Summary of Contents for Viatom Checkme Lite

  • Page 1 Checkme Health Monitor Lite User ’s Manual………………….…….English 1-19 Manuel de l'utilisateur……..…Français 20-34 Benutzerhandbuch …………...Deutsch 35-50 Manuale d’uso……………………Italiano 50-63...
  • Page 2 About Checkme Intended Use The Checkme Lite health monitor is intended to be used for measuring, displaying, storing and reviewing of multiple physiological parameters including ECG, pulse oxygen saturation (SpO ) and systolic blood...
  • Page 3 ECG and SBP are intended for use with adult. The data and results provided by this device are for pre-check screening purpose only and cannot be directly used for diagnostic or treatment. 1.2 Outline Touch Screen Internal SpO sensor Micro-USB connector It connects with USB cable for charging.
  • Page 4 (Put right forefinger or middle finger on it.) 1.3 Main Screen The Main Screen is shown as below. Press an icon in the Main Screen will start a measurement, activate a function, or open corresponding menu. The device will enter Main Screen when: No operation is detected for 60 seconds in result screen, the device ...
  • Page 5: Getting Started

    No alarm system. Getting Started 2.1 Power On/Off Press the Home button to power on the device. Press and hold Home button for 2 seconds to power off the device. 2.2 Initial Settings The first time when the Checkme is powered on, you can set up your Checkme step by step.
  • Page 6 Any tests recently performed on you that required an injection of intravascular dyes. The Oximeter may not work if you have poor circulation. Rub your  finger to increase circulation, or place the SpO sensor on another finger. The Oximeter measures oxygen saturation of functional ...
  • Page 7 Don’t exert too much pressure on the ECG electrode, which may result in EMG (electromyograph) interference. Just hold gently and ensure good contact with the ECG electrode. Do not exert pressure on the finger that put in the SpO sensor. Just fit it inside but gently to ensure good blood perfusion.
  • Page 8 Find a qualified traditional cuff BP meter. Sit down and stay calm. Place the cuff on you left arm according to instructions. Select the <Settings> icon. Select <BP Calibration>, and then choose the right user. Ensure that the cuff is at the same level as your heart. Then start the blood pressure measurement from the cuff BP meter.
  • Page 9 electrode; For Lead I, press the left electrode to the left palm;  For Lead II, press the left electrode to the left lower abdomen;  Do not press the device too firmly against your skin, which may result in EMG ( electromyography ) interference.
  • Page 10 Settings 4.1 Opening Settings Menu To open the Settings menu, tap the <Settings> icon to open the menu as below. 4.2 Changing Sound Volume In the Settings menu, tap the <Volume> area to change the volume directly. Or use the quick setting by tapping the area on the left side of the screen.
  • Page 11 device will be deleted, and the device will be restored to the factory default settings. 4.9 Identify Software Version Choose <About> in the <Settings> menu to identify the software version of your device. Telling the version information when reporting a problem may help to identify and solve your problem.
  • Page 12: Maintenance

    Maintenance 6.1 Battery This monitor is designed to operate on rechargeable Lithium-ion battery. The battery is charged automatically when the monitor is connected to a powered USB port. On-screen battery symbols indicate the battery status as follow: The battery is fully charged. The solid portion represents the remained battery energy.
  • Page 13 drifts on the electrode is not stably and gently. stable or too much. 2. Try to keep perfectly 2. Hand or body may be still and test again. moving. or pulse 1. Finger may not be 1. Remove finger and rate shows no inserted correctly.
  • Page 14 Display 2.4” touch screen, color, backlit Connector Micro-USB connector Power Supply Battery type Rechargeable lithium-polymer battery Battery run time Daily check: > 200 times Charge time Less than 2 hours to 90% Lead type Integrated ECG electrodes Lead set Lead I, lead II Record length Sampling 500 Hz / 16 bit...
  • Page 15 , PR, PI, Summary (Normal Blood Measurement summary Oxygen, Low Blood Oxygen, Unable to analyze) Blood Pressure Variation Measurement method Cuff-free non-invasive technology systolic pressure based on individual Measurement summary calibration coefficient Review Waveform review Full disclosure waveform Storage 100 pcs of records...
  • Page 16: Electromagnetic Compatibility

    Electromagnetic Compatibility The device meets the requirements of EN 60601-1-2. All the accessories also meet the requirements of EN 60601-1-2 when in use with this device. Warnings and Cautionary Advices Using accessories other than those specified in this manual may ...
  • Page 17 Fluctuations domestic purposes. Flicker Emissions 61000-3-3 Guidance and Declaration - Electromagnetic Immunity The Health Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Health Monitor should assure that it is used in such an environment. Immunity test IEC60601 Compliance...
  • Page 18 for 5 cycles for 5 cycles product requires 61000-4-11 70 % UT 70 % UT continued (30 % dip in (30 % dip in operation during power mains for 25 cycles for 25 cycles interruptions, it is <5 % UT <5 % UT recommended that (>95 % dip in...
  • Page 19 MHz to MHz to distances: IEC61000- 2.5 GHz 2.5 GHz 80 MHz~800 MHz:  800MHz-2.5GHz:  Where, P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
  • Page 20 RF communications equipment and the device The Health Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Health Monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the monitor as recommended below, according to the maximum output power of the communications equipment.
  • Page 21 Borkstrasse 10 · 48163 Muenster · Germany TEL: +49 251 32266-0 FAX: +49 251 32266-22 PN:255-00425-00 Version: A Dec, 2017 Contents of this manual are subject to changes without prior notice. ©Copyright 2014-2017 Shenzhen Viatom Technology Co., Ltd. All rights reserved.