Metal sheath standard platinum resistance thermometer
Table of Contents

Advertisement

5680/5682/5689 SPRT
User's Guide
22
info@Fluke-Direct.com
Problem
Solution
Temperature readout different
Measure the SPRT resistance at TPW.
than expected, e.g. the heat
source is set at 300°C, the SPRT
measures 275°C
Fluke-Direct
If the resistance of the SPRT is less than the rated resis-
tance, there may be a short in the sensor. Contact an
Authorized Service Center.
If the resistance of the SPRT is only a few ohms, there
may be a short in the four lead-wires. Contact an Autho-
rized Service Center.
If the SPRT is open, the resistance will be "Out of Limits"
or in the kilo-ohm or mega-ohm range. Contact an Au-
thorized Service Center.