Page 3
Be aware of the risks of using the required reagents by reading the MSDS (Material Safety Data Sheets). Failure could result in serious injury to the operator or damage to the instrument. MSDS: www.orbeco.com SP600 Spectrophotometer_Revision 1 01/2008 SP600 Spectrophotometer 01/2008...
7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted prepare Test 8. Press TEST key. press TEST The result is shown in the display as Acid demand to pH 4.3 in mmol/l. SP600 Spectrophotometer 01/2008...
Page 17
1.1 Methods Notes: 1. The terms total Alkalinity, Alkalinity-m, m-Value and Acid demand to pH 4.3 are identical. 2. For accurate results exactly 10 ml of water sample must be taken for the test. SP600 Spectrophotometer 01/2008...
7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted prepare Test 8. Press TEST key. press TEST The result is shown in the display as Alkalinity-p. SP600 Spectrophotometer 01/2008...
Page 19
Case 2: Alkalinity-p > 0 and Alkalinity-m > 2p Hydrogen carbonate = m – 2p Carbonate = 2p Hydroxide = 0 Case 3: Alkalinity-p > 0 and Alkalinity-m < 2p Hydrogen carbonate = 0 Carbonate = 2m – 2p Hydroxide = 2p – m SP600 Spectrophotometer 01/2008...
7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted prepare Test 8. Press TEST key. press TEST The result is shown in the display as total Alkalinity. SP600 Spectrophotometer 01/2008...
Test 9. Press TEST key. press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Aluminium. SP600 Spectrophotometer 01/2008...
Page 23
0.45 0.09 0.20 0.29 0.37 0.48 Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is 0.17 mg/l Al. SP600 Spectrophotometer 01/2008...
(this is the sample). 10. Close the vials tightly and swirl several times to mix the contents. Countdown 2 11. Press [ ] key. 5:00 Wait for a reaction period of 5 minutes. start: SP600 Spectrophotometer 01/2008...
Page 25
0.45 0.09 0.20 0.29 0.37 0.48 Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is 0.17 mg/l Al. SP600 Spectrophotometer 01/2008...
Test 9. Press TEST key. press TEST Wait for a reaction period of 10 minutes. Countdown 10:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Ammonium. SP600 Spectrophotometer 01/2008...
Page 27
Fill the test tube with the sample to the 10 ml mark and add one level spoonful of Conditioning Powder. Mix to dissolve, then continue as described in the test instructions. 5. Conversion: mg/l NH = mg/l N x 1.29 mg/l NH = mg/l N x 1.22 SP600 Spectrophotometer 01/2008...
12. Place the vial (the sample) in the sample chamber making sure that the marks are aligned. Zero accepted prepare Test 13. Press TEST key. press TEST The result is shown in the display in mg/l Ammonium. SP600 Spectrophotometer 01/2008...
Page 29
Glycine, Hydrazine, Less common interferences such as Hydrazine and Colour, Turbidity Glycine will cause intensified colours in the prepared sample. Turbidity and colour will give erroneous high values. Samples with severe interferences require distillation. SP600 Spectrophotometer 01/2008...
10. Place the vial (the sample) in the sample chamber making sure that the marks are aligned. Zero accepted 11. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Ammonium. SP600 Spectrophotometer 01/2008...
Page 31
Determine the amount of total iron present in the water sample. To produce the blank add an iron standard solution with the same iron concentration to the vial (point 1) instead of deionized water 4. Conversion: mg/l NH = mg/l N x 1.29 mg/l NH = mg/l N x 1.22 SP600 Spectrophotometer 01/2008...
10. Place the vial (the sample) in the sample chamber making sure that the marks are aligned. Zero accepted prepare Test 11. Press TEST key. press TEST The result is shown in the display in mg/l Ammonium. SP600 Spectrophotometer 01/2008...
Page 33
Determine the amount of total iron present in the water sample. Add an iron standard solution with the same concentration to the vial (point 1) instead of deionized water to produce the blank. 4. Conversion: mg/l NH = mg/l N x 1.29 mg/l NH = mg/l N x 1.22 SP600 Spectrophotometer 01/2008...
1 hour and filtrate to get a clear solution. Notes: • use only dry glass vessels • stored in a dark glass bottle at max. 20°C the absorp- tion solution can be used for about 1 week • store Silver diethyldithiocarbamate at 4°C. Part list for glass apparatus: • 100 ml Erlenmeyer flask (NS29/32) Order code: 37 05 01 • glass stopper (NS29/32) Order code: 37 05 02 • absorption tube (NS29,2/32) Order code: 37 05 03 Assembling of apparatus: SP600 Spectrophotometer 01/2008...
Page 35
13. Fill the cell with the colored absorption solution. (see picture 14. Place the cell in the sample chamber making sure that the marks are aligned. Zero accepted 15. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Arsenic. SP600 Spectrophotometer 01/2008...
Page 36
Positioning: insert cell on the left side in the sample chamber (c = clip). 4. According to literature (G. Ackermann, J. Köthe: Fresenius Z. Anal. Chem. 323 (1986), 135) Sb, Se and Te interfere due to the same reaction; Thiosulfate interferes differently. SP600 Spectrophotometer 01/2008...
Test 9. Press TEST key. press TEST Countdown Wait for a reaction period of 20 minutes. 20:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Boron. SP600 Spectrophotometer 01/2008...
Page 39
2. The sample solution should have a pH value between 6 and 7. 3. Interferences are prevented by the presence of EDTA in the tablets. 4. The rate of colour development depends on the temperature. The temperature of the sample must be 20°C ± 1°C. SP600 Spectrophotometer 01/2008...
8. Fill the 10 mm cell with the colored test solution. 9. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted 10. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Bromine. SP600 Spectrophotometer 01/2008...
Page 41
0 mg/l. In this event, the water must be diluted with water free of Bromine. 10 ml of the diluted sample will be mixed with the reagent and the measurement repeated. Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Bromine. SP600 Spectrophotometer 01/2008...
Page 42
8. Fill the 50 mm cell with the colored test solution. 9. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted 10. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Bromine. SP600 Spectrophotometer 01/2008...
Page 43
0 mg/l. In this event, the water must be diluted with water free of Bromine. 10 ml of the diluted sample will be mixed with the reagent and the measurement repeated. Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Bromine. SP600 Spectrophotometer 01/2008...
Page 44
8. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted 9. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Bromine. SP600 Spectrophotometer 01/2008...
Page 45
0 mg/l. In this event, the water sample must be diluted with water free of Bromine. 10 ml of the diluted sample will be mixed with the reagent and the measurement repeated. Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Bromine. SP600 Spectrophotometer 01/2008...
Test 11. Press TEST key. press TEST Countdown Wait for a reaction period of 2 minutes. 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display as Cadmium. SP600 Spectrophotometer 01/2008...
Page 47
6. Sample and reagent volumes should always metered by using volumetric pipettes (class A). 7. This test determines only Cd -ions. Samples must be pre-treated or decomposed by digestion before colloidal, undissolved and complex-bounded cadmium can be mea- sured. mg/l μg/l SP600 Spectrophotometer 01/2008...
Test 9. Press TEST key. press TEST Wait for a reaction period of 2 minutes. Countdown After the reaction period is finished the measurement starts 2:00 automatically. The result is shown in the display in mg/l Chloride. SP600 Spectrophotometer 01/2008...
Page 49
3. Ions which also form deposits with Silver nitrate in acidic media, such as Bromides, Iodides and Thiocyanates, interfere with the analysis. 4. Highly alkaline water should – if necessary – be neutralised using Nitric acid before analysis. SP600 Spectrophotometer 01/2008...
11. Place the vial (the sample) in the sample chamber ma- king sure that the marks are aligned. Zero accepted 12. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Chloride. SP600 Spectrophotometer 01/2008...
Page 51
2. The test sample should have a pH of between 3 and 9. 3. Store the reagent bottles in a cool, dry place ideally at between 4°C and 8°C. 4. Interferences: Thiocyanate, Sulfide, Thiosulfate, Bromide and Iodide interfere because they react like Chloride. SP600 Spectrophotometer 01/2008...
Chlorine >> free for the determination of free Chlorine for the determination of total Chlorine >> total Select the desired determination with the arrow keys [ ] and [ ]. Confirm with [ ] key. SP600 Spectrophotometer 01/2008...
Page 53
* it is not possible to give exactly values, because the development of turbidity depends on nature and ingredients of the sample. 7. If ??? is displayed at a differenciated test result see page 325. Oxidizing agents such as Bromine, Ozone etc. interfere as they react like Chlorine. SP600 Spectrophotometer 01/2008...
Page 54
After the reaction period is finished the measurement 2:00 starts automatically. The result is shown in the display in: *,** mg/l free Cl mg/l free Chlorine *,** mg/l comb Cl mg/l combined Chlorine *,** mg/l tot Cl mg/l total Chlorine SP600 Spectrophotometer 01/2008...
