AP O3 Desktop Technical Data Manual

Ozone monitor, temperature and humidity modbus, analog, relay outputs

Advertisement

Quick Links

Ozone Monitor, Temperature and Humidity
Modbus, Analog, Relay Outputs
O
Desktop
3
Technical Data
Gas Detected
Sensing Element
Sensor lifetime
Temperature Sensor
Humidity Sensor
Power Supply
Power Consumption
Response Time
Signal Update
Warm up Time
Ozone Measuring Range
Display Resolution
Accuracy
Nonlinear
Repeatability
Zero Drift
Sound
Display, LCD backlight
Temperature
measurement
Humidity measurement
Analog Output
Analog Output Resolution 16Bit
Relay dry contact Output
db_ozone_wall_duct_ap_uk
O
Wall
3
Ozone - O
3
Electrochemical gas sensor
>2 years, removable and easy to
replace
NTC 5K
HS series capacitive sensor
24Vac/dc (power adaptor selectable)
2.8W
<50s @T90
1s
<60 seconds
0 to 5000ppb (0-5ppm)
(0 to 9.81mg/m
)
3
0 to 1000ppb (0 to 1 ppm)
(0.01mg/m
)
3
1ppb (0,001ppm) (0,01mg/m
3
±0.01ppm + 10% reading
<1%FS
<0.5%
<1%
When LCD backlight is red the inner
buzzer alarm will be activated
Green - ozone level < 1500 ppb,
Yellow - 1500 ppb <ozone level
< 3500 ppb,
Red - 3500ppb < ozone level (default)
-20 to 50o C (-4 to 140o F)
5 to 99% rH
0 to 10Vdc (default) or 4 to 20mA
linear output selectable
Two dry-contact outputs
Max, switching current 3A
(220Vac/30Vdc), resistance Load

Features

- Real time detection and monitoring ambiant ozone level
- Design for real time detecting and monitoring ambiance
ozone level and temperature
- Electrochemical ozone sensor with high sensitivity
- Particular LCD display with three color backlights (Green/
Yellow/Red)
- Maximum ozone measuring range:
0 to 5000ppb (0 to 9.81mg/m
- Enables reset of measurement range by end user
- 2 x On/Off dry contact outputs for two stages alarm device,
or control a ozone generator or a ventilator
- Buzzer alarm and 3-color backlight LCD indication
- Preset two stage alarm points for ozone measurement
- Provide 1 x analog output (0,2 to 10Vdc/4 to 20mA) (can be
used as transmitter)
- Modbus RS485 communication interface, 15 kV antistatic
protection, individual IP address
- Easy way for calibration and setup alarm points via the RS
485 interface
- Temperature measurement and display
- Humidity measurement and display optional
- Multiple application, wall mounting type and desktop type
- Great performance with high quality

Application

- Specially used for detecting indoor ambiance quality
- Real time measuring ozone level and analyzing:
alarm when ozone level is over proof, control ozone
)
concentrations by controlling related devices in order to reach
a certain level in some special places
- Portable type for real time measuring and analyzing
- Anti-epidemic and disease prevention, sanitary inspection,
environment protection, health protection, etc.
- O3-5000 is the best device used to detect and analyze ozone
concentration.

Ordering Codes

O3-5000
Automatikprodukter
O3
)
3
Ozone Monitor 0 - 5000ppb (0 - 5ppm), incl.
remote control, desktop bracket and power
supply.
Sep. 18
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the O3 Desktop and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for AP O3 Desktop

  • Page 1 Ozone Monitor, Temperature and Humidity Modbus, Analog, Relay Outputs Sep. 18 Features - Real time detection and monitoring ambiant ozone level - Design for real time detecting and monitoring ambiance ozone level and temperature - Electrochemical ozone sensor with high sensitivity - Particular LCD display with three color backlights (Green/ Yellow/Red) - Maximum ozone measuring range:...
  • Page 2 Ozone Monitor, Temperature and Humidity Modbus, Analog, Relay Outputs Sep. 18 Modbus Communication Ordering Codes - Alarm Units Modbus RTU protocol with Interface 19200bps(default) Warning Siren 24Vdc, 98 - 108 dB Other bps rate selectable in order AAW 230 Warning Siren 230Vac, 98 - 108 dB 15KV antistatic protection OA 24 Red Flash Light, 24Vdc...
  • Page 3 Ozone Monitor, Temperature and Humidity Modbus, Analog, Relay Outputs Sep. 18 Important Safety Information Fig. 1 Adaptor power supply - Always turn off power before mounting, removing and clearing the alarm. Notice the supply power voltage: 24Vac/dc Mounting and Wire Connection - Do not mount it behind a door, in a corner or near a heat source, diffuser or any steam source or in direct sunlight.
  • Page 4 Ozone Monitor, Temperature and Humidity Modbus, Analog, Relay Outputs Sep. 18 Parameter Setup Power off and remove the front cover. There is a set of 4 dip switches on the upper left corner on the circuit board. DIP 1 OFF - Volume concentration ON - Mass concentration Leaving factory (default): OFF DIP 2...
  • Page 5 Ozone Monitor, Temperature and Humidity Modbus, Analog, Relay Outputs Sep. 18 Modbus Register LCD Display Parameter Setting Range Default Temperature differential value ±3.0OC/±6OF Manual humidity calibration 5~99%rH Real-time ozone value 0~ upper limit of measuring calibration range as 16ppm e.g. if set this value at 16 through RS485 or remote con- troller, the sensor will take the present ozone level"...
  • Page 6 Ozone Monitor, Temperature and Humidity Modbus, Analog, Relay Outputs Sep. 18 Parameter Setting Range Default Display Sensor calibration, 1st 0ppb~ upper limit of measuring detection point range Sensor calibration, 1st AD value 0 ~ 32767 Sensor calibration, 2nd 0ppb~ upper limit of measuring 2500 detection point range...

This manual is also suitable for:

O3 wallO3 series

Table of Contents