1 Liter Bag Sampler, Model 1060
1. SAMPLE PREPARATION AND COLLECTION:
1 Place the sample bag inside sampler and close sampler lid.
2. Insert a 6" long piece of 1/4"0DX1/8"ID T y g o n © tubing through the Inlet Tube Fitting.
3. Adjust lenght of tubing so half of the tubing is inside the sampler.
4. Tighten the inlet tube fitting just enough to create a seal around the T y g o n © tubing.
5. Attach the inside portion of the T y g o n © tubing to the inlet valve on the sample bag.
6. Open sample valve on the sample bag and close the lid of the sampler.
7. Make sure the Purge Valve is closed(Closed for fastest fill rate, Open to slow fill rate).
8. Turn on the sample pump.
9. To ensure proper sample bag preparation, fill & empty the new sample bag at least three
times with a clean gas. To empty the sample bag using the sampler pump: disconectthe tubing
at the Outlet Vacuum Port; Attach a short piece of 1/4"0DX1/8"ID T y g o n © tubing from the
Pump Discharge Port to the Outlet Vacuum Port to create a presure inside the sample camber.
10. Fill pre purged sample bag for collection and analysis.
11. Turn off sample pump, open Purge Valve, open sampler lid, CLOSE SAMPLE BAG
VALVE and remove sample bag from inlet tubing.
CAUTION: It is recommended that you give the battery an overnight charge after six hours of use.
PUMP
DISCHARGE
PORT
INLET TUBE
FITTING
POWER SWITCH
BATTERY
CHARGER
PORT
/
PURGE VALVE
OUTLET VACUUM
PORT
FIGURE 1
Page 3