Page 55
9. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted prepare Test 10. Press TEST key. press TEST The result is shown in the display in mg/l free Chlorine. Notes: see page 51 SP600 Spectrophotometer 01/2008...
Page 56
TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l total Chlorine. Notes: see page 51 SP600 Spectrophotometer 01/2008...
After the reaction period is finished the measurement 2:00 starts automatically. The result is shown in the display in: *,** mg/l free Cl mg/l free Chlorine *,** mg/l comb Cl mg/l combined Chlorine *,** mg/l tot Cl mg/l total Chlorine SP600 Spectrophotometer 01/2008...
9. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted prepare Test 10. Press TEST key. press TEST The result is shown in the display in mg/l free Chlorine. Notes: see page 51 SP600 Spectrophotometer 01/2008...
Page 59
10. Press TEST key. press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l total Chlorine. SP600 Spectrophotometer 01/2008...
Page 60
11. Add one DPD No. 3 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod. 12. Close the vial tightly with the cap and swirl several times until the tablet is dissolved. SP600 Spectrophotometer 01/2008...
Page 61
After the reaction period is finished the measurement starts automatically. The result is shown in the display in: *,** mg/l free Cl mg/l free Chlorine *,** mg/l comb Cl mg/l combined Chlorine *,** mg/l total Cl mg/l total Chlorine Notes: see page 51 SP600 Spectrophotometer 01/2008...
Page 62
8. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted prepare Test 9. Press TEST key. press TEST The result is shown in the display in mg/l free Chlorine. Notes: see page 51 SP600 Spectrophotometer 01/2008...
Page 63
TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l total Chlorine. Notes: see page 51 SP600 Spectrophotometer 01/2008...
10. Remove the vial from the sample chamber. 11. Add 3 drops of DPD 3 solution to the same water sample. 12. Close the vial tightly with the cap and swirl several times to mix the contents. SP600 Spectrophotometer 01/2008...
Page 65
*,** mg/l total Cl mg/l total Chlorine Notes: 1. After use replace the bottle caps securely noting the colour coding. 2. Store the reagent bottles in a cool, dry place ideally between 6°C and 10°C. 3. Also see page 51. SP600 Spectrophotometer 01/2008...
Notes (free and total Chlorine): 1. After use replace the bottle caps securely noting the colour coding. 2. Store the reagent bottles in a cool, dry place ideally between 6°C and 10°C. 3. Also see page 51. SP600 Spectrophotometer 01/2008...
Zero accepted prepare Test Wait for a reaction period of 2 minutes. press TEST Countdown After the reaction period is finished the measurement starts 2:00 automatically. The result is shown in the display in mg/l total Chlorine. SP600 Spectrophotometer 01/2008...
10. Add one VARIO Chlorine TOTAL-DPD / F10 powder pack straight from the foil to the water sample. 11. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds). SP600 Spectrophotometer 01/2008...
Page 69
After the reaction period is finished the measurement starts automatically. The result is shown in the display in: *,** mg/l free Cl mg/l free Chlorine *,** mg/l comb. Cl mg/l combined Chlorine *,** mg/l total Cl mg/l total Chlorine Notes: see page 51 SP600 Spectrophotometer 01/2008...
7. Place the vial in the sample chamber making sure that marks are aligned. 8. Press TEST key. Zero accepted prepare Test press TEST The result is shown in the display in mg/l free Chlorine. Notes: see page 51 SP600 Spectrophotometer 01/2008...
Wait for a reaction period of 3 minutes. press TEST Countdown After the reaction period is finished the measurement starts 3:00 automatically. The result is shown in the display in mg/l total Chlorine. Notes: see page 51 SP600 Spectrophotometer 01/2008...
8. Place the vial in the sample chamber making sure that the marks are aligned. Zero accepted 9. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Chlorine. SP600 Spectrophotometer 01/2008...
Page 73
1.1 Methods Notes: 1. Oxidizing agents interfere as they react like Chlorine. SP600 Spectrophotometer 01/2008...
8. Fill the 50 mm cell with the colored test solution. 9. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted prepare Test 10. Press TEST key. press TEST The result is shown in the display in mg/l Chlorine dioxide. SP600 Spectrophotometer 01/2008...
Page 75
0 mg/l. In this event, the water sample must be diluted with water free of Chlorine dioxide. 10 ml of the diluted sample will be mixed with the reagent and the measurement repeated. 6. Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Chlorine dioxide. SP600 Spectrophotometer 01/2008...
>> with Cl of Chlorine. for the determination of Chlorine dioxide in the absence >> without Cl of Chlorine. Select the desired determination with the arrow keys [ ] and [ ]. Confirm with [ ] key. SP600 Spectrophotometer 01/2008...
Page 77
Chlorine dioxide. 10 ml of the diluted sample will be mixed with the reagent and the measurement repeated. 5. If ??? is displayed at a differentiated test result see page 325. Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Chlorine dioxide. SP600 Spectrophotometer 01/2008...
10. Close the vial tightly with the cap and swirl several times until the tablet is dissolved. 11. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted 12. Press TEST key. prepare T1 press TEST SP600 Spectrophotometer 01/2008...
Page 79
Chlorine dioxide in mg/l Chlorine, *,** mg/l ClO as Chlorine dioxide in mg/l ClO mg/l free Chlorine *,** mg/l free Cl mg/l combined Chlorine *,** mg/l comb. Cl mg/l total Chlorine *,** mg/l total Cl Notes: See next page. SP600 Spectrophotometer 01/2008...
Page 80
2. The total Chlorine result given includes the contribution by the Chlorine dioxide (as Chlorine) reading. For true total Chlorine value subtract the Chlorine dioxide (as Chlorine) reading from the quoted total Chlorine reading. 3. Also see page 75. SP600 Spectrophotometer 01/2008...
Test 9. Press TEST key. press TEST The result is shown in the display *,** mg/l ClO [Cl] as Chlorine dioxide in mg/l Chlorine, *,** mg/l ClO as Chlorine dioxide in mg/l ClO Notes: see page 75 SP600 Spectrophotometer 01/2008...
Chromium (VI) for the determination of total Chromium (sum Cr (III) + >> Cr (III + VI) Cr (VI)) Select the desired determination with the arrow keys [ ] and [ ]. Confirm with the [ ] key. SP600 Spectrophotometer 01/2008...
13. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted prepare T1 14. Press TEST key. press TEST Wait for a reaction period of 5 minutes. Countdown After the reaction period is finished the measurement 5:00 starts automatically. SP600 Spectrophotometer 01/2008...
Page 85
3. For information about interferences especially in waste water and chemical waste water through metals and reductive or oxidic agents see DIN 38 405 – D 24 and Standard Methods of Water and Wastewater, 20th Edition; 1998. mg/l μg/l SP600 Spectrophotometer 01/2008...
Page 86
Test press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display as Chromium (Vl). Notes: see previous page SP600 Spectrophotometer 01/2008...
Test 14. Press TEST key. press TEST Wait for a reaction period of 5 minutes. Countdown After the reaction period is finished the measurement 5:00 starts automatically. The result is shown in the display as total Chromium. SP600 Spectrophotometer 01/2008...
Page 88
11. Place the vial in the sample chamber making sure that the marks are aligned. Zero accepted prepare T1 12. Press TEST key. press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. SP600 Spectrophotometer 01/2008...
Page 89
3. For information about interferences especially in waste water and chemical waste water through metals and reductive or oxidic agents see DIN 38 405 – D 24 and Standard Methods of Water and Wastewater, 20th Edition; 1998. SP600 Spectrophotometer 01/2008...
Page 90
TEST Countdown Wait for a reaction period of 5 minutes. 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Chromium (Vl). Notes: see previous page SP600 Spectrophotometer 01/2008...
12. Press TEST key. press TEST Wait for a reaction period of 5 minutes. Countdown After the reaction period is finished the measurement 5:00 starts automatically. The result is shown in the display in mg/l total Chromium. SP600 Spectrophotometer 01/2008...
9. Place the vial (the sample (Note 3, 4)) in the sample chamber making sure that the marks are aligned. Zero accepted prepare Test 10. Press TEST key. press TEST The result is shown in the display in mg/l COD. SP600 Spectrophotometer 01/2008...
Page 93
5. Samples can be measured when the Chloride contents does not exceed 1000 mg/l. 6. In exceptional cases, compounds contained in the water cannot be oxidized adequate, what results in minimum findings, compared with the reference method. SP600 Spectrophotometer 01/2008...
9. Place the vial (the sample (Note 3, 4)) in the sample chamber making sure that the marks are aligned. Zero accepted 10. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l COD. SP600 Spectrophotometer 01/2008...
Page 95
6. In exceptional cases, compounds contained in the water cannot be oxidized adequate, what results in minimum findings, compared with the reference method. 7. For samples under 100 mg/l COD it is recommendable to repeat the test with the tube test for COD LR. SP600 Spectrophotometer 01/2008...
