CTI-CRYOGENICS Cryo-Torr Series Installation And Operation Instructions

Temperature indicator

Advertisement

Quick Links

8040024 Rev. A (6/94)
Cryo-Torr
Installation and Operation Instructions
The Cryo-Torr Vacuum-Pump Temperature Indicator, Part No. 8042001G003, or
G004 shown in Figure 1, is used with Cryo-Torr high-vacuum pumps, to monitor
cryopump operating temperatures. With User-supplied switch gear, the indicator
can be used in multiple installations. In cryopumps provided with a silicon diode
sensor, the indicator is easily connected to the cryopump with an electrical
cable. The panel readout is in Kelvin, and has a range from 10K to 320K. The
indicator has two set points that are adjustable over this range. When the vacuum
pump temperature reaches either setpoint, electrical contacts on the rear
terminal board can be used to activate indicators or operate solenoid valves for
controlling temperature-related functions. A switch on the front panel permits
momentary display of the two setpoints.
Copyright © 1994, Helix Technology Corporation
Mansfield Corporate Center, Nine Hampshire Street, Mansfield, Massachusetts 02048-9171 Telephone (508) 337-5000
®
Temperature Indicator
Figure 1: Cryo-Torr Temperature Indicator
HELIX TECHNOLOGY CORPORATION
Printed in U.S.A
-
CR Y OGENICS
C
TI
HELIX TECHNOLOGY CORPORATION
DEO-10302
1

Advertisement

loading
Need help?

Need help?

Do you have a question about the Cryo-Torr Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for CTI-CRYOGENICS Cryo-Torr Series

  • Page 1 8040024 Rev. A (6/94) CR Y OGENICS HELIX TECHNOLOGY CORPORATION ® Cryo-Torr Temperature Indicator Installation and Operation Instructions The Cryo-Torr Vacuum-Pump Temperature Indicator, Part No. 8042001G003, or G004 shown in Figure 1, is used with Cryo-Torr high-vacuum pumps, to monitor cryopump operating temperatures.
  • Page 2: Specifications

    1. Connect the cryopump to the temperature indicator SENSOR connector using the 10 - foot interconnecting cable provided with the indicator. Contact CTI-CRYOGENICS if the supplied cable is too short to order a longer cable. 2. Insert input power cable plug in appropriate power receptacle. The unit is furnished with either 115V or 208/230V three prong as specified in the sales order.
  • Page 3: Operation

    8040024 Rev. A (6/94) Operation 1. Connect the contacts of the HI and LO relays, on the two terminal strips on the rear panel, either to indicators or to solenoid valves, as desired. 2. Turn the rocker power switch on the left side of the front panel to ON. The panel meter will read the temperature in Kelvin.
  • Page 4 8040024 Rev. A (6/94) Table 2: Troubleshooting Fault Possible Cause Corrective Action Panel meter fails to Power switch is off. Turn switch on. indicate a reading. Power cord not plugged in. Plug in power cord Fuse blown on rear panel Replace the fuse.
  • Page 5 8040024 Rev. A (6/94) Table 2: Troubleshooting (Continued) Fault Possible Cause Corrective Action Check to see if the proper sig- Solenoid valves or nal is being provided by the indicator are not being electronics. Repair as operated at proper Defective relays or electronics. required by a qualified temperature in accor- technician.