MAS 640 Instruction Manual Rev. 0A Table of Contents 1. Introduction................................3 2. Specification...............................3 2.1. General Technical Data ........................3 Figure 2: Mounting holes distance in mm ....................4 2.2. Applications.............................4 2.3. Technical Data A/D Converter........................4 3. Putting Into Operation............................5 3.1. Opening the Instrument..........................5 3.2.
MAS 640 Instruction Manual Rev. 0A 1. Introduction The weight indicator MAS 640SS was designed for fast and accurate high resolution measurements. Its performance is based on a 24 bit delta sigma A/D converter. Water resistant housing allows this instrument to work under very demanding industrial environments with years of constant quality.
MAS 640 Instruction Manual Rev. 0A 3. Putting Into Operation 3.1. Opening the Instrument WARNING! Before opening the instrument disconnect the power cord from the power supply. To open the instrument unscrew the 8 screws from the back and lift the cover.
MAS 640 Instruction Manual Rev. 0A 3.3. Connecting to the Serial Interface ● Insert the interface cable into the instrument through the cable gland. ● Connect the interface cable to the J10 (3-pin terminal) according to the table below. Assignment Figure 3: Location of J10 terminal on the digital module inside the instrument.
MAS 640 Instruction Manual Rev. 0A 4. Basic Functions 4.1. User Interface Figure 5. Front panel interface 4.2. Operation Keys DESCRIPTION The <POWER> key is used to turn the instrument on and off. <POWER> The <ZERO> key is used to execute zero level adjustment.
MAS 640 Instruction Manual Rev. 0A The <TARE> key is used to display zero on the display when an empty container is placed on the weighing platform. Use the <TARE> key to cancel mass of an empty container <TARE> The <FUNCTION> key is used to select one of the special function. Refer to the Special Functions [ section 5.] for the details on the available functions.
MAS 640 Instruction Manual Rev. 0A 4.4. Front Panel Annunciators (LEDs) Symbols on the Description front panel I, II Indicates which one of two ranges is actually in operation Ʃ Indicates that total weight is being displayed Indicates stable readings for weighing or counting...
MAS 640 Instruction Manual Rev. 0A 4.8. Weighing with Tare Place the empty container on the weighing platform and press the <TARE> key. Instrument will cancel a container mass by displaying zero. On the main display, after canceling a container mass, the net LED will light up just above the <TARE>...
MAS 640 Instruction Manual Rev. 0A 5.3. Setting a Filtering Level Press the <FUNCTION> key. Press the <PIECES> key repeatedly until message 'AuE' displayed. Press the <FUNCTION> key to enter filtering sub menu. Change the last digit blinking from 0 to 3 to adjust strenght of the filter.
MAS 640 Instruction Manual Rev. 0A 5.5.1. Printout Select Options. Description Net weight only printout. Example: N: 12.4 kg Net and Gross weight printout. Example: N:12.4 kg G:13.0 kg Net, Gross and Tare weight printout . Example: N: 12.4 kg G: 13.0 kg...
MAS 640 Instruction Manual Rev. 0A 5.6.1. Autozero/Zero Tracking Options Autozero option No. Description: Disable Autozero and Zero tracking Enable Zero tracking only Enable Zero tracking and Autozero 6. Setup 6.1. Entering the Setup Mode To enter the setup mode: ●...
MAS 640 Instruction Manual Rev. 0A 6.2. Main Menu Setup Functions See the Table below for available functions. Firmware version A0.01 Function name Description Setting position of decimal point for display readings. Allowed numbers: 0 to 3. 0 – no decimal point.
MAS 640 Instruction Manual Rev. 0A 6.2.2. Editing Function Content To edit content of a chosen function press the <FUNCTION> key while function name is displayed. The flashing digit and the function name will light on the display. <FUNCTION> To change the content of a chosen function use the <PIECES> key to increment the value of the flashing digit.
MAS 640 Instruction Manual Rev. 0A 7. Calibration Routines 7.1. Zero Calibration After entering the setup mode [section 6.1] repeatedly press the <PIECES> key until ZerCAL is displayed. < PIECES> Make sure that there is no load at the weighing platform.
MAS 640 Instruction Manual Rev. 0A 7.2. Span Calibration After entering the setup mode [see 6.1] repeatedly press the <PIECES> key until <PIECES> SPAn is displayed. Make sure the weighing platform is loaded by a load of weight declared in the E function of the main menu.
MAS 640 Instruction Manual Rev. 0A 8. Saving Changes < PIECES> In the main menu setup, press the <PIECES> key repeatedly until 'SauE' is displayed. Accept the savings by pressing the <FUNCTION> key. <FUNCTION> 9. Leaving Setup mode After completion of all the adjustments select the 'SAuE' function from the main menu and accept it by pressing the <FUNCTION>...