9. Place the vial (the sample (Note 3, 4)) in the sample chamber making sure that the marks are aligned. Zero accepted 10. Press TEST key. prepare Test press TEST The result is shown in the display in g/l COD. SP600 Spectrophotometer 01/2008...
Page 97
7. For samples under 1 g/l COD it is recommendable to repeat the test with the test kit for COD MR or for samples under 0.1 g/l COD with the tube test COD LR. SP600 Spectrophotometer 01/2008...
Copper. >> free for the determination of free Copper. for the determination of total Copper. >> total Select the desired determination with the arrow keys [ ] and [ ]. Confirm with [ ] key. SP600 Spectrophotometer 01/2008...
Page 99
1.1 Methods Note: 1. If ??? is displayed at the diffentiated test result see page 325. SP600 Spectrophotometer 01/2008...
12. Add one COPPER No. 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod. Dissolve the tablet. SP600 Spectrophotometer 01/2008...
Page 101
14. Press TEST key. prepare T2 press TEST The result is shown in the display in: *,** mg/l free Cu mg/l free Copper *,** mg/l comb Cu mg/l combined Copper *,** mg/l tot Cu mg/l total Copper SP600 Spectrophotometer 01/2008...
Page 102
7. Fill the 50 mm cell with the colored test solution. 8. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted prepare Test 9. Press TEST key. press TEST The result is shown in the display in mg/l Copper. SP600 Spectrophotometer 01/2008...
Page 103
7. Fill the 50 mm cell with the colored test solution. 8. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted 9. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l total Copper. SP600 Spectrophotometer 01/2008...
Page 104
10. Add one COPPER No. 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod. 11. Close the vial tightly with the cap and swirl several times until the tablet is dissolved. SP600 Spectrophotometer 01/2008...
Page 105
13. Press TEST key. prepare T2 press TEST The result is shown in the display in: *,** mg/l free Cu mg/l free Copper *,** mg/l comb Cu mg/l combined Copper *,** mg/l total Cu mg/l total Copper SP600 Spectrophotometer 01/2008...
Page 106
7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted prepare Test 8. Press TEST key. press TEST The result is shown in the display in mg/l free Copper. SP600 Spectrophotometer 01/2008...
Page 107
7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted 8. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l total Copper. SP600 Spectrophotometer 01/2008...
8. Press TEST key. prepare Test press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Copper. SP600 Spectrophotometer 01/2008...
Page 109
If a turbidity remains and turns black, silver interferences is likely. Add 10 drops of saturated Potassium chloride solution to 75 ml of water sample. Filtrate through a fine filter. Use 10 ml of the filtered water sample to perform test. SP600 Spectrophotometer 01/2008...
Test 11. Press TEST key. press TEST Wait for a reaction period of 10 minutes. Countdown 10:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display as Cyanide. SP600 Spectrophotometer 01/2008...
Page 111
2. In the present of Thiocyanate, heavy metal complexes, colorants or aromatic amines, the cyanide must be separated out by distillation before analysis is performed. 3. Store the reagents in closed containers at a temperature of + 15°C to + 25°C. mg/l μg/l SP600 Spectrophotometer 01/2008...
Page 112
Test 10. Press TEST key. press TEST Wait for a reaction period of 10 minutes. Countdown After the reaction period is finished the measurement 10:00 starts automatically. The result is shown in the display in mg/l Cyanide. SP600 Spectrophotometer 01/2008...
Page 113
2. In the present of Thiocyanate, heavy metal complexes, colorants or aromatic amines, the cyanide must be separated out by distillation before analysis is performed. 3. Store the reagents in closed containers at a temperature of + 15°C to + 25°C. SP600 Spectrophotometer 01/2008...
(Note 2, 3). 7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted 8. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Cyanuric acid. SP600 Spectrophotometer 01/2008...
Page 115
2. If Cyanuric acid is present a cloudy solution will be given. Single particles are uncaused necessarily by Cyanuric acid. 3. Dissolve the tablet completely (therefore swirl the vial approx. 1 minute). Not dissolved particles of the tablet can cause too high results. SP600 Spectrophotometer 01/2008...
Test 10. Press TEST key. press TEST Wait for a reaction period of 10 minutes. Countdown After the reaction period is finished the measurement 10:00 starts automatically. The result is shown in the display as DEHA. SP600 Spectrophotometer 01/2008...
Page 117
10 mg/l Sulfate 1000 mg/l Zinc 50 mg/l 6. There is an option to change the unit from mg/l to μg/l. The unit mg/l is rounded, e.g.: 25 μg/l = 0.025 mg/l display 0.03 mg/l. mg/l μg/l SP600 Spectrophotometer 01/2008...
11. Place the vial (the sample) in the sample chamber ma- king sure that the marks are aligned. Zero accepted prepare Test 12. Press TEST key. press TEST The result is shown in the display as DEHA. SP600 Spectrophotometer 01/2008...
Page 119
10 mg/l Sulfate 1000 mg/l Zinc 50 mg/l 7. There is an option to change the unit from mg/l to μg/l. The unit mg/l is rounded, e.g.: 25 μg/l = 0.025 mg/l display 0.03 mg/l. mg/l μg/l SP600 Spectrophotometer 01/2008...
7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted prepare Test 8. Press TEST key. press TEST The result is shown in the display in mg/l Fluoride. SP600 Spectrophotometer 01/2008...
Page 121
2. 6. SPADNS reagent solution contains Arsenite. Chlorine concentrations up to 5 mg/l do not interfere. 7. Seawater and wastewater samples must be distilled. 8. It is convenient to use special vials with larger volume. SP600 Spectrophotometer 01/2008...
12. Fill the 10 mm cell with the test solution. 13. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted 14. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Form- aldehyde. SP600 Spectrophotometer 01/2008...
Page 123
4. Appropriate safety precautions and good lab technique should be used during the whole procedure. 5. Because reaction depends on temperature, sample and tube temperature must be between 20 and 25°C. 6. Sample volume should always metered by using volumetric pipette (class A). SP600 Spectrophotometer 01/2008...
Page 124
12. Fill the 50 mm cell with the test solution. 13. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted prepare Test 14. Press TEST key. press TEST The result is shown in the display in mg/l Form- aldehyde. SP600 Spectrophotometer 01/2008...
Page 125
4. Appropriate safety precautions and good lab technique should be used during the whole procedure. 5. Because reaction depends on temperature, sample and tube temperature must be between 20 and 25°C. 6. Sample volume should always metered by using volumetric pipette (class A). SP600 Spectrophotometer 01/2008...
Page 126
After reaction period is finished proceed as follows: 9. Place the vial in the sample chamber making sure that the marks are aligned. Zero accepted 10. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Form- aldehyde. SP600 Spectrophotometer 01/2008...
Page 127
4. Appropriate safety precautions and good lab technique should be used during the whole procedure. 5. Because reaction depends on temperature, sample and tube temperature must be between 20 and 25°C. 6. Sample volume should always metered by using volumetric pipette (class A). SP600 Spectrophotometer 01/2008...
8. Press TEST key. prepare Test press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display as total Hardness. SP600 Spectrophotometer 01/2008...
8. Press TEST key. prepare Test press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display as total Hardness. SP600 Spectrophotometer 01/2008...
7. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted 8. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Pt-Co-units. SP600 Spectrophotometer 01/2008...
Page 133
Do not stir the sample; avoid lengthy contact with the air. The sample may be stored in a dark place at a temperature of 4°C for 24 hours. Before performing measurements, the water sample must be brought up to room temperature. SP600 Spectrophotometer 01/2008...
(Note 4). 9. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted prepare Test 10. Press TEST key. press TEST The result is shown in the display as Hydrazine. SP600 Spectrophotometer 01/2008...
Page 135
0.05 mg/l, you should only use the reagent with reservations (major result deviation). 6. There is an option to change the unit from mg/l to μg/l. The unit mg/l is rounded, e.g.: 25 μg/l = 0.025 mg/l display 0.03 mg/l. mg/l μg/l SP600 Spectrophotometer 01/2008...
11. Press TEST key. prepare Test Wait for a reaction period of 12 minutes. press TEST Countdown After the reaction period is finished the measurement 12:00 starts automatically. The result is shown in the display as Hydrazine. SP600 Spectrophotometer 01/2008...
Page 137
Principle: Hydrazine is oxidised by the household bleach. The interference by colour will eliminated by zeroing. 5. There is an option to change the unit from mg/l to μg/l. The unit mg/l is rounded, e.g.: 25 μg/l = 0.025 mg/l display 0.03 mg/l. mg/l μg/l SP600 Spectrophotometer 01/2008...
10. Press TEST key. press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Hydrogen peroxide. SP600 Spectrophotometer 01/2008...
