HAIRUIS P1 User Manual

Coating thickness gauge

Advertisement

Quick Links

User Manual
Coating Thickness Gauge
V1.7
P1
P1S
P2
P3
P4

Advertisement

Table of Contents
loading

Summary of Contents for HAIRUIS P1

  • Page 1 User Manual Coating Thickness Gauge V1.7...
  • Page 2: Table Of Contents

    Table of Contents 1. Introductions--------------------1 2.7 Probe Mode---------------------------12 1.1 Features--------------------1 2.8 Storage-------------------------------12 1.2 Application------------------1 2.9 Recall and Delete Readings--------------12 1.3 Principle of Measurement-------2 2.10 High and Low Limit Alarm--------------13 1.4 Supply Information------------2 2.11 Unit--------------------------------14 1.5 View------------------------3 2.12 Speed-----------------------------14 1.6 Specifications----------------6 2.13 Data Download-----------------------15 2.
  • Page 3: Introductions

    1. Introductions This compact gauge can be used for non-destructive coating thickness measurement of non-magnetic coatings, e.g. paint, enamel, chrome on steel, and insulating coatings, e.g. paint and anodizing coatings on non-ferrous metals. 1.1 Features 128*128 dot matrix LCD display, standard menu operations; Two measure mode: single and continuous;...
  • Page 4: Principle Of Measurement

    1.3 Principle of Measurement The coating thickness gauge work either on the magnetic induction principle or on the eddy current principle, depending on the type of probe used. This gauge has internal F probe and N probe. F probe work on magnetic induction principle and should be used for non-magnetic coatings such as aluminium, chrome, copper, zinc, paint and varnish,enamel, rubber etc., on an iron or steel substrate;...
  • Page 5: View

    1.5 View 1.5.1 Applied to P1 and P1S 1.Current work group(DIR and GENn,n=1-4) CoatingThickness Tester 2.High and low limit alarm AUTO 3.Probe mode:AUTO, MAG, EDDY 4.Measuring readings display 5.Statistics display Mean 101um 100um 6.Auto-power off indication 103um 7.USB connecting indication 8.Reserved...
  • Page 6 1.5.2 Applied to P2 1.Current work group(DIR and GENn,n=1-4) 2.High and low limit alarm 3.Probe mode:AUTO, MAG, EDDY AUTO 4.Measuring readings display AUTO 5.Statistics display 6.Auto-power off indication Mean 101um Mean 101um 100um 7.USB connecting indication 100um 103um 103um 8.Reserved 9.Unit(um、mm、mils) 10.Low battery indication 11.CAL-Calibration key...
  • Page 7 1.5.3 Applied to P3 and P4 1.Current work group(DIR and GENn,n=1-4) 2.High and low limit alarm AUTO 3.Probe mode:AUTO, MAG, EDDY 4.Measuring readings display 5.Statistics display Mean 101um 6.Auto-power off indication 100um 103um 7.USB connecting indication 8.Reserved 9.Unit(um、mm、mils) 10.Low battery indication 11.Reserved 12.POWER on/off key 13.
  • Page 8: Specifications

    One direct group(readings not be stored to memory) Four general groups(max. 2000 readings will be stored automatically) Four general groups(max. 320 readings for P1 and max. 2000 readings NOTE: each group have individual statistics, alarm limit settings and for P1S be stored). NOTE: each group have individual statistics, alarm...
  • Page 9 1.7 Specifications for P3 and P4 Product Code Probe F Probe N Probe F Probe N Probe Magnetic induction Eddy Currents Magnetic induction Eddy Currents Measuring principle Measuring range 0 to 1500um (0 to 59mils) 0 to 1500um (0 to 59mils) 0 to 500um (0 to 19.68mils) 0 to 500um (0 to 19.68mils) Accuracy...
  • Page 10: Use The Gauge

    2. Use the Gauge If you use the gauge for the first time, please read the chapter 3(factors of affecting measurement accuracy). 2.1 Replacing the Battery Place the gauge upside down on a suitable surface, remove the screws from the battery compartment with a crosstip screwdriver, raise the lid of the compartment, remove battery, insert new battery according to the positive and negative poles and close the lid and fasten with screws.
  • Page 11: User Interface

