savillex HPX-100 & HPX-200 Hotplates 5.3 Setting Countdown Timer Function (optional) The temperature controller can be programmed to perform a timed heating function. The user programs a low temperature setting, a high temperature setting, and a total heating time up to 99 hours and 59 minutes.
Page 13
Savillex HPX-100 & HPX-200 Hotplates 5. Toggle the Advance Key to the ‘time remaining’ display as indicated by the digital readout t.rl (Figure 6) 6. Press the EZ Key to start the programmed countdown timer heating cycle. The numbers shown are the hours and minutes that have been entered for the timed heating cycle.
Multiple occurrences of the automatic shutdown feature WARNING! may indicate an unsafe operating condition. Disconnect and discontinue use of the HPX and contact Savillex or its authorized distributor. 6.0 Maintenance, Troubleshooting and Repair 6.1 Product Maintenance The HPX requires little maintenance to provide a reliable means of performing laboratory procedures at accurately controlled temperatures.
Contact Savillex 6.3 Service and Repair Service and/or parts may be obtained from Savillex. Service should always be performed by qualified technicians using Savillex parts and instructions. Please contact Savillex or its authorized distributor with any service or repair issues.
Savillex HPX-100 & HPX-200 Hotplates 7.0 Equipment Disposal This equipment must not be disposed of with unsorted waste. It is the user’s responsibility to correctly dispose of the equipment at life-cycle-end by handing it over to an authorized facility for separate collection and recycling.
Buyer for such product. This remedy is subject to return of the defective product to Savillex or its agent, freight prepaid. EXCEPT AS STATED IN THIS DOCUMENT, NO OTHER WARRANTY, EXPRESS OR IMPLIED, IS GIVEN OR AUTHORIZED BY SAVILLEX.
In the event that the programmer inadvertently makes a change to parameters that should not have been changed, the controller may not function as WARNING! intended. Contact Savillex its authorized distributor for instructions on how to restore the controller to the original settings. Advance Key...
Page 19
Savillex HPX-100 & HPX-200 Hotplates 2. Program a Calibration Offset 1. Press both the Up/Down Arrow Keys until Ai and SEt are displayed • Approximately 7 seconds The display will read oPEr prior to Ai and SEt. The Up and Down Arrow Keys must be pressed and held until Ai and SEt are displayed.
Page 20
Savillex HPX-100 & HPX-200 Hotplates 3. Lock the Controller 1. Press the Infinity Key and the Advance Key simultaneously for six seconds. The display will read CUSt FCtY (Figure 12). 2. Press the Down Arrow Key until LoC and FCtY are displayed (Figure 13) 3.
Savillex HPX-100 & HPX-200 Hotplates 8.3 CE Declaration of Conformity The Undersigned, representing the manufacturer Savillex Corporation 10321 West 70 Street Eden Prairie, MN 55344 USA Herewith declare that the products: HPX-100 (all models), HPX-200 (all models) laboratory hotplates conform to the essential requirements of the following EC Directive(s) when installed in accordance with the...