Page 139
0 mg/l. In this event, the water must be diluted with water free of Bromine. 10 ml of the diluted sample will be mixed with the reagent and the measurement repeated. Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Hydrogen peroxide. SP600 Spectrophotometer 01/2008...
Page 140
9. Press TEST key. prepare Test press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Hydrogen peroxide. SP600 Spectrophotometer 01/2008...
Page 141
0 mg/l. In this event, the water sample must be diluted with water free of Hydrogen peroxide. 10 ml of the diluted sample will be mixed with the reagent and the measurement repeated. Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Hydrogen peroxide. SP600 Spectrophotometer 01/2008...
8. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted prepare Test 9. Press TEST key. press TEST The result is shown in the display in mg/l Iodine. SP600 Spectrophotometer 01/2008...
Page 143
1.1 Methods Notes: 1. Oxidising reagents, such as Chlorine, Bromine, etc. interfere as they react like Iodine. SP600 Spectrophotometer 01/2008...
*Determination of all soluble iron and most insoluble forms of iron; most insoluble iron oxides are recovered by the reagent. * This information refer to analysis of the water sample without digestion. Further information you will find at the method notes. SP600 Spectrophotometer 01/2008...
Page 145
Therefore add 1 ml concentrated sulfuric acid and 1 ml concentrated nitric acid to 100 ml of water sample and boil to approx. half volume. After cooling down proceed as described above. SP600 Spectrophotometer 01/2008...
Page 146
9. Press TEST key. prepare Test Wait for a reaction period of 5 minutes. press TEST Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Iron. SP600 Spectrophotometer 01/2008...
Page 147
IRON (II) LR tablet is used, as described above, instead of the IRON LR tablet. 3. For the determination of total dissolved and undissolved iron digestion is required. An example is described on page 143. SP600 Spectrophotometer 01/2008...
Page 148
9. Press TEST key. prepare Test press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Iron. SP600 Spectrophotometer 01/2008...
Page 149
IRON (II) LR tablet is used, as described above, instead of the IRON LR tablet. 3. For the determination of total dissolved and undissolved iron digestion is required. An example is described on page 143. SP600 Spectrophotometer 01/2008...
Page 150
Test 8. Press TEST key. press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Iron. SP600 Spectrophotometer 01/2008...
Page 151
IRON (II) LR tablet is used, as described above, instead of the IRON LR tablet. 3. For the determination of total dissolved and undissolved iron digestion is required. An example is described on page 143. SP600 Spectrophotometer 01/2008...
8. Press TEST key. press TEST Wait for a reaction period of 3 minutes (Note 5). Countdown 3:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Iron. SP600 Spectrophotometer 01/2008...
Page 153
3. Very strong alkaline or acidic water samples must be adjusted to a ph-Value between 3 and 5 before analysis. 4. Accuracy is not affected by undissolved powder. 5. Water samples containing visible rust should be allowed to react at least five minutes. SP600 Spectrophotometer 01/2008...
9. Place the vial (the sample) in the sample chamber ma- king sure that the marks are aligned. Zero accepted prepare Test 10. Press TEST key. press TEST The result is shown in the display in mg/l Iron. SP600 Spectrophotometer 01/2008...
8. Fill the 10 mm cell with the colored test solution. 9. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted prepare Test 10. Press TEST key. press TEST The result is shown in the display in mg/l Lead. SP600 Spectrophotometer 01/2008...
Page 157
5. Volumes for sample and reagent should always metered by using volumetric pipettes (class A). 6. The test determines only Pb -ions. Samples must be pre-treated or decomposed by digestion before colloidal, undissolved and complex-bounded lead can be measured. SP600 Spectrophotometer 01/2008...
70 mg/l (approx. 10°d). Procedure B Select this test for the determination of lead in hard to very hard waters with a Ca contents between 70 mg/l and 500 mg/l (approx. 10°d to 70°d). SP600 Spectrophotometer 01/2008...
Page 159
6. Sample volume should always metered by using volumetric pipette (class A). 7. The test determines only Pb -ions. Samples must be pre-treated or decomposed by digestion before colloidal, undissolved and complex-bounded lead can be measured. Test performance: see next page SP600 Spectrophotometer 01/2008...
Page 160
8. Place the vial in the sample chamber making sure that the marks are aligned. Zero accepted 9. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Lead. Notes: see previous page SP600 Spectrophotometer 01/2008...
Page 161
13. Place the vial in the sample chamber making sure that the marks are aligned. T1 accepted prepare T2 14. Press TEST key. press TEST The result is shown in the display in mg/l Lead. SP600 Spectrophotometer 01/2008...
Test 9. Press TEST key. press TEST Wait for a reaction period of 5 minutes. Countdown After the reaction period is finished the measurement starts 5:00 automatically. The result is shown in the display in mg/l Manganese. SP600 Spectrophotometer 01/2008...
13. Place the vial (the sample) in the sample chamber making sure that the marks are aligned. Zero accepted prepare Test 14. Press TEST key. press TEST The result is shown in the display in mg/l Manganese. SP600 Spectrophotometer 01/2008...
Page 165
The turbidity should disappear after point 7. 4. Water samples containing more than 5 mg/l iron should be allowed to react at least 10 minutes. 5. Conversion: mg/l MnO = mg/l Mn x 2.17 KMnO SP600 Spectrophotometer 01/2008...
Test 10. Press TEST key. press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Mangan. SP600 Spectrophotometer 01/2008...
Page 167
Sodium hydroxide before test. Do not exceed pH 5, as manganese may precipitate. 3. Interferences: Interfering substance Interference level Calcium greater than 700 mg/l Chloride greater than 70 000 mg/l Iron greater than 5 mg/l Magnesium greater than 100 000 mg/l KMnO SP600 Spectrophotometer 01/2008...
10. Close the vial tightly with the cap. 11. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted 12. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l Molybdate / Molybdenum. SP600 Spectrophotometer 01/2008...
Page 169
2. Under test conditions (pH 3.8 – 3.9) iron does not interfere nor do other metals at levels likely to be found in industrial water systems. 3. Conversions: mg/l Mo = mg/l MoO x 0.6 mg/l Na = mg/l MoO x 1.3 SP600 Spectrophotometer 01/2008...
12. Press TEST key. prepare Test Wait for a reaction period of 5 minutes. press TEST Countdown After reaction period is finished the measurement starts 5:00 automatically. The result is shown in the display in mg/l Molybdate / Molybdenum. SP600 Spectrophotometer 01/2008...
Page 171
5 minutes is increased. So it is very important to perform the test procedure continuously. 4. Substances who may interfere when present in concentrations at: Aluminium 50 mg/l Chromium 1000 mg/l Iron 50 mg/l Nickel 50 mg/l Nitrite all levels SP600 Spectrophotometer 01/2008...
10. Press TEST key. prepare Test press TEST Wait for a reaction period of 3 minutes. Countdown 3:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Nickel. SP600 Spectrophotometer 01/2008...
Page 173
1.1 Methods Notes: 1. The test sample and the reagents should have room temperature for test performance. 2. The test sample should have a pH of between 3 and 9. SP600 Spectrophotometer 01/2008...
10. Press TEST key. prepare Test press TEST Wait for a reaction period of 3 minutes. Countdown 3:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Nickel. SP600 Spectrophotometer 01/2008...
Page 175
1.1 Methods Notes: 1. The test sample and the reagents should have room temperature for test performance. 2. The test sample should have a pH of between 3 and 9. SP600 Spectrophotometer 01/2008...
10. Place the vial (the sample) in the sample chamber making sure that the marks are aligned. Zero accepted prepare Test 11. Press TEST key. press TEST The result is shown in the display in mg/l Nitrate. SP600 Spectrophotometer 01/2008...
Page 177
1.1 Methods Notes: 1. Some solids may not dissolve. 2. Conversion: mg/l NO = mg/l N x 4.43 SP600 Spectrophotometer 01/2008...
Test 9. Press TEST key. press TEST Wait for a reaction period of 15 minutes. Countdown 15:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Nitrate. SP600 Spectrophotometer 01/2008...
Page 179
1.1 Methods Notes: – 1. Nitrite concentrations greater than 2 mg/L NO lead to higher test results. 2. Great quantities of COD lead to higher test results. SP600 Spectrophotometer 01/2008...
Test press TEST 8. Press TEST key. Wait for a reaction period of 10 minutes. Countdown 10:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Nitrite. SP600 Spectrophotometer 01/2008...
Page 181
Copper (II)-ions may cause lower test results as they accelerate the decomposition of the Diazonium salt. It is improbable in practice that theses interfering ions will occur in such high concentrations that they cause significant reading errors. 2. Conversion: mg/l NO = mg/l N x 3.29 SP600 Spectrophotometer 01/2008...
8. Press TEST key. prepare Test press TEST Wait for a reaction period of 20 minutes. Countdown 20:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Nitrite. SP600 Spectrophotometer 01/2008...
