Contents 1.Introduction 2.Technical Specifications 3.Names and Functions of Various Parts for UTG‐2650 4. Operation 4.1 Calibration 4.2 Measuring 4.3 Saving Test Results 4.4 Reading Saved Results 4.5 System Setting‐Change Resolution and/or Units 4.6 Minimum Capture 4.7 Two Point Calibration 5. Methods of Thickness Measurements 5.1 Measuring planar surface 5.1.1 General methods 5.1.2 Method of precision measurement 5.1.3 Method of continuous measurements 5.2 Methods for measuring pipe walls 6. Maintenance 7.Special Applications & Couplant Gels 8. Standard Package Contents...
1. Introduction The UTG‐2650 is a simple to use, highly accurate, hand‐held Ultrasonic Thickness Gauge with the ability to measure ultra‐thin materials in super high resolution. Utilizing it’s state of the art microprocessor and ultrasonic technology the 2650 offers you many outstanding features such as quick scan and extended memory. It features reliable and stable readings, standard or dual element measuring modes, convenient data display (in both millimeters and inches),high resolution .0001”...
Page 4
3. Names and Functions for UTG-2650 ● Power on/Reset key ON/RST –to turn on or reset the tester. When the tester is turned on, LCD will display the value of sound velocity used for the last measurement. The gauge will shut off automatically if idle for 4 minutes.
4.Operation 4.1Calibration Set the sound velocity to be 0.232”(5900m/s) .Put a pea sized drop of couplant gel on calibration block built-in on the gauge. Press the probe on the face of the block to obtain measurement .LCD screen will display a value. Lift the probe off the block.
4.6 Minimum Capture (Scan Feature) If this function is set to “ON” from system setting, it is ready to capture minimum reading in the measurement. LCD will show the current thickness while the probe is coupled with the sample .as soon as you release the probe from your part, the LCD displays the minimum reading and “MIN”...
5.1.2Method of Precision Measurement This method involves increasing the number of measurement spots around a certain measurement point and expressing the thickness fluctuation in terms of isothickness contour lines, fig.3. Fig. 3 Fig 3. 5.1.3Method of continuous measurements Fig.4 As in fig.4, this method involves taking continuous measurements along a specified line according to the single measurement method, at intervals of 5mm or less.
Fig.6 Pipe sample Fig.7 Bottom view of probe 6.Maintenance Be sure to clean the probe and cable after each use. Grease, oil and dust will cause the cable of the probe to dry out and shorten life expectancy. If the unit is not to be used for a long period of time, remove the batteries to avoid battery leakage and corrosion of the battery contacts.
TWICE what it should be. The Responsibility for proper use of the instrument and recognition of these types of phenomenon rests solely with the user of the instrument. 8.Standard Package Contents Items UTG-2650(main gauge unit) Probe Coupling gel Operation Manual...
Page 11
Main Headquarters: U.S.A Phase II Machine & Tool, Inc. 21 Industrial Ave Upper Saddle River, NJ. 07458 USA Tel: (201) 962‐7373 Fax: (201) 962‐8353 General E‐Mail: info@phase2plus.com BEIJING, CHINA Phase II Measuring Instruments (Beijing) Ltd. Room 301, Bldg 2 Qing Yuan Xi Li, Haidian District, Beijing 100192,China Tel:+86‐10‐59792409 Fax:+86‐10‐59814851 General E‐mail: info@phase2china.com.cn MEXICO Phase II de Mexico Calle A No. 4 Promer Piso Col. San Marcos Azcapotzalco C.P 02020 Mexico Tel: 011‐525‐5538‐39771 Fax: same General E‐mail: phase2mexico@hotmail.com VENEZUELA Phase II Herramientas Universales EDCM. CA. Av. Francisco Lazo Marti CC Plaza Santa Monica PB Local Santa Monica, Caracas 1040 Venezuela Tel: 212‐690‐28‐21 Fax: 212‐693‐29‐16 E‐mail: edcphm@movistar.net...
Need help?
Do you have a question about the UTG-2650 and is the answer not in the manual?
Questions and answers