Drager Alcotest 7510 MBRS Instructions For Use Manual

Breath alcohol monitor

Advertisement

Quick Links

Dräger Alcotest
Breath Alcohol Monitor
i
®
7510 MBRS
Instructions for use
en
3

Advertisement

Table of Contents
loading

Summary of Contents for Drager Alcotest 7510 MBRS

  • Page 1 Instructions for use 3 ® Dräger Alcotest 7510 MBRS Breath Alcohol Monitor...
  • Page 3: Table Of Contents

    Contents Safety-related information ....4 6.1.3 Show cal. dates ......13 6.1.4 Accuracy check .
  • Page 4: Safety-Related Information

    Safety-related information Not for use in areas of explosion hazard This device is neither approved nor certified for use in areas Before using this product, carefully read these instructions  where combustible or explosive gas mixtures are likely to oc- for use and those of the associated products.
  • Page 5: Description

    Description Product overview 1 Mouthpiece receptacle 2 Display - key (UP) - key (ON/OFF, Confirm) 11:53:47 04.12.17 5 Opening to the audible alarm 6 USB connection Alcotest 7510 - key (DOWN/MENU) 8 Mouthpiece ejector at right or left dependent on the mouthpiece receptacle 9 LED red, yellow, and green Alcotest 7510 00200085.eps...
  • Page 6: Intended Use

    Intended use ®1) Dräger Alcotest 7510 MBRS Installing the wrist strap Breath alcohol measuring device for the fast determination  1. Guide the mounting strap (thin section) of the wrist strap th- of the presence of alcohol in breath. rough the housing drills on the underside of the measuring For portable, battery-powered operation.
  • Page 7: Symbols

    Symbols Battery or charge state of rechargeable battery: full According to the device status, the following symbols are dis- Battery or charge state of rechargeable battery: played in the top row, next to the time and date information full (one below the other on the right side) Battery or charge state of rechargeable battery: full 11:53:47...
  • Page 8: Switch On And Off

    Switch on and off Automatic shut down 17:42:31 08.10.2018 In the default setting, the device shuts Switch on down automatically 4 minutes after the 17:42:31 08.10.2018 Press and hold the key for  last operation. approx. 1 second until the start The device also shuts down automatical- screen appears.
  • Page 9: Performing The Measurement

    Performing the measurement Preparation Switch on the device.  17:42:31 Prerequisites After a few seconds, the start screen 08.10.2018 appears. 1. Insert a new mouthpiece for each test subject. 2. Do not allow any high alcohol concentrations on the sen- sor, e.g., by rinsing the mouth with a high proof alcohol Alcotest 7510 shortly before the measurement.
  • Page 10: Procedure

    The main user screen then appears Device then moves from 17:42:31 17:42:31 with » Driver «, » Specified » WARMUP «, » AIR BLANK 08.10.2018 08.10.2018 Driver « and » Menu « options. CHECK « to » READY «. Driver Driver READY Specified Driver...
  • Page 11: Test Result

    5.4.2 Test result Error during the delivery of the breath sample If the volume of the breath sample is in- After 3 to 20 seconds (dependent on the 17:42:31 17:42:31 sufficient, the following appears in the 08.10.2018 detected concentration) the measure- 08.10.2018 display: »...
  • Page 12: Removing The Mouthpiece

    Menu Removing the mouthpiece Push the eject button upwards.  Menu settings The mouthpiece is pushed out of the mouthpiece recep- tacle. 6.1.1 Main user screen Select » Menu «.  17:42:31 08.10.2018 Driver Specified Driver Menu 01600085.eps 01000085.eps Use blue keys to navigate and green ...
  • Page 13: Show Cal. Dates

    6.1.3 Show cal. dates Display of last date the device was calibrated. Back with green key. 6.1.4 Accuracy check 1. Select » Accuracy check « and press the key. The display shows the calibration check gas to be used. » Next « is automatically highlighted at the bottom of the screen.
  • Page 14: Troubleshooting

    Troubleshooting Error message in Cause Remedy the display Fault Cause Remedy Measured value Use lower concent- Device auto- Discharge the Replace batteries. exceeds display rations. matically battery / batteries area. switches off No activity for 4 Switch on device again. »...
  • Page 15: Maintenance

    Maintenance 3. Remove used alkaline bat- teries insert alkaline batteries – obser- – Replacing the batteries ve correct polarity! 4. Replace the battery cover For devices with alkaline batteries or NiMH rechargeable bat- and lock the catch with a teries. clockwise rotation.
  • Page 16: Cleaning And Disinfecting

    Disposal Cleaning and disinfecting Do not submerge in liquids, no liquids can touch the  This product must not be disposed of as household connectors! waste. It is therefore marked with the adjacent symbol. For cleaning: You can return this product to Dräger free of charge. For information please contact the national marketing wipe with a disposable towel.
  • Page 17: Technical Data

    Technical data CE marking Electromagnetic compatibility Measuring Principle (Directive 2014/30/EC) Electrochemical sensor Measuring range (Default setting) Breath alcohol concent- 0.00 to 3.0 mg/L ration (Ethyl alcohol mass per breath vo- lume at 34 C and 1013 hPa.) Display Area The display is identical with the measuring range. If the measurement result is outside of the measuring range, ↑...
  • Page 20 Dräger Safety AG & Co. KGaA Revalstraße 1 93 00 085 - GA 4754.602 en 23560 Lübeck, Germany +49 451 882-0 © Dräger Safety AG & Co. KGaA Fax +49 451 882-20 80 Edition 01 - February 2019 www.draeger.com Subject to alteration...