Page 183
Silver ions interfere by causing precipitation. • In samples with very high concentrations of Nitrate (> 100 mg/l N) a small amount of Nitrite will be found. Such high levels of Nitrate appear to undergo a slight amount of reduction to Nitrite, either spontaneously or during the reaction time of the test. SP600 Spectrophotometer 01/2008...
9. Press TEST key. prepare Test press TEST Wait for a reaction period of 10 minutes. Countdown 10:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Nitrite. SP600 Spectrophotometer 01/2008...
Page 185
1.1 Methods Notes: 1. Store the reagents in closed containers at a temperature of + 4°C to + 8°C. 2. The test sample and the reagents should have room temperature for test performance. SP600 Spectrophotometer 01/2008...
Test 9. Press TEST key. press TEST Wait for a reaction period of 10 minutes. Countdown 10:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Nitrite. SP600 Spectrophotometer 01/2008...
Page 187
1.1 Methods Notes: 1. Store the reagents in closed containers at a temperature of + 4°C to + 8°C. 2. The test sample and the reagents should have room temperature for test performance. SP600 Spectrophotometer 01/2008...
14. Add 2 ml of the digested, treated water sample to the other TN Acid LR/HR vial (this is the sample). 15. Close the vials with the caps and swirl the vials gently several times to mix the contents (10 x, Note 9). (CAUTION: Vials warm up). SP600 Spectrophotometer 01/2008...
Page 189
12. Application: for water, wastewater and seawater 13. Interferences: Interfering substances that resulted in a concentration change of 10%: Bromide more than 60 mg/l and Chloride more than 1000 mg/l produce positive interferences. TN = Total Nitrogen SP600 Spectrophotometer 01/2008...
14. Add 2 ml of the digested, treated water sample to the other TN Acid LR/HR vial (this is the sample). 15. Close the vials with the caps and swirl the vials gently several times to mix the contents (10 x, Note 9). (CAUTION: Vials warm up). SP600 Spectrophotometer 01/2008...
Page 191
12. Application: for water, wastewater and seawater 13. Interferences: Interfering substances that resulted in a concentration change of 10%: Bromide more than 60 mg/l and Chloride more than 1000 mg/l produce positive interferences. TN = Total Nitrogen SP600 Spectrophotometer 01/2008...
11. Remove the vial from the sample chamber. 12. Open one reagent tube and add 0.5 ml of the pre- treated sample (step 8). 13. Close the vial tightly with the cap and invert several times to mix the contents. (Caution: Vial becomes warm!) SP600 Spectrophotometer 01/2008...
Page 193
1. This test determines the inorganic compounds Ammonium, Nitrate and Nitrite, as well as organic compounds like amino acid, urea, complexing agents etc. 2. Nitrogen compounds which are hardly to oxidise, as may be found in industrial sewage, are not digested or only partially. SP600 Spectrophotometer 01/2008...
11. Remove the vial from the sample chamber. 12. Open one reagent tube and add 0.5 ml of the pre- treated sample (step 8). 13. Close the vial tightly with the cap and invert several times to mix the contents. (Caution: Vial becomes warm!) SP600 Spectrophotometer 01/2008...
Page 195
1. This test determines the inorganic compounds Ammonium, Nitrate and Nitrite, as well as organic compounds like amino acid, urea, complexing agents etc. 2. Nitrogen compounds which are hardly to oxidise, as may be found in industrial sewage, are not digested or only partially. SP600 Spectrophotometer 01/2008...
8. Press TEST key. prepare Test press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l active Oxygen. SP600 Spectrophotometer 01/2008...
Page 197
* Active Oxygen is a synonym for a common disinfectant (based on “Oxygen”) in Swimming Pool Treatment. 1. When preparing the sample, the escape of Oxygen gases, e.g. by pipetting or shaking, must be avoided. 2. The analysis must take place immediately after taking the sample. SP600 Spectrophotometer 01/2008...
Ozone in the presence of Chlorine. >> with Cl for the determination of Ozone in the absence of Chlorine. >> without Cl Select the desired method with the arrow keys [ ] and [ ]. Confirm with [ ] key. SP600 Spectrophotometer 01/2008...
Page 199
10 ml of the diluted sample will be mixed with the reagent and the measurement repeated. 6. If ??? is displayed at the diffentiated test result see page 325. Oxidizing agents such as Bromine, Chlorine etc. interfere as they react like Ozone. SP600 Spectrophotometer 01/2008...
14. Fill a second beaker with 10 ml of water sample. Add one Glycine tablet straight from the foil and crush the tablet using a clean stirring rod. Dissolve the tablet. SP600 Spectrophotometer 01/2008...
Page 201
Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in: *,** mg/l O mg/l Ozone *,** mg/l tot Cl mg/l total Chlorine Notes: see page 197 SP600 Spectrophotometer 01/2008...
Test press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Ozone. Notes: see page 197 SP600 Spectrophotometer 01/2008...
Page 204
2-3 drops of water sample. 11. Add one DPD No. 1 tablet and one DPD No. 3 tablet straight from the foil and crush the tablets using a clean stirring rod. SP600 Spectrophotometer 01/2008...
Page 205
Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in: *,** mg/l O mg/l Ozone *,** mg/l total Cl mg/l total Chlorine Notes: see page 197 SP600 Spectrophotometer 01/2008...
Page 206
Test press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Ozone. Notes: see page 197 SP600 Spectrophotometer 01/2008...
7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted 8. Press TEST key. prepare Test press TEST The result is shown in the display as pH-value. SP600 Spectrophotometer 01/2008...
Page 209
1 molar 2 molar 3 molar - 0.21 - 0.26 - 0.29 The values of Parson and Douglas (1926) are based on the use of Clark and Lubs buffers. 1 Mol NaCl = 58.4 g/l = 5.8 % SP600 Spectrophotometer 01/2008...
7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted 8. Press TEST key. prepare TEST press Test The result is shown in the display as pH-value. SP600 Spectrophotometer 01/2008...
Page 211
This can be minimised by using a pipette (0.18 ml PHENOL RED solution is equivalent to 6 drops). 3. After use replace the bottle cap securely. 4. Store the reagent in a cool, dry place ideally at between 6°C and 10°C. SP600 Spectrophotometer 01/2008...
9. Press TEST key. prepare Test press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Phenol. SP600 Spectrophotometer 01/2008...
Page 213
3. Interferences can be caused in the presence of reducing agents, oxidising reagents, sulphides or suspended solids. Distillation of the sample is necessary then (see: “Standard Methods for Examination of Water and Wastewater, 20 Edition, 4-40 f.”). 4. Wastewater and seawater samples may also require a distillation. SP600 Spectrophotometer 01/2008...
Tube Test, 3 – 60 mg/l PO Determination of ortho-Phosphate ions Phosphate, acid hydrolyzable with Tube Test, 0.02 – 1.6 mg/l P Determination of ortho-Phosphate ions + condensed inorganic Phosphates More information you can find in the notes according to the methods. SP600 Spectrophotometer 01/2008...
Page 215
10 mg/l Iron greater than 100 mg/l Nickel greater than 300 mg/l Silica (Silicium dioxide) greater than 50 mg/l Silicate greater than 10 mg/l Sulfide at any level Zinc greater than 80 mg/l Phosphate, ortho = Phosphorus, reactive SP600 Spectrophotometer 01/2008...
14. Press TEST key. prepare Test press TEST Wait for a reaction period of 2 minutes. Countdown After the reaction period is finished the measurement 2:00 starts automatically. The result is shown in the display in mg/l total Phos- phate. SP600 Spectrophotometer 01/2008...
Page 217
2. Use a funnel to add the reagent. 3. The reagent dissolves not completely. 4. See also page 213. 5. Conversions: mg/l PO = mg/l P x 3.07 mg/l P = mg/l P x 2.29 SP600 Spectrophotometer 01/2008...
14. Press TEST key. prepare Test press TEST Wait for a reaction period of 10 minutes. Countdown After the reaction period is finished the measurement 10:00 starts automatically. The result is shown in the display in mg/l total Phos- phate. SP600 Spectrophotometer 01/2008...
Page 219
1.1 Methods Notes: 1. If the analysis is performed without digestion only ortho-Phosphate ions are determined. 2. See also page 213. SP600 Spectrophotometer 01/2008...
14. Press TEST key. prepare Test press TEST Wait for a reaction period of 10 minutes. Countdown After the reaction period is finished the measurement 10:00 starts automatically. The result is shown in the display in mg/l total Phos- phate. SP600 Spectrophotometer 01/2008...
Page 221
1.1 Methods Notes: 1. If the analysis is performed without digestion only ortho-Phosphate ions are determined. 2. See also page 213. SP600 Spectrophotometer 01/2008...
9. Press TEST key. prepare Test press TEST Wait for a reaction period of 10 minutes. Countdown 10:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l ortho- Phosphate. SP600 Spectrophotometer 01/2008...
