Summary of Contents for Zeiss FixAssist XTR

  • Page 1 ® FixAssist  Probe presetting device for probe systems with XTR plate Operating manual...
  • Page 3 The replication of trade names and such in this document does not justify the assumption that such names should, in the sense of trademark and brand protection legislation, be regarded as unused and therefore usable by anyone. Layout, illustration, photos and text: Carl Zeiss 3D Automation GmbH Printed in Germany 06/2017...
  • Page 4: Table Of Contents

    6.3 Transport ..........18 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 5: Introduction

    – Unauthorised structural modifications made to the product. Importance of these instructions These instructions are intended for the user. They are used to operate the FixAssist XTR safely and properly. 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 6: Safety

    Parts must be replaced immediately if not in perfect working order. Only use original spare and wear parts. If third-party parts are used, it is not possible to guarantee that they have been designed and manufactured to the relevant load and safety requirements. 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 7: Structure, Function And Technical Data

    All of the specifications in these operating instructions must be heeded during use. Misuse The FixAssist XTR is exclusively designed for configuring and assembling XTR measuring probes. Use of any other kind or outside of this scope is regarded as improper. The manufacturer cannot accept any liability for damage arising from such use.
  • Page 8: Scope Of Delivery

    – 2x DG20 hook wrench – 1x M5, V2A probe extension – 1x angle adjustment device for DG11 angle elements – 2x 2.9 pin wrench – 5x Allen key; sizes 1.5–4 mm Fig. 2: Tool kit 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 9: Design

    5. Fixing screw (yellow) for vertical guide 6. Fixing screw (black) for angle adjustment 7. Fixing screw for horizontal guide 8. Horizontal guide 9. Fixing screw (red) for plate support 10.Wing nut for locking the plate support around the Z-axis 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 10: Technical Data

    Structure, function and technical data Technical data Condition Value Weight 19 kg Dimensions 700 mm x 230 mm x 515 mm (LxWxH) Ambient temperature 5–40°C Humidity 30–95% (non-condensing) Storage temperature 5–40°C 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 11: Operation

    Extension as counter support NOTE: If possible do not use your original probe for configuring your probe system. Use old probes with a large sphere diameter for configuration, and your original probe exclusively for your measurements. 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 12: Inserting Probe Plates

    Push the red fixing screws (1) inwards and tighten them.  Important! Do not tighten the fixing screws too much, because the plate could become damaged.  Fig. 7: The probe plate is clamped in to the precise position. 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 13: Configuring The Angle Around The Z-Axis

    Re-apply the fixing screws  again.   Release the yellow fixing screw (4) of the horizontal guide and push the prism slide carefully toward the probe until it is against the Fig. 10: prism. 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 14  sphere is still centred in the  prism.   If the configuration is cor- rect, tighten the screws with the torque screwdriver.  The angle around the Z-axis is configured. Fig. 12: 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 15: Configuring The Angle Around The Second Axis

    Re-apply the fixing screw of  the vertical guides.    The prism is now set to the  correct height.   Push prism slides Fig. 14: horizontally. Remove the centring aid and probe once again, 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 16   12. Check the angle once more.   13. Replace the stainless steel  extension with the probe.  The probe system is set up Fig. 17: and ready for operation. 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 17: Maintenance

    5%. – Compressed air for cleaning by blowing out. Maintenance The FixAssist does not require any additional maintenance. Service contact Carl Zeiss 3D Automation GmbH Carl-Zeiss-Straße 32 D-73431 Aalen Germany Tel.: +49 7361 6336-0 626109-9160-000 FixAssist XTR Operating manual V1.0...
  • Page 18: Taking Out Of Operation, Disposal And Transport

    Notes on storage If the FixAssist XTR is put into storage, it must be stored in a dry and dust-free location. The recommended storage temperature is +5°C to 40°C.