Electronic Fever Thermometer – Express
Suitable also for measuring the basal body temperature!
Article no. 1771
Waterproof – Signalling Tone – Last Measurement Memory
This thermometer complies with the European Standard EN 12470-3.

TECHNICAL SPECIFICATIONS

Measurement Range:
Increments:
Signalling Tone:
Accuracy:
Operation Temperature:
LCD Display:
Battery:
Battery Life:
Scheiber Gmbh, D-97892 Kreuzwertheim, CE 0124
SPECIAL NOTE
Should the ambient temperature exceed +35°C, please hold the sensor tip in cold water for 30 – 60 seconds
before starting to measure.

HOW TO USE THE THERMOMETER

1. Press the on/off button; all symbols show on the LCD display for a few seconds as a means of checking that
the thermometer is functioning correctly.
2. The following display Ok indicates readiness after the internal self-test. If the display shows Err , a malfunc-
tion has occurred. Start again. If the error message is still shown after repeatedly turning the thermometer on
and off, the thermometer is broken and has to be disposed off.
3. Position the sensor tip in the preferred place of measurement – mouth, rectum or armpit (see Measuring
Methods).
4. Measuring starts with a flashing "°C" on the display. Lo is shown as long as the temperature stays below
32°C.
5. The currently measured temperature is shown as long as the "°C" is flashing.
6. As soon as the "°C" stops flashing, the maximum temperature has been measured and the signalling tone
sounds for approx. 5 seconds.
7. Should the measured maximum temperature fall below 32°C or exceed 43.9°C, either Lo- or Hi is shown on
the display.
8. To prolong the life of the battery, please switch off the thermometer after use. If the power is left on, the
thermometer will automatically shut down after 8 minutes 40 seconds to save power.
9. To safe the last measurement to the thermometer's memory simply press the on/off button for more than
1.5 seconds. The last measured temperature will be displayed with the "M" in the lower right-hand corner.
For example 37,0.
32.0°C – 43.9°C
0.1°C
approx. 5 seconds
± 0.1°C
ambient temperature +10°C – +45°C
8.0 mm, 3 ½ digits
1.5 V (LR41, AG3 or similar)
over 4,000 measurements
Normal Temperatures
Average Values:
Axillary
36.1°C
Oral
36.6°C
Rectal
37.0°C

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the B401585 and is the answer not in the manual?

Questions and answers

Summary of Contents for Kramer B401585

  • Page 1 Electronic Fever Thermometer – Express Normal Temperatures Average Values: Axillary 36.1°C Oral 36.6°C Rectal 37.0°C Suitable also for measuring the basal body temperature! Article no. 1771 Waterproof – Signalling Tone – Last Measurement Memory This thermometer complies with the European Standard EN 12470-3. TECHNICAL SPECIFICATIONS Measurement Range: 32.0°C –...
  • Page 2 Electronic Fever Thermometer – Express MEASURING METHODS 1. Armpit = Axillary approx. 15 – 25 seconds Wipe the armpit dry with a cloth. Place the sensor tip in the armpit and firmly press the thermometer in place with your arm. 2.

Table of Contents