Conducting An Optics Check Test - Cardiochek; What Happens If The Check Strip Fails? - CardioChek PA Quick Manual

Analyser
Table of Contents

Advertisement

Conducting an optics check test - CardioChek

Turn CardioChek unit on by pressing
install memo chip. Do not insert the Memo chip at this stage.
Press the "Next" button three times
Press "Enter" button
Insert the grey optics check strip with strip guides facing upwards, as shown, and press
the "Enter" button.
The unit will conduct test and display passed. The unit has
passed the optics check strip test. Remove the check strip
and store safely.
Press the "Next" button
then press the "Enter" button.
This takes you back to the main menu. Press THE "Next"
until you see the word run test. Now put your
button
memo chip in and use the machine for testing.
When inserting and removing the optics check strip and any of the test strips you must keep them as
flat as possible and

What happens if the check strip fails?

If when conducting an optics check, the unit displays failed inspect the check strip and
make sure the strip is not dirty or damaged. Clean the CardioChek optical window with a
soft, lint-free, damp cloth and allow to dry, then test again with the second grey optics
check strip supplied with your unit. Further strips can be purchased from Health Diagnostics.
These are classed as a consumable item and should be replaced every 6 months or before
depending on the level of usage.
If the unit fails again, please email qc@healthdiagnostics.co.uk
@healthdiags
www.healthdiagnostics.co.uk
and the unit will display check strip.
until you see the word exit and
IMPORTANT INFORMATION
DO NOT INSERT AT AN
ANGLE. Failure to follow this procedure may result in damage
to the plastic locators on either side of the optical window
or the next button,
and the unit will display utility.
Quick Guide
Side 1 of 1
the unit will display

Advertisement

Table of Contents
loading

Table of Contents