The description of this surgical technique has been compiled by medical experts and trained staff of aap Implantate AG with utmost diligence and to the best of their knowledge. However, aap Implantate AG excludes any liability for the completeness, accuracy, currentness, and quality of the information as well as for material or immaterial damages arising from the use of this information.
Additional recesses at the underside of the proximal holes for K-wires allow the attachment of suture material • Available as right and left version • 6 and 10 holes in the plate shaft Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany • •...
Additional recesses at the underside of the proximal holes for K-wires allow the attachment of suture material • Available as right and left version with 3 holes in the plate shaft Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
All materials employed comply with national and international standards. They are characterized by good biocompatibility, a high degree of safety against allergic reactions and good mechanical properties. Indications / Contraindications Indications The aap LOQTEQ® Proximal Medial Tibia Plate 3.5 is indicated for: • Metaphyseal fractures of the medial tibial plateau •...
The proximal tibia must be clearly visible during intraoperative fluoroscopy in both planes, AP and ML. Access • Medial, according to the fracture situation. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany • •...
For this purpose, apply slight pressure onto the screw from the underside of the targeting device, and remove the screw. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
3.5 mm of appropriate length. The plate can be simultaneously pulled against the shaft using this standard cortical screw. • If necessary, readjust the plate position. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
We recommend drilling under fluoroscopic guidance to ensure optimal plate positioning and screw orientation in AP and lateral view. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
Page 13
Depth gauge for locking screws, small IS 7904-00 Handle with quick coupling, with torque limiter, 2.0 Nm IU 7707-20 Torque limiting screwdriver 2.0 Nm, quick coupling IU 7825-55 Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany • •...
After insertion of the first shaft screw or a compression screw, a clinical examination or check using fluoroscopy is recommended to assess the alignment of the tibial plateau relative to the tibial shaft. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
Therefore it is permissible in exceptionally hard bone of the diaphysis to finish the screw without the torque limiter. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
Drill to the desired depth using a drill bit ø2.7 (blue-red) and determine the depth with the depth gauge. Please refer to the LOQTEQ® Application Notes, pp. 12 and 13. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
Page 17
ø2.5. • When all required screws have been inserted, perform the final check using fluoroscopic monitoring, AP and lateral views, and close the wound. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
Place an incision on the old scar. Manually undo all screws and sequentially remove them. After manually undoing all screws, the removal can be perfor- med in automated mode in the second step. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
However, they must not interfere with the screws of the lateral plate. When combining lag or standard screws with locking screws, the lag screws must be placed first. Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
Insert the set screw (item 3) into the compression block, middle hole • Insert the retaining screws (items 4.1 and 4.2) using a hexagonal screwdriver SW 2.5 Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
Trays • LOQTEQ® Proximal Tibia Plates 3.5, Implants IC 6936-00 Small Fragment Set, Instruments IC 6931-10 Small Fragment Set, Screw Rack IC 6931-35 Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany • •...
LOQTEQ® Proximal Medial Tibia Pate 3.5, 10 holes, L 164, L, sterile PA 3552-10-2S * No Instruments! Please complete with Small Fragment Set IC 6931-05 or IC 6931-10 and IC 6931-35 Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany •...
Cortical Screw 3.5, small head, self-tapping, L 80 SK 3512-80-02 Cortical Screw 3.5, small head, self-tapping, L 85 SK 3512-85-02 Cortical Screw 3.5, small head, self-tapping, L 90 SK 3512-90-02 Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany • •...
IU 8166-20 K-wire with trocar point, ø1.6, L 150 NK 0016-15 OPTIONAL Load drill guide LOQTEQ® 3.5, adjustable up to 2 mm IU 8166-03 Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany • •...
LOQTEQ® Cortical Screw 3.5, T15, self-tapping, L 60 SK 3525-60-2 LOQTEQ® Cortical Screw 3.5, T15, self-tapping, L 65 SK 3525-65-2 LOQTEQ® Cortical Screw 3.5, T15, self-tapping, L 70 SK 3525-70-2 Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany • •...
Page 26
Cortical Screw 3.5, self-tapping, L 60 SK 3510-60-2 Cortical Screw 3.5, self-tapping, L 65 SK 3510-65-2 Cortical Screw 3.5, self-tapping, L 70 SK 3510-70-2 Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany • •...
Page 27
Cortical Screw 3.5, small head, self-tapping, L 60 SK 3512-60-2 Cortical Screw 3.5, small head, self-tapping, L 65 SK 3512-65-2 Cortical Screw 3.5, small head, self-tapping, L 70 SK 3512-70-2 Surgical Technique aap Implantate AG Proximal Medial Tibia Plate 3.5 Lorenzweg 5 12099 Berlin Germany • •...