Operating instructions ALPRO Weekend-System II LIST OF CONTENTS I. Operating instructions Paragraph..................................Page 1 Directions for use ......................4 2 Operating conditions of the Weekend-System ..............5 3 Operation .......................... 6 3.1 Switching the Weekend-System on and off .................... 6 3.2 Filling the tank ............................6 3.3 Chronological procedure of application ....................
Operating instructions ALPRO Weekend-System II I. Operating instructions These operating instructions must always be available near the device. They have to be read and used by each person who is instructed to the works at the device (Installation, operation and maintenance).
Operating instructions ALPRO Weekend-System II 2 Operating conditions of the Weekend-System 1 Non-operation Water exit Water supply Hand part (during the week) Tumbler Bowl rinsing Water stop Air supply Weekend-System is switched off: Pressure valve Water flows through without any hindrance.
Operating instructions ALPRO Weekend-System II 3 Operation 3.1 Switching the Weekend-System on and off The operating agents are on the side of the device, above Pressure Water stop the cover for the supplies. valve for air You can switch the compressed air for the tank on and off using the toggle switch of the pressure valve.
Operating instructions ALPRO Weekend-System II 3.3 Chronological procedure of application “On Fridays” (or before downtimes of several days’ duration): Inject BILPRON Procedure Remarks Check if enough Filling of the tank: see Chapter 3.2 BILPRON is in the tank: The level indicator at the side has to show a liquid level at „Max.“.
Page 8
Operating instructions ALPRO Weekend-System II e) Switch off the treatment unit (if necessary at the main switch) min. 24 h The operating agents are on the side Depressurize the tank: cover. Set the toggle switch of the pressure valve on “0”...
Operating instructions ALPRO Weekend-System II 4 Maintenance Filter The water filter should be examined by a dental engineer at least once a year and changed if necessary. Information Tank cap The threads of the tank cap and tank neck should be cleaned from time to time and should be slightly greased with vaseline.
Installation instructions ALPRO Weekend-System II II. Installation instructions (for qualified technical staff only) The procedures described below must only be carried out by suitably qualified technicians for dental devices in accordance with the valid instructions for electrical safety and the drinking water decree.
2.3 Connection to the treatment unit The Weekend-System must only be connected by an expert authorized by ALPRO in accordance with the valid instructions for the drinking water decree. Caution: Unauthorised interference to the treatment unit can Attention! invalidate the manufacturers guarantee.
Installation instructions ALPRO Weekend-System II (included in delivery) which has to be connected to the central water supply of the treatment unit. If necessary, the pipes of the treatment unit to the instruments, to the tumbler filler and to the bowl rinsing are to change so that they are connected together to one pipe.
Installation instructions ALPRO Weekend-System II 2.5 Instructions for installation a) Switch off power supply of the treatment unit using the main switch! b) Close angle valve in the connection box and relieve the water pipe system of the treatment unit of pressure, e.g.
Installation instructions ALPRO Weekend-System II 3.3 Spare parts Article Art. no. Tank cap 80770020 Seal for tank cap 80770023 Filter for tank 80770021 Filter insert 80µ 80770316 Installation kit 8220 3.4 Circuit diagram Check valve Water stop Filter Throttle valve...
Need help?
Do you have a question about the Weekend-System II and is the answer not in the manual?
Questions and answers