Page 223
Higher concentrations of Cu, Ni, Cr (III), V (V) and W (VI) interfere due to their colour. Silicates doe not interfere (masked by Citric acid in the tablets). 5. See also page 213. 6. Conversion: mg/l P = mg/l PO x 0.33 mg/l P = mg/l PO x 0.75 SP600 Spectrophotometer 01/2008...
9. Press TEST key. prepare Test press TEST Wait for a reaction period of 10 minutes. Countdown After the reaction period is finished the measurement 10:00 starts automatically. The result is shown in the display in mg/l ortho-Phos- phate. SP600 Spectrophotometer 01/2008...
Page 225
320 ”Posphate LR, ortho with Tablet“. 2. Only ortho-Phosphate ions react. 3. See also page 213. 4. Conversions: mg/l P = mg/l PO x 0.33 mg/l P = mg/l PO x 0.75 SP600 Spectrophotometer 01/2008...
8. Press TEST key. press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l ortho-Phos- phate. SP600 Spectrophotometer 01/2008...
Page 227
1.1 Methods Notes: 1. The reagent dissolves not completely. 2. See also page 213. 3. Conversions: mg/l P = mg/l PO x 0.33 mg/l P = mg/l PO x 0.75 SP600 Spectrophotometer 01/2008...
8. Press TEST key. prepare Test press TEST Wait for a reaction period of 2 minutes. Countdown 2:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l ortho-Phos- phate. SP600 Spectrophotometer 01/2008...
Page 229
1.1 Methods Notes: 1. Use a funnel to add the reagent. 2. The reagent dissolves not completely. 3. See also page 213. 4. Conversions: mg/l P = mg/l PO x 0.33 mg/l P = mg/l PO x 0.75 SP600 Spectrophotometer 01/2008...
7. Press TEST key. prepare Test press TEST Wait for a reaction period of 3 minutes. Countdown 3:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l ortho-Phos- phate. SP600 Spectrophotometer 01/2008...
Test press TEST Wait for a reaction period of 2 minutes. Countdown After the reaction period is finished the measurement 2:00 starts automatically. The result is shown in the display in mg/l acid hydroly- zable Phosphate. SP600 Spectrophotometer 01/2008...
Page 233
2. Use a funnel to add the reagent. 3. The reagent dissolves not completely. 4. See also page 213. 5. Conversions: mg/l PO = mg/l P x 3.07 mg/l P = mg/l P x 2.29 SP600 Spectrophotometer 01/2008...
7. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted prepare Test 8. Press TEST key. press TEST The result is shown in the display in mg/l Potassium. SP600 Spectrophotometer 01/2008...
Page 235
1.1 Methods Notes: 1. If Potassium is present a cloudy solution will be given. Single particles are uncaused necessarily by Potassium. SP600 Spectrophotometer 01/2008...
7. Immediately place the cell in the sample chamber ma- king sure that the positioning is correct. Zero accepted 8. Press TEST key. prepare Test press TEST The result is shown in the display in (m SP600 Spectrophotometer 01/2008...
Page 237
525 nm and 620 nm. In case of doubt, you should perform a wavelength scan from 330 to 780 nm using the spectrum function (mode 53). SP600 Spectrophotometer 01/2008...
Test 12. Press TEST key. press TEST Wait for a reaction period of 1 minute. Countdown After the reaction period is finished the measurement 1:00 starts automatically. The result is shown in the display in mg/l Silica. SP600 Spectrophotometer 01/2008...
Page 239
2. Phosphate ions does not interfere under the given reaction conditions. 3. If Phosphate is known to be absent, the addition of the SILICA PR tablet may be omitted. 4. Conversion: mg/l Si = mg/l SiO x 0.47 SP600 Spectrophotometer 01/2008...
9. Add one Vario LR Silica Amino Acid F F10 Powder Pack straight from the foil into the vial (the sample). 10. Close the vial tightly with the cap and swirl several times to mix the contents. SP600 Spectrophotometer 01/2008...
Page 241
The nature of these forms is not known. A pre-treatment with Sodium hydrogencarbonate and then with Sulfuric Acid will make these forms reactive to Molybdate (pre-treatment is given in “Standard methods for the Examination of Water and Wastewater” under “Silica-Digestion with Sodium Bicarbonate”). SP600 Spectrophotometer 01/2008...
Test 13. Press TEST key. press TEST Wait for a reaction period of 2 minutes. Countdown After reaction period is finished the measurement starts 2:00 automatically. The result is shown in the display in mg/l Silica. SP600 Spectrophotometer 01/2008...
Page 243
The nature of these forms is not known. A pre-treatment with Sodium hydrogencarbonate and then with Sulfuric Acid will make these forms reactive to Molybdate (pre-treatment is given in “Standard methods for the Examination of Water and Wastewater” under “Silica-Digestion with Sodium Bicarbonate”). SP600 Spectrophotometer 01/2008...
8. Press TEST key. prepare Test press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Sulfate. SP600 Spectrophotometer 01/2008...
Page 245
1.1 Methods Note: 1. If Sulfate ions are present a cloudy solution will be given. SP600 Spectrophotometer 01/2008...
9. Press TEST key. prepare Test press TEST Wait for a reaction period of 10 minutes. Countdown 10:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Sulfide. SP600 Spectrophotometer 01/2008...
Page 247
3. To avoid loss of Sulfide collect the sample carefully with a minimum of aeration. It is essential to test the sample immediately after collection. 4. The temperature of test performance should be 20°C. Difference to this temperature can lead to higher or lower results. 5. Conversion: S = mg/l S x 1.06 SP600 Spectrophotometer 01/2008...
9. Press TEST key. prepare Test press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Sulfite. SP600 Spectrophotometer 01/2008...
Page 250
Test 8. Press TEST key. press TEST Wait for a reaction period of 5 minutes. Countdown 5:00 After the reaction period is finished the measurement starts automatically. The result is shown in the display in mg/l Sulfite. SP600 Spectrophotometer 01/2008...
Swirl the vial and than place the vial in the sample chamber making sure that the marks are aligned. Zero accepted 9. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l MBAS. SP600 Spectrophotometer 01/2008...
Page 253
5. Because reaction depends on temperature, sample and tube temperature must be between 10 and 20°C. 6. Sample volume should always metered by using volumetric pipette (class A). 7. MBAS = Methylene Blue Active Substances, calculated as sodium 1-dodecanesulfonate SP600 Spectrophotometer 01/2008...
14. Place the cooled down vial in the sample chamber ma- king sure that the marks are aligned. Zero accepted 15. Press TEST key. prepare Test press TEST The result is shown in the display in mg/l TOC. SP600 Spectrophotometer 01/2008...
Page 255
MERCK KGaA. ® 4. Appropriate safety precautions and good lab technique should be used during the whole procedure. 5. Sample volume should always metered by using volumetric pipette (class A). 6. TOC = Total Organic Carbon SP600 Spectrophotometer 01/2008...
6. Place the cell in the sample chamber making sure that the positioning is correct. Zero accepted prepare Test 7. Press TEST key. press TEST The result is shown in the display in FAU. SP600 Spectrophotometer 01/2008...
Page 257
3. Collect water samples in clean glass bottles and analyse the water sample as soon as possible. It is possible to store the sample at 4°C for 48 hours. Before measurement warm up the sample to temperature at collection time. SP600 Spectrophotometer 01/2008...
11. Add one AMMONIA No. 2 tablet straight from the foil to the same water sample and mix to dissolve with a clean stirring rod. SP600 Spectrophotometer 01/2008...
7. Before analysing seawater samples, a measuring spoon of Ammonia Conditioning Powder must be added to the sample and swirled to dissolve before AMMONIA No. 1 tablet is added. 8. Fill the vial with the same size drops by holding the bottle vertically and squeeze slowly. SP600 Spectrophotometer 01/2008...
9. Place the vial in the sample chamber making sure that marks are aligned. Zero accepted press ZERO 10. Press TEST key. press TEST The result is shown in the display in mg/l Zinc. SP600 Spectrophotometer 01/2008...
Page 261
To dechlorinate add one DECHLOR tablet to the water sample (point 1). Crush and mix to dissolve the tablet. Then add the COPPER / ZINC LR tablet (point 2) and continue with the test procedure as described above. SP600 Spectrophotometer 01/2008...
Add the same size drops to the water sample by holding the bottle vertically and squeezing slowly. After use replace the bottle caps securely noting the colour coding. Note recommendation for storage (e.g. cool and dry). Powder Packs: SP600 Spectrophotometer 01/2008...
MA 03 (alkaline, phosphate-free) ® ® from Merck KGaA). b. Rinse with tap water thoroughly. c. On demand (see Notes) perform special cleaning at this point, e.g.: rinse with diluted Hydrochloric acid solution. d. Rinse with deionized water thoroughly. SP600 Spectrophotometer 01/2008...
