SafeTest 99
Manual
!
Warning of electrical danger!
Important, follow the documentation!

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Rigel SafeTest 99 and is the answer not in the manual?

Questions and answers

Summary of Contents for Seaward Rigel SafeTest 99

  • Page 1 SafeTest 99 Manual Warning of electrical danger! Important, follow the documentation!
  • Page 2 Calibration Statement The Rigel SafeTest 99 hand-held electrical safety analyzer is fully calibrated and found to be within the specified performance and accuracy at the time of production. The Seaward Group provides its products through a variety of channels, therefore it may be possible that the calibration date on the provided certificate may not represent the actual date of first use.
  • Page 3 Disposal of old product The Rigel SafeTest 99 has been designed and manufactured with high quality materials and components, which can be recycled and reused. When this symbol is attached to a product it means the product is covered by the European Directive 2012/19/EC.
  • Page 4: Table Of Contents

    Index Introduction ________________________________________________ 5 1.1 Key Features ______________________________________________________ 5 1.2 Getting to Know Your SafeTest 99: ____________________________________ 6 1.3 In the Box ________________________________________________________ 7 1.4 Optional Accessories and Replacement Spare Parts ______________________ 7 1.5 Icons used on the SafeTest 99 ________________________________________ 8 Getting Started __________________________________________ - 10 - 2.1 Setting your Language ___________________________________________ - 10 - 2.2 Displaying the SafeTest 99 Device Information _______________________ - 11 - 2.3 Accessing the SafeTest 99 Functions _ _______________________________ - 12 - Ground Continuity Measurement _ ___________________________ - 13 - NFPA-99 Ground Leakage Measurement _ _____________________ - 15 - NFPA-99 Chassis Leakage (Touch Current) Measurement ________ - 17 - NFPA-99 Patient Lead Leakage Measurement _________________ - 19 - Displaying Line Voltage Supply Characteristics _________________ - 21 - Point to Point Measurement _______________________________ - 22 - Warning Messages _______________________________________ - 24 - Maintaining the SafeTest 99 _ _______________________________ - 25 - 10.1 Cleaning the Analyzer ___________________________________________ - 25 - 10.2 User Maintenance ______________________________________________ - 25 - 10.3 Calibration or Repair. _ ___________________________________________ - 25 -...
  • Page 5 Warnings and Cautions These operating instructions are intended for the use by adequately trained personnel. Environmental Conditions The SafeTest 99 has been designed to be operated in a dry environment, at a temperature of 32 to 104 degrees F without moisture condensation, and at an operating altitude 0 – 2000m.
  • Page 6: Introduction

    1 Introduction The SafeTest 99 is a dedicated medical safety analyzer, ideal for testing high volumes of basic medical and laboratory equipment. A robust and reliable design ensures that the SafeTest 99 can withstand a busy schedule of testing medical equipment. With a large color display and a color coded user interface, it’s easy to select the required tests with a single key press, while a fast step-through of the test routine makes the testing process speedy and dependable.
  • Page 7: Getting To Know Your Safetest 99

    1.2 Getting to Know Your SafeTest 99: 1, Start key 2, Stop key 3, Reversed key 4, Function keys F1-F5 5, Open ground key 6, Open neutral key 7, EUT socket, 90-264VAC, 48-64Hz 8, Point to point auxiliary probe. 9, Ground continuity and leakage probe (Applied Parts Connector Box) 10, Line inlet 90-264VAC 48-64Hz SafeTest 99 Instruction Manual –...
  • Page 8: In The Box

    In the Box Carefully unpack all items from the box and ensure the following items are included: AC Power Supply Ground Continuity Test Probe with Clip ECG Snap Adapters (pack of 10) Calibration Certificate Applied Parts Connector Box SafeTest 99 Carry Case SafeTest 99 Instrument SafeTest 99 Quick Start Guide Applied parts Connection Box Quick Start Guide...
  • Page 9: Icons Used On The Safetest 99

    1.5 Icons used on the SafeTest 99 The SafeTest 99 features a high resolution color graphic back lit display that provides a unique user experience and to help guide the user through the different test steps. Below are of some of the icons used in the SafeTest 99: Select Ground Continuity testing Select Leakage testing Select Point to point testing (Ground Continuity and Leakage)
  • Page 10 Applies reversed mains to the EUT Warning, EUT socket live Open Ground single fault condition key Open Neutral single fault condition key Ground continuity test lead compensated Test running (note; not used on all tests) SafeTest 99 Instruction Manual – Revision 1.0...
  • Page 11: Getting Started

    2 Getting Started The SafeTest 99 can perform electrical safety tests in accordance with NFPA 99-2015. To get started, simply follow these instructions; Switch ON: To switch on the SafeTest 99, please insert the mains cable to the power inlet, the SafeTest 99 will automatically power up to the Home screen.
  • Page 12: Displaying The Safetest 99 Device Information

    2.2 Displaying the SafeTest 99 Device Information From the Home screen, select (F5) to enter the settings menu. Select (F2) to view the SafeTest 99 information • Serial Number • Hardware version • Firmware version • Last calibration date SafeTest 99 Instruction Manual – Revision 1.0 - 11 -...
  • Page 13: Accessing The Safetest 99 Functions

    2.3 Accessing the SafeTest 99 Functions From the SafeTest 99 Home screen, select the required test; Select (F1) to enter the Ground Continuity test menu. Please see section 3.0 for further information on performing a Ground Continuity test. Select (F2) to enter the Leakage test menu. Please see section 4.0 to 7.0 for further information on performing Ground, Chassis (Touch Current) and Patient Leakage test, and also displaying mains voltage.
  • Page 14: Ground Continuity Measurement

