OTIS OI-2100 TOCSIN Operation Manual

Single gas personal monitor

Advertisement

Quick Links

Model OI-2100 TOCSIN Single Gas Personal Monitor
________________________________________
Operation Manual
Revision 2.1w
_______________________________________________________________________________________________
________________________________________________________________

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OI-2100 TOCSIN and is the answer not in the manual?

Questions and answers

Summary of Contents for OTIS OI-2100 TOCSIN

  • Page 1 Model OI-2100 TOCSIN Single Gas Personal Monitor ________________________________________ Operation Manual Revision 2.1w _______________________________________________________________________________________________ ________________________________________________________________...
  • Page 2: Table Of Contents

    Table of Contents Introduction..............................3 Warnings Statements/Avertisseement...................... 4 First Time Activation..........................5 Display Details..................................5 Day to Day Usage............................6 Additional Steps..................................7 Failures/FAQ’s............................8 Alarms................................. 9 Alarm Types..................................9 Alarm Set Points.................................9 Event Log..............................10 “Bump Test Interval” (bUP) - Clearing the alert......................10 O2 Detector-Calibration..........................
  • Page 3: Introduction

    Introduction This document is an Operation Manual containing diagrams and step-by-step instruction for proper operation of the Otis Instruments, Inc. Gen II Model OI-2100 TOCSIN Single Gas Personal Monitor. This document should be read before initial operation of the product.
  • Page 4: Warnings Statements/Avertisseement

    Do not attempt to replace the battery or sensor, this product is designed to be disposable. Changing these • components will void the warranty. If you suspect any malfunction or have any technical problems please contact your local Otis Instruments • representative.
  • Page 5: First Time Activation

    Remaining" user options. WARNING: Users must familiarize themselves with the icons in both non-alarm and alarm states WARNING: If the display is missing icons or cannot be clearly read, please contact your local Otis Instruments representative immediately. Alarm Condition Icon...
  • Page 6: Day To Day Usage

    Day to Day Usage Prior to daily use, the detector will prompt the user to perform a self test. This process is a simple and effective way to ensure safe operation of the detector. During the “Self Test” the audio, visual, and vibrating alarms are activated and the sensor is tested.
  • Page 7: Additional Steps

    Additional Steps In the previous section, the most common screen configurations were outlined. If the unit has been programmed via the IR Link or has been exposed to gas additional, though, the following screens may appear. Screen Sequence # Detail Screen 4 If programmed with a “User ID”, after the alarm set points are displayed a combination of numbers and or letters will...
  • Page 8: Failures/Faq's

    If the self test fails, the detector emits five short beeps and flashes before displaying “Test”. Repeat the self test. • If the self test fails three (3) consecutive times the detector will enter a Fail Safe mode. Please contact Otis •...
  • Page 9: Alarms

    Alarms Alarm Types Screen Display Detail Low Alarm: Audible Alarm = one slow beep every second Visual Alarm = one slow flash every second Vibrating Alarm = one slow vibration every second High Alarm and Over Limit (OL) Alarm: Audible Alarm = two fast beeps every second Visual Alarm = two fast flashes every second Vibrating Alarm = two fast vibrations every second Detector Life Countdown Alarm (EOL)
  • Page 10: Event Log

    Event Log By default the detector stores the last twenty five (25) alarm events. The system stores events by first in first out, i.e. the event will replace the first event and so on. This information can be downloaded using the OI-TOCSIN-LINK. For each alarm event the detector records the following: The detector serial number •...
  • Page 11: O2 Detector-Calibration

    O2 Detector-Calibration NOTE: Otis Instruments recommends users of the OI-2100-O2 bump test the detector before using each day. OI-2100-O2 detector factory default will prompt the user to calibrate the detector every 30 days. The user will be prompted by the screen flashing CAL, please see calibration instruction below:...
  • Page 12: Specifications

    Specifications Size: 3.3 in H X 2.0 in W X 1.1 in D/ 85mm X 50 mm X 28mm Weight: 76 g / 2.7 oz Humidity: 5% to 95% non-condensing relative humidity Alarms: Visual, vibrating, audible (minimum 95dB) Battery Life: 24 months of operation/ 2 minutes of alarm per day.
  • Page 13 Otis Instruments products. This warranty extends only to the sale of new and unused products to the original buyer. Otis' warranty obligation is limited, at Otis Instruments' option, to refund of the purchase price, repair, or replacement of a defective product that is returned to a Otis Instruments authorized service center within the warranty period.
  • Page 14 Otis Instruments, Inc. 2200 E. Villa Maria Dr. Bryan, TX 77802 Service Department: 903.566.1300 Corporate Office: 979.776.7700 service@otisinstruments.com www.otisinstruments.com...