– e.g. due to the formation of condensation in the area of the lens or on the vial. 10. To avoid errors caused by stray-light do not use the instrument in bright sunlight. Correct filling of the vial: SP600 Spectrophotometer 01/2008...
The corresponding displayed result is 10 mg/l. Total volume = 100 ml + 5 ml = 105 ml Correction factor = 105 ml / 100 ml = 1.05 Corrected result = 10 mg/l x 1.05 = 10.5 mg/l SP600 Spectrophotometer 01/2008...
2.1 Operating 2.1.1 Start Up Before using the photometer SP600 it is necessary to insert two batteries. Then select language (mode 10) and perform “Delete Data” (mode 34). Following set date and time (see chapter 2.4 Photometer settings). 2.1.2 Batteries Saving data –...
• 50 mm cells: no cell holder required. Cylindric cells (16 and 24 mm diameter): Note: Cylindric cells are called “vials” in the test instructions. Vials are placed as indicated in the cell holder, the 2 marks matching each other. SP600 Spectrophotometer 01/2008...
Displaying date and time / user-countdown 2.2.2 Displaying time and date: Press [“clock”] key. 19:30:22 2006-06-15 The display shows: After 15 seconds the photometer reverts to the previous display automatically or press [ ] key or [ESC]. SP600 Spectrophotometer 01/2008...
2 minutes, 0 seconds = [0][2][0][0]. Confirm with [ ] key. The display shows: Countdown 02:00 Start count down with [ ] key. Start: After countdown has finished the photometer reverts to the previous display automatically. SP600 Spectrophotometer 01/2008...
• function of the tungsten halogen lamp • function of the step motor • accuracy of wavelength with internal Didymium filter and if necessary adjustment (in this case the selftest can be taken up to 5 minutes) • function of data storage When selftest is passed the list for method selection is dis- played. SP600 Spectrophotometer 01/2008...
Line 2: Range with chemical species 1 0.05-4 mg/l PO Line 3: Range with chemical species 2 0.02-1.3 mg/l P Line 4: Range with chemical species 3 0.04-3 mg/l P Chemical Species Method Information Methodlist Information Information Method SP600 Spectrophotometer 01/2008...
ZERO Prepare a clean vial as described in “Method” and place the vial in the sample chamber making sure that the marks are aligned. Zero Press ZERO key. Zero accepted prepare Test The display shows: press TEST SP600 Spectrophotometer 01/2008...
In this case the operator is responsible for ensuring the necessary reaction period by himself. Non-compliance with reaction periods lead to incorrect test results. 2. The time remaining is displayed continuously. The beeper indicates the last 10 seconds. SP600 Spectrophotometer 01/2008...
[ ] directly. (The assignment for the Code-No. is then 0 automatically.) The entire data set is stored with date, time, Code-No., method and test result. Stored! The display shows: The test result is then shown again. SP600 Spectrophotometer 01/2008...
0.02-6 mg/l Cl Profi-Mode: no 2006-07-01 14:53:09 Test No.: 1 Code-Nr.: 007 4.80 mg/l Cl The test No. is an internal number that is set automatically if a test result is stored. It appears only at the print out. SP600 Spectrophotometer 01/2008...
The display shows: press ZERO 2.3.11 Selecting a new method Press ESC key to return to method selection. Or enter the required method number directly, e.g. [1] [6] [0] for Cyanuric acid. Confirm with [ ] key. SP600 Spectrophotometer 01/2008...
Switching the acoustic signal on/off to indicate end of reading Spectrum (Scan) Absorption scanning about a max. range between 300 and 900 nm Storage Displaying all stored results Stor., code Displaying only results of a selected Code-No. range SP600 Spectrophotometer 01/2008...
Press [MODE] [1] [0] keys. Mode Confirm with [ ] key. The display shows: <Language> Deutsch >> English Press arrow key [ ] or [ ] to select the required language Francais from the displayed list. Confirm with [ ] key. SP600 Spectrophotometer 01/2008...
Page 281
14. May 2006 = [0][6][0][5][1][4] Enter hours and minutes yy-mm-dd hh:mm e.g.: 3.07 p.m. = [1][5][0][7] 06-05-14 15:07 Confirm with [ ] key. Note: While conforming date and time with [ ] key the seconds are adjusted to zero automatically. SP600 Spectrophotometer 01/2008...
Page 282
The “user-countdown” is also available if the countdown is switched off. 2. If the countdown function is switched off, the operator is responsible for ensuring the necessary reaction period by himself. Non-compliance with reaction periods lead to incorrect test results. SP600 Spectrophotometer 01/2008...
Page 283
• Press [0] key to switch the signal-beep off. • Press [1] key to switch the signal-beep on. Confirm with [ ] key. Note: In the case of methods with reaction periods, an acoustic signal still sounds during the last 10 seconds of the countdown even if the key-beep is switched off. SP600 Spectrophotometer 01/2008...
Press [ ] key for printing out all stored test results. Test No.: The display shows e.g.: After printing the photometer goes back to mode menu automatically. Note: It is possible to cancel the entering by [ESC]. All stored data are printed out. SP600 Spectrophotometer 01/2008...
Page 285
After printing the photometer goes back to mode menu automatically. Note: It is possible to cancel the entering by [ESC]. If you want to print only results of one day enter the same date twice to characterise the period. SP600 Spectrophotometer 01/2008...
Page 286
It is possible to cancel the entering by [ESC]. If you want to print only results of one Code-Number enter the same Code-Number twice. If you want to print all results without Code-No. (Code-Nr. is 0) enter Zero [0] twice. SP600 Spectrophotometer 01/2008...
Page 287
[ ] key. <Print> The display shows: method 30 Alkalinity-total Start: Press [ ] key and all stored results of the selected method cancel: ESC are printed. After printing the photometer goes back to mode menu automatically. SP600 Spectrophotometer 01/2008...
Page 288
Confirm with [ ] key. Finish with ESC key. Flow Control will be set to the selection displayed at “is”. Press [2] key to select “Baud rate”. <Baud rate> is: 19200 The display shows: select: [ ] [ ] save: cancel: SP600 Spectrophotometer 01/2008...
• Press [F3] key to print the displayed result. • Press [F2] key to print all results. • End with [ESC]. • Press arrow key [ ] to display the following test result. • Press arrow key [ ] to display the previous test result. no data If there are no test results in memory the display shows: SP600 Spectrophotometer 01/2008...
Page 290
F2 • Press [F3] key to print the displayed result. • Press [F2] key to print all selected results. • End with [ESC]. Note: It is possible to cancel the entering by [ESC]. If you want to recall only results of one day enter the same date twice to characterise the time period. SP600 Spectrophotometer 01/2008...
Page 291
It is possible to cancel the entering by [ESC]. If you want to recall only results of one Code-Number enter the same Code-Number twice. If you want to recall all results without Code-No. (Code-Nr. is 0) enter Zero [0] twice. SP600 Spectrophotometer 01/2008...
Page 292
[ ] key. The display shows: <Storage> method • Press [ ] key and all stored results of the selected 30 Alkalinity-total method are displayed. Start: cancel: ESC print: F3 • Press [F3] key to print the displayed result. print all: F2 • Press [F2] key to print all selected results. • End with [ESC]. SP600 Spectrophotometer 01/2008...
1. Fill a clean vial (24 mm Ø) with exact 10 ml of deionized Zero: deionised water water, close tightly with the cap. press ZERO 2. Place the vial in the sample chamber making sure that the marks are aligned. SP600 Spectrophotometer 01/2008...
Page 294
As the test result is highly dependent on exact sample and reagent volumes, the sample and reagent volumes should always be metered by using a 10 ml resp. 2 ml volumetric pipette (class A). Error, absorbance T2>T1 If there appears an error message please repeat adjust- ment. SP600 Spectrophotometer 01/2008...
Page 295
If the user calibration is deleted the factory calibration is automatically activated. Remarks: The method “Fluoride” cannot be calibrated with mode 45 since the test requires a calibration related to the batch of the liquid reagent (SPADNS) (mode 40, chapter calibration (fluoride)”). SP600 Spectrophotometer 01/2008...
Page 296
Cancel user calibration by pressing [ESC] key. Jus Factor The display shows: saved Chlorine T Now the method name is displayed inverse and the test 0.02-6 mg/l Cl2 result is calculated with the new calibration factor. 1.00 mg/l free Cl2 SP600 Spectrophotometer 01/2008...
Page 297
The display shows: <user calibration> Chlorine T 0.02-6 mg/l Cl2 clear user calibration? YES: 1, NO: 0 Press [1] key to delete user-calibration. Press [0] key to keep the valid user-calibration. The instrument goes back to Zero-query automatically. SP600 Spectrophotometer 01/2008...
Storage of test results is possible. In case of stored test results the display shows “Profi-Mode” additionally. The selected settings are kept by the photometer also after it was switched off. To change photometer setting a new setting is required. SP600 Spectrophotometer 01/2008...
