  • Page 1 Slit lamp inStructionS for uSe H Series...
  • Page 2 Copyright © Keeler Limited 2013. Published in the UK 2013. S L I T L A M P b y K E E L E R...
  • Page 3: Table Of Contents

    contentS Please click on the contents to go straight to your chosen section or navigate by using the ‘Next’ and ‘Back’ buttons to the right. Clicking on ‘Home’ will bring you back to this page. Introduction Assembly Symbols used in these Instructions for Use Instructions for use and the Slit Lamp packaging Description of filters, apertures and magnifications...
  • Page 4: Introduction

    Thank you for choosing this Keeler product. Please read this manual carefully before using your Keeler Slit Lamp, this will ensure the safety of the patient and ensure you get the best performance from this precision optical device. S L I T L A M P b y K E E L E R...
  • Page 5: Symbols Used In These Instructions For Use

    SymbolS uSed in theSe inStructionS for uSe and the Slit lamp packaging Manufacturer’s name and address Reference Number Serial Number Mandatory action sign Date of manufacture Follow Instructions for Use The CE mark on this product indicates it has been tested to and conforms Optical radiation hazard with the provisions noted within the 93/42/EEC Medical Device Directive Warning: Dangerous Voltage...
  • Page 6: Indications For Use

    Keeler or can be mounted on a third parties table top the cornea epithelium to the posterior capsule. It is used to aid in the (refraction unit) by suitably trained technicians.
  • Page 7: Safety

    The retinal exposure dose for a photochemical hazard is a product of the Keeler Ltd shall on request, provide the user with a graph showing the radiance and the exposure time. If the value of radiance were reduced in half, relative spectral output of the instrument twice the time would be needed to reach the maximum exposure limit.
  • Page 8 • C heck your Keeler product for signs of transport / storage damage prior • C are should be taken when handling halogen bulbs. Halogen bulbs can to use. shatter if scratched or damaged.
  • Page 9: Cleaning And Disinfection Instructions

    Storage and Working conditionS Before any cleaning of the instrument or the base unit, The following ambient condition limits are recommended for the Keeler ensure the power lead is disconnected. Slit Lamp, for transport and storage it is recommended that the Slit Lamp is kept in its original manufacturers packaging.
  • Page 10: Names Of Controls And Components

    7. Chinrest height adjuster 27. Runner covers 8. Patient grab handles 9. Power lead, chinrest to lamp housing keeler Slit lamp, h SerieS 28. R type tonometer mounting hole 10. Lamp cover 29. Yellow filter knob (up = out) 11. Lamp cover release screws 30.
  • Page 11: Assembly

    3rd edition compliant line with each other. Check that the guide wheels are tight. If you are fitting or have fitted your Slit Lamp to a medical or Keeler table Fit the Runner covers to the Runners by gently sliding them inwards, leg/base ensure it is situated on a firm and level floor.
  • Page 12 10.2 illumination toWer aSSembly procedure 10.3 cable attachment procedure Remove the hex bolt (a) from the base of the Illumination tower, and Connect the main lamp cable from the Chinrest to the then place the illumination tower on the Slit Lamp base with the base illumination tower.
  • Page 13 Connect the mains power to the Slit Lamp transformer using the power lead provided. 3 pin to 3 pin Fixation light socket Only a hospital grade 3-conductor electrical power supply cable must be used. For USA and Canada: Detachable power supply cord set, UL listed, type SJE, SJT or SJO, 3-conductor, not smaller than 18 AWg.
  • Page 14 10.4 fitting applanation tonometerS, t type and r typeS KEELER APPLANATION TONOmETER (T-TyPE) APPLANATION TONOmETER ‘KEELER FIxED’ (R-TyPE) This instrument is for those who wish the tonometer to • P osition the guide plate in the tonometer/test bar remain permanently on the slit lamp. support hole on the slit lamp.
  • Page 15: Instructions For Use

    inStructionS for uSe 11.1 Setting the binocularS Adjust the eyepieces pupillary distance by holding both eyepiece bodies and rotating them inwards or outwards until they are correct for your PD. Turn both eyepieces (32) to maximum plus (+) correction. It is vital that the binoculars are optimised for the user’s optical correction in order to obtain focused binocular images.
  • Page 16 Tightening the knob will re- to the examination. Keeler recommends the use of disposable centre the slit image in the centre of the visual field of the microscope.
  • Page 17: Description Of Filters, Apertures And Magnifications

    deScription of filterS, apertureS and magnificationS Stereo microScope filterS Eyepieces 12.5x Dioptric adjustment +/- 8D PD range 49mm – 77mm Convergent angle of optical axis 13° An Infra Red (IR) cut filter is permanently fitted for safety reasons 5 step magnification change Clear Neutral Density Red Free magnification...
  • Page 18: Routine Maintenance

    LED life. • D isconnect the power cord to the lamp housing and Lift In the unlikely event of an LED failure please contact Keeler or remove the lamp cover on tower illumination systems your local distributor for guidance on the replacement procedure. by loosening the two screws on the lamp housing.
  • Page 19: Warranty

    Warranty = 3 yearS The mirror should be cleaned only with a soft clean lens cloth. The Keeler H-Series Slit Lamps are guaranteed for three years (3) against faulty workmanship materials or factory assembly. Warranty is on a Return The mirror is an interference fit in its holder and can be removed by To Base (RTB) basis at the cost of the customer and may be void if the Slit grasping it firmly and pulling it out of its holder.
  • Page 20: Specifications And Electrical Ratings

    SpecificationS and electrical ratingS The Keeler Slit Lamp is a medical electrical instrument. The instrument 15.2 interference immunity requires special care concerning electromagnetic compatibility (EMC).This guidance and manufacturer’s declaration – electromagnetic immunity Section describes its suitability in terms of electromagnetic compatibility The Keeler Slit Lamp is intended for use in the electromagnetic of this instrument.
  • Page 21 Keeler Slit Lamp is used exceeds the applicable RF compliance level above, the Keeler Slit Lamp should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orientating or relocating the Keeler Slit Lamp.
  • Page 22 15.5 technical SpecificationS SyStem componentS Part number Slit Lamp 40H, Standard Set complete with Table Top 3020-P-2000 optical SyStem Slit Lamp 3020-P-5032 3020-P-5040 Type Galilean converging binoculars @ 8° Slit Lamp 40H, Refraction Unit Set 3020-P-2003 magnification Rotating drum change x6, x10, x16, x25 & x40 Slit Lamp 3020-P-5032 Power Supply and Socket Kit...
  • Page 23: Accessories And Spares

    acceSSorieS and SpareS Supplied With your Slit lamp Part name Part number Part name Part number Test bar EP39-80243 Chinrest papers 3104-L-8200 Chinrest assembly 3020-P-5036 Bulb 1030-P-7160 Power supply to Slit Lamp base 3020-P-7011 cable (Table top variant only) Small projection mirror EP39-80250 Base assembly (Table top variant 3020-P-5007...
  • Page 24 Supplied With your Slit lamp additional acceSSorieS Part name Part number Part name Part number mains Cables – EU MIS138 Volk digital lens set 2105-L-2010 Tonometer R type 2414-P-2040 mains Cables – Brazil 3020-P-7007 mains Cables – Japan 3020-P-7008 mains Cables – UK MIS100 Tonometer T type 2414-P-2030...
  • Page 25: Contact, Packaging And Disposal Information