Summary of Contents for Quart Healthcare Diamond 8 Plus
Page 1
User Manual – Diamond 8 Plus Diamond 8 Plus Alternating + Micro Low Air Loss Pressure Relief System User Manual Manufactured by: Moxi Enterprises, LLC. Distributed by: Quart Healthcare Inc. www.quarthealthcare.com...
Page 2
User Manual – Diamond 8 Plus Warning Connect the Master Control unit to a proper power source Do not use the system in the presence of any flammable gases (such as Anesthetic Agents) Keep the pump and mattress away from sources of liquid and open flames ...
User Manual – Diamond 8 Plus Intended Use The Diamond 8 Plus system is a unique and innovative specialized mattress replacement unit. It features a patented 8-shape twin cell design; whereby, the bottom half of the cell is always inflated and the top half is free to alternate. The system utilizes micro low air loss technology with a low rate of flow.
Page 4
User Manual – Diamond 8 Plus Master Control Unit Part No. FC-QHI0002 Device Identifier (Health Canada). FM-MOX0003S Size (cm) 34.5(L)x12.5 (W)x22.5(H) Weight (Kg) Cycle Time (min) 20 min 20 min Min/Max Pressure 20 ~ 60 mmHg +/- 5 mmHg Max Flow-rate...
User Manual – Diamond 8 Plus Symbols Used Class II Equipment Refer to Accompanying Documents Type BF Applied Part Waste Disposal Warning 1. Instructions for Proper Use Remove the existing mattress from the bed frame Replace the standard mattress with mattress replacement system (orient mattress so that the air tube is at the foot of the bed) iii.
Page 6
User Manual – Diamond 8 Plus Pressing MAX button will activate quick inflation to bring the mattress to a flat surface for performing nursing tasks with maximum pressure. The pump is set to After the mattress is fully default back to “ALTER” after 30 minutes of inflation.
Page 7
User Manual – Diamond 8 Plus 2. Cleaning The mattress should be cleaned on the bed weekly using a damp soft cloth and mild Mattress Overlay detergent. If top sheet (Top cover) or base (Bottom cover) becomes grossly soiled, put on clean gloves, plastic gown and eye protection before removing top sheet or base and disposing according to standard hospital procedures for contaminated waste.
Page 8
User Manual – Diamond 8 Plus Master Control Unit SWITCH OFF THE ELECTRICAL SUPPLY TO THE PUMP AND DISCONNECT THE POWER CORD FROM THE MAIN SUPPLY BEFORE CLEANING AND INSPECTION The casing of the pump is manufactured from ABS plastic. If soiled it can be wiped down with a sodium hypochlorite solution to dilution of 1000ppm or any EPA approved hospital grade disinfectant.
User Manual – Diamond 8 Plus 3. Storage and Care Control Unit: Check the power cord and plug for abrasions and excessive wear Plug in the unit and verify air flow from the hose connection ports Place in plastic bag for storage Mattress Replacement System: ...
In no event shall Quart Healthcare Products be liable for any direct, indirect or consequential damages or losses resulting from the use of equipment...