Page 299
Test chamber. press TEST Test Press TEST key. < Abs / Trans > The result is shown in the display as Extinction (in Abs) and wavelength: 540 nm Transmission (in %). E: 0.596 T: 25, 3 % SP600 Spectrophotometer 01/2008...
Page 300
Place a filled cell (the blank) into the sample chamber (Note 1). Zero Press ZERO to start null balance (the baseline). Zero accepted prepare Test The display shows: press TEST Place the cell with the test solution into the sample cham- ber. Press TEST key. Test SP600 Spectrophotometer 01/2008...
Page 301
1. It is possible to perform the baseline against air. For the measurement of water samples it is recommend to use a cell filled with deionized water. 2. 1000 mAbs = 1.000 E (or Abs) 1 mAbs = 0.001 E (or Abs) SP600 Spectrophotometer 01/2008...
Page 302
[6] [0]. Confirm with [ ] key. Time delay: _ _ _ s Enter the interval time in the range from 6 to 999 sec., Interval tine: e.g. [2] [0]. Confirm with [ ] key. _ _ _ s SP600 Spectrophotometer 01/2008...
Page 303
E + _ _ • Press [ ] or [ ] key to change between plus- and minus-sign. • Enter the exponent value, e.g. – 1 Confirm with [ ] key. After confirmation the sample measurement starts automa- tically (see ‘Flow of Standard/Sample Measurement’). SP600 Spectrophotometer 01/2008...
Page 304
If the same zero shall be used press [Test] key. To measure a new Zero press [Zero] key. Time delay: 28 s The remaining time delay is displayed as count-down. After expiration the first measurement takes place. SP600 Spectrophotometer 01/2008...
Page 305
121.3673 = 1.213567 E+02. All displayed values are limited to 9.999 E ± 09. This value is an error-indicator, there is no additional error message. standard: conc. will be displayed during the If a standard measuring is performed measurement. SP600 Spectrophotometer 01/2008...
If only a few methods are required it is recommendable to perform Mode 62 first, followed by Mode 60. All user-Polynomials (1-25) and -Concentrations (1-10) are displayed in the method list, although they are not programmed by the user. Non-programmed user-methods can’t be activated! SP600 Spectrophotometer 01/2008...
Page 307
Press [MODE] [6] [2] keys. Confirm with [ ] key. <Mlist all off> The display shows: switch off all methods YES: 1, NO: 0 • Press [1] key to display only one method in the method selection list. • Press [0] key to keep the valid method selection list. The instrument goes back to mode menu automatically. SP600 Spectrophotometer 01/2008...
Page 308
Enter the required wavelength in the range from 330 to 900 nm, e.g.: 550 nm. Confirm with [ ] key. choose unit: >> mg/l Press [ ] or [ ] keys to select the required unit. mmol/l mAbs µg/l SP600 Spectrophotometer 01/2008...
Page 309
Confirm with [ ] key. S1: 0.05 mg/l 0.012 Enter the concentration of the second standard; e.g.: 0.1 S1 accepted • One step back with [ESC]. S2: +________ | ESC | F1 • Press [F1] key to reset numerical input. Confirm with [ ] key. SP600 Spectrophotometer 01/2008...
Page 310
Entering a User-Polynomial: Press [MODE] [6] [5] keys. Mode Confirm with [ ] key. The display shows: <User polynoms> choose no.: ____ (800-824) Enter a method-number in the range from 800 to 824, e.g.: [8] [0] [0] SP600 Spectrophotometer 01/2008...
Page 311
• Press [ ] or [ ] key to change between plus and minus Min E: +________ sign. Max E: +________ • Enter the values in Absorbance (E = extinction) for the upper limit (Max) and the lower limit (Min). Confirm every input with [ ] key. SP600 Spectrophotometer 01/2008...
Page 312
Save all your polynomial data in a written form because in case of power outage (e.g. changing the battery) all polynomial data will be lost and must be entered again. You might want to use Mode 67 to transfer all polynomial data to a PC. SP600 Spectrophotometer 01/2008...
Page 313
(in the range from 800 to 824 or 850 to 859), e.g.: 800 Confirm with [ ] key. M800 There is displayed the query: delete? YES: 1, NO: 0 • Press [1] key to delete the selected User-Method. • Press [0] key to keep the valid User-Method. The instrument goes back to mode menu automatically. SP600 Spectrophotometer 01/2008...
Page 314
The display shows: Start: Press [ ] key to print out the data (e.g. wavelength, unit, ...) of all stored User-Methods. M800 The display shows e.g.: M803 After data transfer the photometer goes back to mode menu automatically. SP600 Spectrophotometer 01/2008...
Page 315
Confirm with [ ] key. The display shows: <User m. init> Start: Confirm with [ ] key. There is displayed the query: Initialising? YES: 1, NO: 0 • Press [1] key to start initialisation. • Press [0] key to to cancel without initialisation. The instrument goes back to mode menu automatically. SP600 Spectrophotometer 01/2008...
5 and 800 mg/l CaCO total dissol. solids 0<=TDS<=6000 The display shows: +_ _ _ _ Enter the value for TDS (Total Dissolved Solids) in the range between 0 und 6000 mg/l and confirm with [ ] key. SP600 Spectrophotometer 01/2008...
Page 317
Press [MODE] [7] [1] keys. Confirm with [ ] key. <temperature> The display shows: 1: °C 2: °F Press [1] key to select degree Celsius. Press [2] key to select degree Fahrenheit. The instrument goes back to mode menu automatically. SP600 Spectrophotometer 01/2008...
Press arrow key [Test] to decrease brightness of the display Zero Test about ten units. 0...254 : 200 The display shows: The brightness can be selected between 0 and 254 units, e.g.: 200. Confirm with [ ] key. SP600 Spectrophotometer 01/2008...
, cancel: Esc Press arrow key [ ] to display the number of performed tests and free memory capacity. <System-Info> Number of Tests: free records left Finish with [ESC] key. cancel: Esc SP600 Spectrophotometer 01/2008...
RS232 interface SP600 There are two RS 232 interfaces at the photometer: Interface (F) is for factory use only. Interface (D) the operator needs to transfer data to a PC or printer or to perform an Update. SP600 Spectrophotometer 01/2008...
Caution: Adjust printing parameter in Mode 29. See chapter 2.4.3 Printing Parameter. 2.5.2 Data transfer to a personal computer Transferring test results from the photometer to a personal computer requires a transfer program, e.g. HyperTerminal. Please find detailed information at our homepage on the download-area. SP600 Spectrophotometer 01/2008...
The magnet part of the delivery content is needed for updating the instrument! Remark: To prevent loss of stored test results store or print out them before performing an Update. SP600 Spectrophotometer 01/2008...
1 Photometer 2 batteries AA/LR6 1 Power adapter, 100 – 240 V, 50 – 60 Hz 1 Cable for connection to PC 1 Magnet (for updating SP600) 1 Instruction manual 1 Manufacturer‘s Test Certificate M 1 Guarantee declaration Reagent sets and cells are not part of the standard scope of delivery.
5 – 40 °C at max. 30 –90 % relative humidity (without condensation) Specific accuracy of the photometer only applies at 20 to 25°C Language options English, German, French, Spanish, Italian, Portugese; further languages via Internet-Update Storage capacity approx. 1000 data sets Subject to technical modification! SP600 Spectrophotometer 01/2008...
Underrange for a test result a of higher/lower concentration. not possible user calibration is not possible Zero not Light absorption Refer to chapter 2.3.4 accepted is too great or too low Performing Zero (page 272) Clean sample chamber. Repeat zeroing. SP600 Spectrophotometer 01/2008...
Page 327
Chlorine. The test should be repeated with a diluted sample. Error absorbance calibration of Fluoride was Repeat calibration e.g.: T2>T1 not correct Printer “timeout” printer switched off; Connect printer no connection Check connections Switch printer on SP600 Spectrophotometer 01/2008...
Profi-Mode off with Mode 50. It seems that a method is Method is not activated in Activate the required method not available. the user method list. in the user method list with Mode 60. SP600 Spectrophotometer 01/2008...
4) Take the new pre-adjusted halogen lamp module out of the packing an fix it. 5) Fix the screws and put the metal plate back into position. 6) Connect the instrument to the AC power and start selftest. SP600 Spectrophotometer 01/2008...
3. Remove the old batteries. 4. Place the new batteries into the holder with the correct polarity. 5. Put the cover back into position and lock it into place. Caution: Dispose batteries according to the local regulations. SP600 Spectrophotometer 01/2008...
Electromagnetic interference with field strength greater than 3V/m may increase the specified tolerances. For data transfer and update use the cable delivered with the instrument only. Dortmund, 26. Februar 2007 Cay-Peter Voss, Managing Director SP600 Spectrophotometer 01/2008...