    3 Ground Continuity Measurement To perform a Ground continuity test, select the (F1) icon from Home screen. The SafeTest 99 will automatically start measuring when the Ground continuity test, (F1), is selected from the Home screen. When testing the Test running icon, , is displayed.
  • Page 15 Each time the ground bond probe is placed on a new test point, the zap circuit will be reactivated, ensuring accurate readings at every measurement point. Do not exceed the maximum permitted voltage of 30 V AC/DC with respect to ground potential! Electric Shock danger! To compensate for the test cable resistance, connect the test cable between the black ground bond socket and the EUT ground, then select the...
  • Page 16: Nfpa-99 Ground Leakage Measurement

    4 NFPA-99 Ground Leakage Measurement To perform a NFPA 99-2015 Ground leakage test, select (F2) from the Home screen, to enter the Leakage test menu. (F1), to select the Ground Leakage test menu. Note in the Leakage menu, you can switch between Leakage tests and the SafeTest 99 will alter internal settings accordingly.
  • Page 17 To apply mains voltage in normal polarity, and start the test, select the button on the front panel. The test will run until the button is pressed. To apply mains voltage in reversed polarity, and start the test, select the button on the front panel.
  • Page 18: Nfpa-99 Chassis Leakage (Touch Current) Measurement

    5 NFPA-99 Chassis Leakage (Touch Current) Measurement To perform a NFPA 99-2015 Chassis leakage (Touch Current) test, select (F2) from the Home screen, to enter the Leakage test menu. (F2), to select the Chassis Leakage (Touch Current) test menu. Note in the Leakage menu, you can switch between Leakage tests and the SafeTest 99 will alter internal settings accordingly.
  • Page 19 To apply mains voltage in normal polarity, and start the test, select the button on the front panel. The test will run until the button is pressed. To apply mains voltage in reversed polarity, and start the test, select the button on the front panel.
  • Page 20: Nfpa-99 Patient Lead Leakage Measurement

    6 NFPA-99 Patient Lead Leakage Measurement To perform a NFPA 99-2015 Patient Lead leakage test, select (F2) from the Home screen, to enter the Leakage test menu. (F3), to select the Patient Lead Leakage test menu. Note in the Leakage menu, you can switch between Leakage tests and the SafeTest 99 will alter internal settings accordingly.
  • Page 21 To apply mains voltage in normal polarity, and start the test, select the button on the front panel. The test will run until the button is pressed. To apply mains voltage in reversed polarity, and start the test, select the button on the front panel.
  • Page 22: Displaying Line Voltage Supply Characteristics

    7 Displaying Line Voltage Supply Characteristics To display the mains supply characteristics, select (F2) from the Home screen, to enter the Leakage test menu. In the Leakage test menu, select (F4) to display the line voltage, frequency and load current. Note in the Leakage menu, you can switch between Leakage tests and the SafeTest 99 will alter internal settings accordingly.
  • Page 23: Point To Point Measurement

    8 Point to Point Measurement To perform a Point to point test, select (F3) from the Home screen, to enter the Point to point test menu. Connect the Point to point probes between the black and green socket on the back panel. The EUT socket will power up during leakage tests however the mains cable is not part of the measurement circuit, it is shown as optional and not required.
  • Page 24 (F1) to perform a point to point Ground Continuity measurement (F2) to perform a point to point Leakage measurement When testing the icon is displayed. The test will run until either (F1), (F2) or the stop button is selected. If Home (F5) is selected the test is stopped and the instrument returns to the Home screen.
  • Page 25: Warning Messages

    9 Warning Messages The SafeTest 99 will warn the user of possible incorrect test setups such as secondary grounding and isolated mains supply (line voltage isolated from ground). Secondary ground warning: To perform a valid test, the secondary ground must be removed. Testing with a secondary ground will lead to invalid readings as the leakage current will flow through the low resistance secondary ground rather than the high resistance (1kΩ) body model in the SafeTest 99.
  • Page 26: Maintaining The Safetest 99

    The unit should be regularly calibrated (at least annually). 10.3 Calibration or Repair. For calibration or repair return the instrument to:- Contact details Address details Service, Calibration and Repair Seaward Group USA 6304 Benjamin Road Tel: 813 886 2775 Suite 506 Fax: 813 886 2754 Tampa, FL 33634 Email: service@seaward-groupusa.com...
  • Page 27: Technical Specifications

    11 Technical Specifications Ground Continuity Method 2 wire ZAP technique. Test Current >± 200mADC into 2Ω Max Test Voltage 4-24Vrms o/c Measuring Range (low range) 0.001 – 0.999Ω Resolution 0.001Ω Measuring Range (mid-range) 1.00 – 9.99Ω Resolution 0.01Ω Measuring Range (high range) 10.0 – 19.9Ω Resolution 0.1Ω...
  • Page 28: Appendix A Pass / Fail Limits

    Voltage rating (maximum) 230Vac Appendix A Pass / Fail Limits NFPA 99-2015 Leakage Current Type Ground Leakage – Fixed Leakage 10.0mA Chassis Leakage (Touch Current) 0.5mA Patient Lead Leakage 0.1mA 0.5mA Note pass / fail values from NFPA 99 2015 edition incorporating TIA 15-3, reference 14.2.9.1.1.
  • Page 29: Appendix Bnfpa99-2015 Body Model

    Appendix B NFPA99-2015 Body Model All resistances are ±1% All capacitances are ±5% Frequency Response Frequency (H z) 1000 10000 100000 1000000 SafeTest 99 Instruction Manual – Revision 1.0 - 28 -...