As the User Guide has been arranged in independent chapters, some of the descriptions have been duplicated. If you encounter any problems during operation of the equipment, stop its use immediately, and contact the customer service center of REMED (Refer to chapter 11). Copyright This User Guide is protected by the Copyright Act.
ELPIS-4 / 3 / 2 User Guide There may be changes. These contents are subject to change without notice. Contents ········································································································ 6 1. Information on safety 1.1 Definitions of symbol ······································································· 6 1.2 Requirements for safety ··································································· 6 1.2.1 Precautions for electrical safety ················································· 6 1.2.2 Classifications ········································································...
Page 4
ELPIS-4 / 3 / 2 User Guide 6. Operation instructions ···························································································· 21 6.1 How to connect each mode ···························································· 22 6.1.1 Bipolar connection method ····················································· 22 6.1.2 Monopolar connection method ················································· 23 6.2 How to use Patient Plate ································································ 24 6.3 How to use each mode ··································································...
ELPIS-4 / 3 / 2 User Guide Information on safety 1.1 Definitions of symbols ❑ When there is specific information in the User Guide that needs to be emphasized for safety, the following terms and symbols are indicated. All warnings and precautions shall always be observed.
ELPIS-4 / 3 / 2 User Guide • Installed equipment shall be subject to regular safety inspections from specialized personnel authorized by the manufacturer. • For electrical safety, equipment shall always be connected to a safely grounded power supply for operation.
Page 8
ELPIS-4 / 3 / 2 User Guide ❑ As a fan is installed in the body of the equipment for air circulation, remove curtains or any other objects which can block the flow of air from near the equipment. ❑ From the surroundings of the location where equipment is installed, remove water, alcohol, flammable materials, and so forth.
ELPIS-4 / 3 / 2 User Guide ❑ If the contact and connector connection is poor, the patient may be burned due to output error. Be sure to check before use. ❑ An electrical spark from an electric scalpel may cause a fire or an explosion if flammable anesthetic gas, high oxygen concentration or self-contained flammable gases are present.
ELPIS-4 / 3 / 2 User Guide 1.5 Abnormal reaction ❑ During the operation of the equipment, if the patient experiences any abnormal symptoms, the operator must stop the treatment immediately, and contact the doctor in charge. 1.6 General warnings ❑...
ELPIS-4 / 3 / 2 User Guide Electrosurgical Unit 2.1 Mechanism [Figure 01] Monopolar electrosurgical An electrosurgical unit is a medical device that is used to reduce bleeding in a portion of the body tissue and to dissect or stop the bleeding from the bleeding site. The device consists of three components: main body, electrode, and patient plate.
ELPIS-4 / 3 / 2 User Guide method of placement of the patient plate. In general, the Patient Plate should be close to the treatment site to keep the electrical circuit as short as possible. (2) Bipolar Electrosurgery In bipolar electrosurgery, active electrode and return electrode functions are performed through bipolar forceps.
ELPIS-4 / 3 / 2 User Guide tissue, resulting in hemostasis. You can control the degree of blood and incision by changing the duty cycle. [Figure 03] Electrosurgical waveform 2.2 Objectives This device is an electrosurgical unit. It is prohibited to use it for purposes other than the intended use.
ELPIS-4 / 3 / 2 User Guide 3.2 Electrical installation condition ❑ Input power : 220 VAC / 60 Hz ❑ Power consumption 1. ELPIS-4: max 1,000 VA 2. ELPIS-3: max 950 VA 3. ELPIS-2: max 800 VA 3.3 Environmental condition 3.3.1 Application environment...
ELPIS-4 / 3 / 2 User Guide socket. If the connection is incorrect, unexpected problems can occur. ❑ Grap the plug point when you unplug the power cable. ❑ Do not connect several devices to one power outlet. ❑ In particular, do not use a power cable that generates connection noise.
ELPIS-4 / 3 / 2 User Guide 4.2 Control unit [Figure 08] Device control section 1. Power button – Power On / Off 2. Mono Cut screen – Monopolar cut function, level 3. Mono cut function selection button – Pure, Blend1, Blend2, Blend3 4.
ELPIS-4 / 3 / 2 User Guide Preparations for operation 5.1 Moving and fixing equipment ❑ Before moving the equipment, remove the connected power cable and the connection cables for peripheral equipment. ❑ Do not apply excessive shock or vibration when moving the unit.
ELPIS-4 / 3 / 2 User Guide 5.5 Check for mode use Make sure that each mode and function is selected through the touch screen and that the output strength is adjusted normally. 5.5.1 Bipolar Mode check (1) On the touch screen, press Mode select button of Bi Coag mode and select Foot Soft or Foot Forced.
Turn on the power switch located on the back of the unit. As shown in the figure below, the orange power button will blink, the Remed logo will appear and disappear, and the orange power button will light. This state is the power standby state.
