Page 1
Instructions for use Upper arm BASIS CONTROL PLUS Automatic multiple measurement for increased measuring reliability Arrhythmia Multiple Traf c light Traf c light storage slots users display detection measurement YEARS Also suitable for: WARRANTY Diabetics...
Page 3
Basis Control Plus ® Switches, housing and accessories Display indication On/off switch Systolic value Display Diastolic value M button (Memory) Pulse indicator MAM/Multiple measurements & MAM mode (multiple measurement mode active) time selection button Battery level warning light User selection button...
Page 4
If you have any questions or problems or want to women suffering from preeclampsia, patients order spare parts, please contact the aponorm® suffering from atherosclerosis, terminal kidney customer service at the product website www. diseases, obesity and for the elderly.
Table of contents Using the monitor for the first time Safety, care, accuracy test · Inserting the batteries and disposal · Setting the date and time · Safety and protection · Selecting the correct cuff · Monitor care · Cleaning the cuff Check list for realiable measurement ·...
M - L, one the setting, the date/time setting is com- size fits all. plete. When the monitor is not in operation, U se only aponorm cuffs manufactured by ® it displays the time Microlife.
Page 7
This monitor stores the data of 2 different users. ments is currently taken. Before each measurement, ensure that the correct B etween the individual measurements, there user is selected. is an automatic pause of 15 seconds each, Select the desired user (user 1 or user 2 indicated by the MAM waiting symbol activating the user button by activating the user...
Page 8
3. Taking the blood pressure Always take your blood pressure on the same arm (normally the left one). It is 1. P ress the On/off button to start the measure- recommended that doctors perform double ment arm measurements on a patient’s first visit in order to determine which arm to measure in 2. T he cuff will now inflate automatically.
5. D uring the measurement, the “pulse rate Manual inflation measurement active” icon flashes in the In case of very high systolic blood pressure (e. g. display. exceeding 135 mmHg), it can be of use that you T he result, comprising the systolic and the determine the pressure yourself.
Viewing the stored values Range Systolic Diasto- Recommenda- tion Press the M button briefly while the monitor blood is switched off. Now the display changes over to seek medical pressure ≥ 135 ≥ 85 the data memory display, indicated by «M» advice too high (M»...
Clearing all values In the event of a battery change or power If you are sure that you want to permanently re- failure, the memory retains all values mea- move all stored values, hold down the M button sured; the date and time must, however, be (the monitor must have been switched off before- reset.
For more detailed help in troubleshooting the You can operate this monitor using the Microlife monitor, please refer to the download section mains adapter (DC 6 V, 600 mA). of the www.aponorm.de product website. Only use the aponorm mains adapter ®...
Page 13
Error Designa- Potential cause and Error Designa- Potential cause and tion remedy tion remedy «ERR 2» Interference During the measurement, «ERR 6» MAM mode Too many errors occurred signal interference signals during the measurement were detected at the in the MAM mode so cuff, caused for instance that no final result can by movement or muscle...
8. Safety, care, accuracy check · Do not use this monitor close to strong elec- and disposal tromagnetic fields such as mobile telephones or radio installations. We recommend obser- Safety nd protection ving a minimum distance of 3.3 m from such devices when using this monitor.
9. Warranty terms small that children can swallow them or choke on them. Beware of the risk of We are confident in the quality of our aponorm ® strangulations in case this monitor is fitted blood pressure monitors. Therefore we, WEPA with cables or tubes.
Page 16
Subject matter and scope of the warranty Warranty term The warranty applies to all aponorm blood pres- The warranty covers a period of 5 years from the ® sure monitors which customers in their capacity as date of purchase (warranty period). The relevant consumers purchased in Germany.
GmbH & Co. KG informs the customer of the estimated Measuring method: oszillometric, validated costs in a cost estimate before starting the repair by the Korotkoff method: work. In this case, the customer is free to commission phase I systolic, the repair of the product with costs or to request the phase V diastolic return of the unrepaired product.
Page 18
Average monitor: 5 years or life: 10.000 measurements; accessories: 2 years This monitor complies with the requirements of the Medical Device Directive 93/42/EEC. Technical alterations reserved!
Page 19
= Enhanced functionality compared with basic model aponorm Basis Control ® Important advice: In the download area of www.aponorm.de you fi nd these instructions with updates also in digital form for download, also in German, Turkish, French and Russian. Homepage Sold by:...