URIPHON wireless2M Instructions For Use Manual

5-channel radio alarm for the treatment of bedwetting and for use for incontinence

Advertisement

Quick Links

5-channel radio alarm
for the treatment of bedwetting
and for use for incontinence
Instructions for use

Advertisement

loading

Summary of Contents for URIPHON wireless2M

  • Page 1 5-channel radio alarm for the treatment of bedwetting and for use for incontinence Instructions for use...
  • Page 3 Alarm device for the treatment of bedwetting Description: The new is a wireless radio alarm for the treatment of bedwetting and for incontinence, whose radio receiver can simultaneously manage up to 5 transmitter units (patients). It is ideally suited for use in clinics and care units, as the signals for all incoming alarms arrives to a single station.
  • Page 4 Delivered with: Receiver, Transmitter, 3 absorbent pads with adhesive points, 2 AA type batteries (for Receiver). 1 CR2450 button battery (for the Transmitter), power supply unit, storage bag, detailed manual, brochure "Bedwetting - the latest information for a common problem". Receiver serial number Transmitter serial number Fig.
  • Page 5 Delivery of Instructions for use 3 contact pads with spare adhesive points Power supply unit Transmitter Receiver Button battery CR2450 batteries 2 type AA Fig. 2: Delivery...
  • Page 6 Components - Functions and function indicators - Controls: Transmitter Fig. 3: Transmitter, front Press studs to fix the contact-pads „Pairing Button“ Clips to attach the Transmitter to the underpants Serial number of the Transmitter Fig. 4: Transmitter, back...
  • Page 7 Receiver Beeper LED display for operating mode of Receiver: No display: The device is running on battery and the battery is charged Red LED flashes: The Receiver batteries have to be replaced Green LED: Device runs on power supply Labels, e.g. for names or room numbers LED indicators for the Transmitter and the pairing mode...
  • Page 8 Receiver Battery compartment cover Serial number of the Receiver Fig. 6: Receiver, back with closed battery compartment Battery compartment for two 1.5V AA batteries Connection socket for the operation of the Receiver to the power supply Fig. 7: Receiver, back with open battery compartment...
  • Page 9 Contact pads adhesive points for fixing the contact pads to underpants Press studs for the connection of the contact pads with the Transmitter Fig. 8: Contact pads Power supply plug with cable Power supply plug for connection to 100-240V AC Plug to connect to the receiver (power supply) Fig.
  • Page 10 Starting the device: Important note: Neither Transmitter or Receiver have an on or off button or switch. Once the batteries are inserted in both the Transmitter and the Receiver and the Transmitter has been paired to the Receiver, the unit is ready for use. Depending on the number of alarms triggered their respective durations, after the starting of the device the batteries are functional for several weeks before they have to be replaced.
  • Page 11 Fig. 10b: Receiver, back with open battery compartment Mains supply operation • Connect the power supply unit (fig. 9) with the Receiver. Then connect the power supply unit to a wall socket • Note: Even when using the power supply, it is always recommended to keep batteries inserted in the Receiver.
  • Page 12 Preparing the Transmitter: • On the left side of the transmitter enclosure there’s an opening with a snap in latch behind it and which allows to open the enclosure (Fig. 11a) • With a thin and spiky object (e.g. a paper clip) press against this latch and take off the enclosure cover (Fig.
  • Page 13 Vibration Alarm “On“ Vibration Alarm “Off“ Abb. 11c: Acivate / Deactivate Vibration Alarm Fig. 11d: Inserting the CR2450 button battery...
  • Page 14 Pairing a Transmitter to the Receiver (pairing mode): • (1) press the button on the receiver (fig. 5) for approx. 3 seconds and then release. • (2) a short double beep rings and both LEDs light up for the Transmitter (orange - drop icon and red - battery symbol) of all 5 channels (10 LEDs), for approx.
  • Page 15 is now ready for use. Function test: Perform a function test of the device for each paired Transmitter as follows: • Press the "pairing button" on the Transmitter. For as long as you hold the button, the two LED displays light up on the Receiver (orange - drop icon and Red - battery icon) of the radio channel to which the Transmitter was paired.
  • Page 16 More information on the operation of the unit Operation of with contact pads It is recommended that several pairs of snug-fitting boxers or briefs, in the appropriate size, are available throughout the duration of the treatment. The absorbent contact pads supplied with the device (3 units) are placed in the pants so that the centre of the contact pad is placed in front of the urethral opening.
  • Page 17 • Fasten the transmitter with the fixing clip to the outside of the underpants. Fig. 13: Attaching the Transmitter to the underpants Triggering the alarm: A few drops of urine are enough to trigger the alarm signal. The wearer is woken up (by the staff, depending on where the receiver with the wake-up signal is placed) before large amounts of urine escape from the bladder.
  • Page 18 After having been to the toilet and having emptied the bladder, dry pants and a new contact pad must be worn. Once the connection to the Transmitter is restored, is ready to go again. The patient can go back to sleep. Operating with bed liner (LiSSI®...
  • Page 19 The LiSSI® sensor matte can be purchased via our Web shop (www.uriphon.de) with item number LP0020, as a SET together with with item number WLP0038, from the specialized resellers or through your pharmacy. Operating with several Receivers Especially when operating the device in hospitals or nursing homes, the need arises to signal an acoustic alarm at different locations - as in the room of the patient himself and in the room in which the nursing staff is located.
  • Page 20 Maintenance and cleaning: The contact pads should be checked on a regular basis. To do this, connect them when dry to the device and wet them in the middle of the bottom part with a few drops of water. If the drops of water are immediately absorbed and the alarm is immediately activated, the pads are in perfect condition.
  • Page 21 Safety instructions: Carefully read through the operating instructions and the safety notes before using the product. Always keep the manual within easy reach. If you sell the unit or share it, make sure to hand over these instructions, as well. Never let children use electrical devices unattended.
  • Page 22 The devices are not suitable for operation in the open and should be used only in dry spaces. Never expose the devices to rain or moisture. Should the devices have been subjected to strong differences of ambient temperature, do not put them into operation immediately as the device could be damaged due to condensation.
  • Page 23: Ec Declaration Of Conformity

    Manufacturer: emcotrade GmbH Address: Esselbachstrasse 27 D-75438 Knittlingen Germany URIPHON® wireless 2M Product specification: Class: We hereby declare that the here described medical device complies with the relevant provisions of Directives 93/42/EEC and 2004/108/EC. Adopted standards: EN 55011:2009 + A1: 2010...
  • Page 24 Additional accessories can be purchased in our Web shop www.uriphon.de, at medical supply stores and dealers: 3 set absorbent replacement pads with adhesive points (Uriphon order No. W0012) 3 set absorbent replacement pads with adhesive points (Uriphon order No. W0012) ROE70®...
  • Page 28 Disposal: Do not dispose of electronic equipment in the household waste! Before disposing of the devices, remove the batteries! Return used batteries to a battery collection point! September 2017...