deconta aircontrol S Series Instruction Manual

Negative pressure monitor 110v

Advertisement

Quick Links

aircontrol S 2
Manufacturer:
Description / Type No.:
Instruction manual
Negative Pressure Monitor
aircontrol S 110V
deconta GmbH
Im Geer 20, D - 46419 Isselburg
aircontrol S 1 110V
aircontrol S 2 110V
aircontrol S 3 110V
aircontrol S 1
aircontrol S 3
Type 608
Type 603
Type 609

Advertisement

Table of Contents
loading

Summary of Contents for deconta aircontrol S Series

  • Page 1 Negative Pressure Monitor aircontrol S 110V aircontrol S 1 aircontrol S 2 aircontrol S 3 Manufacturer: deconta GmbH Im Geer 20, D - 46419 Isselburg Description / Type No.: aircontrol S 1 110V Type 608 aircontrol S 2 110V Type 603...
  • Page 2: Table Of Contents

    Instruction manual aircontrol S 110V Table of contents on page Introduction Basic safety advice Intended use Operation Transport Delivery Transport Scope of delivery Technical description Technical data Technical data Basic equipment and possible upgrades Terminal assignment potential-free alarm contact Initial operation Telealarm GSM, inserting a Sim-Card Standby function Display-indication...
  • Page 3: Introduction

     Mechanics with „kick“  Pleasing design The copyright of this instruction manual remains with deconta. This manual is intended for assembly, operation and maintenance personnel. It contains instructions and drafts of technical nature which may neither be distributed nor used in any unauthorised way for competitive purposes or passed on to others.
  • Page 4: Basic Safety Advice

    2.1 Intended use As an operator, you are obliged by deconta to follow the instruction manual and to use this engineered technology equipment only in accordance with the regulations and its suitability! In the event of non-observance, deconta assumes no liability.
  • Page 5: Transport

    3.1. Delivery Unless agreed otherwise, the lock system is packed completely and securely for transport by deconta. Transport damages have to be documented at once during the handing over of the carrier or another supplier. Please note the possible damages additionally on the way bill.
  • Page 6: Technical Description

    Instruction manual aircontrol S 110V 5 Technical description The aircontrol S is designed to measure, document and react to alarm conditions in regard to the Negative Pressure Monitoring. The devices can be equipped with individual extension modules according to customer requirements.
  • Page 7: Technical Data

    Instruction manual aircontrol S 110V 6 Technical data 6.1 Technical data Measurements in mm (L x W x H) 305 x 350 x 177 475 x 350 x 177 495 x 350 x 187 Weight 5,5 kg 7,9 kg 8,4 kg Power connection 100-120 V / 16 A 100-120 V / 16 A...
  • Page 8: Terminal Assignment Potential-Free Alarm Contact

    Instruction manual aircontrol S 110V 6.3 Terminal assignment dry alarm contact (only aircontrol S 1 und S 2) In normal condition: contact 1 and 2 open, contact 1 and 3 closed In alarm condition: contact 1 und 2 closed, contact 1 and 3 open Without main voltage: contact 1 and 2 closed, contact 1 and 3 open 7 Initial operation...
  • Page 9: Telealarm Gsm, Inserting A Sim-Card

    Instruction manual aircontrol S 110V Important: Rooms adjoining to the area of operations have to be chosen as reference measuring points. A configuration of the reference measuring point can be cancelled, if the Negative Pressure Monitor is placed at the location of the reference measuring point. Attention: Never use the Negative Pressure Monitor in the contaminated work area! 7.1 Telealarm GSM, inserting of Sim-Card (only aircontrol S 3)
  • Page 10: Standby Function

    Instruction manual aircontrol S 110V 7.2 Standby Function (only aircontrol S 3) The Standby Function is activated via a Code-Entry (Code-chart aircontrol, Code 60). As soon as the Negative Pressure Monitor aircontrol notices a negative pressure in the removal area that is too low (constant alarm via X seconds, adjustable at aircontrol), an integrated Standby-mode is activated.
  • Page 11: Display-Indication

    Instruction manual aircontrol S 110V 8 Display-indication 8.1 aircontrol Time Indication for inserted options (e.g. battery) Device status Alarm status Dynanometer(1 – 8) Measuring value After switching on the aircontrol there are 3 information displayed in the top line:  Top left => device status Possible status indications (S followed by 3 numbers): First digit 1 to 8 Number of the present measuring channels...
  • Page 12: Telealarm

    Instruction manual aircontrol S 110V 8.2 Telealarm (only aircontrol S 3) Charging indicator Device status (battery) Alarm status Network operator GSM- field strength After switching on the aircontrol there are 3 information displayed in the top line:  Top left => Device status Possible status indicators (S followed by 3 numbers): S500 Sim-Card is missing...
  • Page 13: Settings / Operation

    Instruction manual aircontrol S 110V 9 Settings / Operation Adjustments are always carried out according to the same procedure via the keypad:  Push the key # => C will appear on the display  Enter the code for the relevant adjustment (see following Code-chart) ...
  • Page 14: Possible Settings / Code Chart Aircontrol

    Instruction manual aircontrol S 110V 9.1 Possible adjustments / Code-chart aircontrol Code Function Value 0 = inactive Activate / Inactivate Unit 1 = active Alarm low channel 1 0 - 290 Pa Alarm low channel 2 0 - 290 Pa Alarm low channel 3 0 - 290 Pa Alarm low channel 4...
  • Page 15: Possible Settings / Code Chart Telealarm

    Instruction manual aircontrol S 110V 9.2 Possible adjustments / Code-chart Telealarm (only aircontrol S 3) Code Function Value Recording of message (record text via microphone, max. 20 seconds) Listen to announcement 0 = inavtive Activate / inactivate unit 1 = active Phone number 1 Phone number 2 Phone number 3...
  • Page 16: Extension Modules

    Instruction manual aircontrol S 110V 10 Extension modules aircontrol S 1 offers space for 4 extension modules, aircontrol S 2 offers space for 10 extension modules and aircontrol S 3 offers space for 5 more extension modules. 10.1 Additional measuring channels For the aircontrol S 1, 4 more measuring channels can be upgraded or managed, in addition to the existing measuring channel For the aircontrol S 2, 7 more measuring channels can be upgraded or managed, in...
  • Page 17: Emergency Battery

    Instruction manual aircontrol S 110V  When the top of the paper comes out, release the green key.  Insert the top of the paper through the slot of the printer cover  Shut the printer cover Changing of ink ribbon: ...
  • Page 18: Measurement Data Memory Module

    Instruction manual aircontrol S 110V 10.4 Measuring data-memory module (from Software version 1.3) Saving of the measuring data (max. 90000 Data sets) with adjustable memory intervals. Export as txt-Data for further treatment via USB-Stick. The USB-Stick must be formatted as FAT or FAT 32. If the maximal capacity of memory is exceeded, the oldest data sets are overwritten.
  • Page 19 Instruction manual aircontrol S 110V The file containing the saved data can be edited freely using a text editor. Example: Project: Munich Bauleitung: Mr. Mustermann Measuring device: aircontrol S2 Type: 481 Series: 583 Removal period: 27.04.2010 08:00 to 29.04.2010 08:29 Channel 1: Working area 1 Setting: min.
  • Page 20: Declaration Of Conformity

    Instruction manual aircontrol S 110V 11 Declaration of conformity EU Declaration of Conformity deconta GmbH Im Geer 20 D-46419 Isselburg Product: Negative Pressure Monitor Type: 603, 608, 609 The design of the units complies with the EU- Machine directive2006/42/EG following directives:...