    Down Key a. Move down or roll down; c. Zero calibration(For P3, P4) Right Key Left Key b. Decreasing; d. Refresh LCD in measure mode(For P1, P2) Down Key Zero Calibration Key(For P1, P2) c. Key Define(For P1, P2) a. Press and hold to do zero calibration;...
  • Page 12: Measure Mode

    2.4 Measure Mode Single measure mode - Place the probe on sample vertically and rapidly, and after one beep, raise it rapidly. Continuous measure mode - Place the probe on sample, and don’t raise the probe, the measurement will be done continuously in a certain cycle. 2.5 Group Mode Direct Group(DIR)-Be intended for quick occasional readings.
  • Page 13: Statistics

    Method of switch group mode a. Press “Left Key” once to enter into menu mode(”Root” directory); b. Press “Up Key” or “Down Key” to activate “Options” item, and press “Left Key” once to enter in; c. Press “Up Key” or “Down Key” to activate ”Group Mode” item, and press “Left Key” once to enter in; d.
  • Page 14: Probe Mode

    2.7 Probe Mode There are three probe modes: AUTO, MAG and EDDY. In AUTO mode, the probe can automatically determine the type of substrate measured. In MAG mode, the probe can only measure on magnetic substrates. In EDDY mode, the probe can only measure on non-ferrous metal substrates. When a magnetic substrate detected, an “F”...
  • Page 15: High And Low Limit Alarm

    b. Press “Up Key” or “Down Key” to activate “Delete Data” item, and press “Left Key” once to enter in; c. Press “Up Key” or “Down Key” to activate ”Current Data” item, and press “Left Key” once. A dialog box will be displayed.
  • Page 16: Unit

    b. Press “Up Key” or “Down Key” to activate “Limit” item, and press “Left Key” once to enter in; c. Press “Up Key” or “Down Key” to activate “Settings” item, and press “Left Key” once to enter in; d. Press “Up Key” or “Down Key” to activate your intended item, and press “Left Key” once to enter in; e.
  • Page 17: Data Download

    a. Press “Left Key” once to enter into menu mode(”Root” directory); b. Press “Up Key” or “Down Key” to activate “Options” item, and press “Left Key” once to enter in; c. Press “Up Key” or “Down Key” to activate ”Speed” item, and press “Left Key” once to enter in; d.
  • Page 18: Factors Of Affecting Measurement Accuracy

    immediately. The final reading derives from the statistical calculation and from the guaranteed tolerance levels of the gauge. T(coating thickness)=M(mean value)±S(standard deviation)±A(measuring accuracy). 3. Factors of Affecting Measurement Accuracy User needs to know the factors of affecting measurement accuracy before using the gauge. The factors are listed as below.
  • Page 19: Calibration

    The measured sample must correspond to the calibration sample in substrate material properties, curvature radius, or the more closely the measured sample matches the calibration sample, the more accurate the measuring. In addition, curvature radius should meet the minimum value as specified in chapter 1.6. The minimum substrate thickness and the minimum measuring areas as specified in chapter 1.6 should be taken into account.
  • Page 20: Zero Calibration

    4.1 Zero Calibration Tips: For P3 or P4, press and hold “DOWN Key” to start zero calibration, and press again to exit a. Press and hold “ZERO Key” until hearing “beep...beep beep...”. The LCD will flash “ZERO” symbol. b. Place the probe on uncoated sample and raise it after one beep. c.
  • Page 21: Shot-Blasted Surfaces

    4.3 Shot-blasted Surfaces The physical nature of shot-blasted surfaces results in coating thickness readings that are too high. The mean thickness over the peaks can be determined as follows. Method one: a. The gauge should be calibrated according to 4.2(one or more points calibration). Use a smooth calibration sample with the same curvature radius and the same substrate as the later measuring sample.
  • Page 22: Trouble Shooting

    The following errors can be remedied by system reset: -Illogical readings -Several of the keys don’t work Reset to factory settings(For P1, P2): 1. Power off the gauge. 2. Press and hold “ZERO Key” and then power on the gauge.

This manual is also suitable for:

P1sP2P3P4