ELPIS-4 / 3 / 2 User Guide 6.1 How to connect each mode 6.1.1 Bipolar connection method • Do not connect a wet accessory to an electrosurgical unit. • Ensure that all accessories are properly connected and never use them if the connection cable cover is damaged.
ELPIS-4 / 3 / 2 User Guide 6.1.2 Monopolar connection method • Do not connect a wet accessory to an electrosurgical unit. • Ensure that all accessories are properly connected and never use them if the connection cable cover is damaged.
ELPIS-4 / 3 / 2 User Guide (2) Footswitching Connect the Monopolar Footswitching Pencil and Patient Plate to the connections on the front of the unit and the Double Foot Switch to the back of the unit as follows. [Figure 14] Monopolar connection method - Footswitching 6.2 How to use Patient Plate...
ELPIS-4 / 3 / 2 User Guide 6.3 How to use each mode 6.3.1 Bipolar mode When the unit is in standby mode, press and hold the power button for 2 to 3 seconds, the unit will turn on and the power indicator will change from the orange standby indicator to the purple indicator.
Page 26
ELPIS-4 / 3 / 2 User Guide ② Connect Bipolar Forceps to the Bipolar Forceps connection on the front of the unit (see Figure 10). ③ Use the Up / Down arrows to select the output strength. Use the output as you need for surgery.
ELPIS-4 / 3 / 2 User Guide ④ Use the Up / Down arrows to select the output strength. Use the output as you need for surgery. If you do not know the proper output strength, you should increase the output strength from the lowest output strength '1'.
ELPIS-4 / 3 / 2 User Guide Incision and hemostasis are possible at the same Pure Cut is used for time. As Blend 1 → Blend 3, the deepening of the hemostasis. bleeding effect occurs, while the incision is slow.
Page 29
ELPIS-4 / 3 / 2 User Guide ③ Make sure that the Patient Plate connection / use indicator does not blink and is lit. If it is blinking, make sure that the Patient Plate is connected or that the Patient Plate is correctly attached to the patient's body.
Page 30
ELPIS-4 / 3 / 2 User Guide ④ Connect the Monopolar Handswitching Pencil to the Monopolar 1 or Monopolar 2, which is the Monopolar Handswitching Pencil connection on the front of the machine ⑤ Use the Up / Down arrows to select the output strength. Use the output as you need for surgery.
Page 31
ELPIS-4 / 3 / 2 User Guide ⑤ Connect the Monopolar Footswitching Pencil to the Monopolar Footswitching Pencil connection on the front of the unit (see Figure 13). ⑥ Use the Up / Down arrows to select the output strength. Use the output as you need for surgery.
Page 32
ELPIS-4 / 3 / 2 User Guide ⑤ Connect the Monopolar Footswitch to the Monopolar Footswitch connection on the back of the unit (see Figure 14). ⑥ Use the Up / Down arrows to select the output strength. Use only as much output as you need for surgery.
7. Messages 7.1 REM system The ELPIS-4, 3, 2 Checks the patient's electrical contact with the Patient Plate to reduce the risk of burns. Through the REM system, the output is interrupted and the operation is stopped in the following cases.
ELPIS-4 / 3 / 2 User Guide 8.ELPIS-4, 3, 2 relating journal papers Electrosurgery: history, principles, and current and future uses. J Am Coll Surg. 2006 Mar;202(3):520-30. Massarweh NN1, Cosgriff N, Slakey DP. Abstract Within the surgeon’s armamentarium, electrosurgical devices stand out as some of the most useful and most used units.
Guidance and manufacturer’s Declaration – Electromagnetic Emissions The ELPIS-4, 3, 2 System is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in such an environment Electromagnetic environment –...
Page 38
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The ELPIS-4, 3, 2 System is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in such an environment.
Page 39
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ELPIS-4, 3, 2 System is used exceeds the applicable RF compliance level above, the ELPIS-4, 3, 2 System should be observed to verify normal operation. If...
Page 40
The ELPIS-4, 3, 2 System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ELPIS-4, 3, 2 System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ELPIS-4, 3, 2 System as recommended below, according to the maximum output power of the communications.
ELPIS-4 / 3 / 2 User Guide Maintenance 11.1 Cleaning procedure ❑ Cleanliness of the ELPIS- and accessories can be maintained through various 4, 3, 2 methods. Avoid damage or contamination of the equipment using the following recommended methods. ❑ Periodically clean the equipment’s exterior surface and touch screen with a soft cloth moistened with alcohol.
ELPIS-4 / 3 / 2 User Guide Keep out of water Keep away from direct sunlight - Do not store near heaters - Avoid locations subject to excessive shock or vibration, exposure to chemicals or explosive gases. 11.4 Troubleshooting If the equipment does not operate normally during use, please check the items listed in the table below before requesting service.
ELPIS-4 / 3 / 2 User Guide In the following cases, stop the operation by cutting off the power to the equipment, and contact the service center. → When the strength of output of set function is increased, no heat is generated in electric scalpel →...