Model 954 4-Channel MFC Power Supply/Controller INSTRUCTION MANUAL October Sierra Instruments, Inc., Headquarters Sierra Europe, European Headquarters Sierra Asia, Asia-Pacific Headquarters...
All pertinent data, required by the microprocessor at power-up to re-initialize the system, is also stored in the same RAM. The Model 954 utilizes a 4-line by 20 character back lighted LCD display. A built in Totalizer, for each channel automatically recognizes the units of measure selected and adjusts the time base for the integrator accordingly.
FRONT PANEL MENU SELECT PURGE CLOSE ENTER ANNUNCIATORS: Displays OVERRIDE signal status of each channel. If annunciators are not illuminated, the Setpoint (Control) voltage is active. DISPLAY AREA Column 1: Reserved for displaying (*) Active Channel, (M) Master Channel, (S) Slave Channel or (T) totalizer. Column 2: Reserved for polarity indicator (minus sign for negative signal, none for positive).
REAR PANEL ANALOG OUT RS485 ALARMS T250 1A CHANNEL 2 RS485 CHANNEL 1 CHANNEL 3 CHANNEL 4 RS232 POWER INLET WITH FUSE (1 amp TIME DELAY or SLO BLO) POWER SELECTOR SWITCH RS232 SERIAL PORT (J6) TRANSDUCER CONNECTORS (J1, J2, J3, J4) ANALOG OUTPUT (J5) ALARMS (J8) RS-485 SERIAL PORT (J7, J9)
Page 7
CONNECTOR PIN DESIGNATIONS TRANSDUCER CONNECTORS (J1, J2, J3, J4) SIGNAL COMMON SIGNAL INPUT GROUND VALVE OPEN GROUND -15Vdc SETPOINT SIGNAL GROUND Transducer Connector (Female) GROUND Rear Panel View VALVE OFF +15Vdc CHASSIS GROUND ANALOG OUTPUT (J5) SIGNAL CH 1 SIGNAL COMMON CH 1 SIGNAL CH2 SIGNAL COMMON CH2 Analog Output Connector (Female)
(one per channel) on the Model 954 PCA (inside the housing). Please refer to the PCA schematic in Appendix A for the location of these jumpers. All Model 954 units come factory set for 15 Vdc valve circuit operation (unless ordered new with a 30 Vdc instrument).
START-UP The Model 954 has a 100 Vac, 115 Vac or 230 Vac, 50/60 Hz power selector switch located at the rear of the instrument. Please refer to page 4 to locate this switch. Verify the power selector switch is in the proper position prior to connecting the power cable to the unit.
Page 11
MANUAL SETUP, Selecting Valve Override (Open, Close or Run) on page 9. The default at power-up is Valve Close. 16. If the Units of Measure are in flow units, The Model 954 automatically calculates TOTAL flow using a Riemann Sum Integration method. To display TOTAL or to reset the TOTAL display, refer to MANUAL/SETUP, Selecting Display (Flow or Total) on page 9.
MFC Setpoint control is activated. If the valve override does not appear to function correctly, you may need to re-configure The Model 954 using the 15/30 VDC jumpers on the PCA. Refer to page 6B to verify valve circuit configuration of The 954 is correct for your MFM or MFC.
Response: "MULTIDROP ADDRESS: 01" *00x<aa> Set 954 Address to aa Note: All Model 954's will respond to * 0 0 X. To prevent bus Example: Send * 0 0 x 2 2 contention, connect only 1 Model 954 to the RS485 Model 954 will respond with a "spade"...
Page 22
*aaUM4<dd> Set CH2 Unit of Measure at Address 02 to selection dd Example: Send *03UM366 Model 954 with Address 02 will have CH3 Unit of Measure set to PSI Setting Gas Idenfifier Reading Gas Identifier RS232 Command: RS232 Query: GS1<ddd>...
Page 23
Set CH4 Signal Input at Address aa to selection d *aaIN4 Response: IN4<>d<>zzzzz Example: Send *01IN31 Example: Send *10IN2 Response: IN2 1 0-5V Model 954 with Address 01 will have CH3 Signal Input set for 0-5V. Setting Filter Reading Filter RS232 Command: RS232 Query: FL1<d>...
Set CH3 Display at Address aa to selection d *aaD4d Set CH4 Display at Address aa to selection d *aaD2 Response: D2d Example: Send *03D12 *aaD3 Response: D3d Model 954 with Address 03 will display FLOW on *aaD4 Response: D4d PAGE 21...
Reset CH3 Total at Address aa to zero *aaIT4R Reset CH4 Total at Address aa to zero Example: Send *01T1R Model 954 with Address 01 CH1 Total reset to zero Reading Range Value Setting Range Value RS232 Query: RS232 Command: Response: SN1d.dddd...
UNITS OF MEASURE TABLE Description Abbrev Total Standard Cubic Centimeters per Minute SCCM Standard Liters per Minute Percent Volts Millivolts Counts Normal Liters per Minute Standard Liters per Second Normal Liters per Second Standard Liters per Hour Normal Liters per Hour Standard Milliliters per Minute SMLM Normal Milliliters per Minute...
Page 30
Appendix A Model 954 Internal Circuit Assembly Board per channel...
Page 31
Appendix A Model 954 Display Assembly FRONT Zero Adj Zero Adj Zero Adj Zero Adj 10V Range 10V Range 10V Range 10V Range (CH4) (CH3) (CH1) (CH2) FS Adj FS Adj FS Adj FS Adj 10V Range 10V Range 10V Range...