1. Color touch screen 2. Support for stylus and multifunction electrodes 3. Aluminum Casing 4. Printer paper Exit (xCell Pro version only) 5. Start Button 6. Printer (xCell Pro version only) 7. Power switch 8.
Page 9
MODES PROGRAMS HF TECHNIQUES CUSTOMIZATION PicoFlash 22 regions (11/male - 11/female) MeloFlash 10 sizes of hair 3 types of probes (INOX, ITH, IBL) MultiPlex 5 sizes of probes Synchro Customizable programs (90/user) BLEND TECHNIQUES Synchro option PicoBlend CLIENT RECORDS MANAGEMENT OmniBlend 1000 client files MultiBlend...
Page 10
What’s in the box? Before installing and turning on your Apilus xCell, we suggest that you make sure that there is no visible damage that may have been caused during OTHER FEATURES shipping. If the box is damaged, contact the carrier for any claim.
Connect the bar or plate electrode to the other Install your Apilus xCell on a solid table of about 86 end of this cable. cm (34 inches) high for easy access to the screen.
25 microprocessor and resume regular operations. illustrated on your left. However, if you have the xCell Pro version, you will first need to select your user name and enter your password. The "i" icon will open a help page that can explain the functions and features that are avai- lable.
Page 14
Time/Date Language/Volume Screen Configuration Software Version: x.x Clinic/Users Rates Invoice Manage user interfaces, passwords, and admin access Apilus xCell Insertions count Statistics on the treatments Phone (cumulative) 418-555-1234 Visit count Reset button for statistics E-mail apilus@dectro.com Average time 00 h 07 m...
(01-0507) and the timer will continue counting seconds even To add an option to your Apilus xCell unit, first if there is a period of inactivity. press the Option ( ) button, located at the bottom left in the Configuration menu.
C H A P T E R 2 Epilate! CHOOSING A hair follicle, while avoiding heat diffusion into the surrounding tissues. It allows for the targeted des- TREATMENT MODE truction of cells responsible for the regrowth of all hair types and delivers a level of comfort that will To choose the treatment mode, simply press on impress your customers.
Page 19
The OmniBlend is a proprietary technique to zero. Therefore it is not recommended to use Blend the Apilus xCell. Simultaneously, it applies one combined currents in such cases. It is also sug- 27.12 MHz thermolysis current of low intensity with gested to avoid applying the Blend combined cur- a galvanic current.
Page 20
Thermolysis Combined Beauty and 27.12 MHz currents Post-treatment Care PicoFlash PicoBlend Cata + Ana - MultiPlex OmniBlend MeloFlash MultiBlend Synchro EvoluBlend SynchroBlend Current range to use depending on the treatment INTENSITY LENGHT OF TREATMENT REGION (MA) TREATMENT Face-Neck 0.25 - 0.40 Anaphoresis 1-3 min Body-Legs...
Page 21
NOTE : ted. ments, first verify that the electrode handle is connected to the black terminal of the Apilus xCell If necessary, a lotion soaked cotton compress can (or that the VitaPhase IM is connected to the Vita- be placed on the area to be treated so as to not Phase port on the right side of the unit) and that the apply the active electrode directly onto the skin.
Page 22
A | 3 INO Probe and probeholder Unlocking of the Settings With/Without Pedal Enabled Files Client file Adjustment Arrows (xCell Pro version only) Starting HF/Glv. currents Home Page MultiBlend 0.00 00.00 1 5h 18 Date and Time Slow Thermolysis FastThermolysis...
ADJUSTING THE SETTINGS Units of Lye in Blend/Galvanic The energy deployed into the follicle according to To modify your settings, you must first select the the settings used in Blend or Galvanic. See chart desired mode you wish to use and press the below.
Page 24
Cancel the selected program User programs (xCell Pro version only) Female Programs for women Chest Selected region Numbers that relate to the size of the hair, press the OK button to confirm the selection and return to the work screen...
Page 25
( ) button that has an arrow going to client’s file, to bill or debit time off the Apilus xCell the right. The device will then display the working smartcard.
Page 26
A | 3 INO Probe and probeholder Locking and Unlocking of the Settings Enabled With/Without Pedal Files Client file (xCell Pro version only) Adjustment Arrows Starting HF/Glv. currents Home Page MultiBlend 0.00 00.00 15 h 18 Date and Time December 11th...
Page 27
Settings Synchro (Time, intensity, number of impulses, etc.)To modify Transit time and intensity setting of pico-impulses, the settings, drag the circle along the arrow shaft time/delay factor. (This factor allows you to choose from top to bottom. You can also press directly on the selected length of each of the pico-impulses the arrow up or down to increase or decrease the and the thermal relaxation time between each pico-...
Page 28
Region Programs by region A | 3 INO Probe and probeholder Enabled With/Without Pedal Files Client file (xCell Pro version only) Home Page Cata + 1 5 h 18 Date and Time December 11th Chronometer and Timer 00:00:00 1.06 mA...
Page 29
MultiBlend Transit time and intensity setting of galvanic cur- rent, the intensity of thermolytic heating (the com- puter automatically calculates the intensity of heating according to the chosen settings for galva- nic current - actual value is displayed); transit time and impulse intensity for thermolysis at end of the treatment.
CONTRAINDICATIONS General Contraindications to direct current: - Paralysis TO TREATMENT - Pregnancy Indication: - Multiple sclerosis This instrument is designed to perform professional hair - Pacemaker removal treatment by application of RF and/or galvanic current. - Epilepsy General Contraindications: - Cancer - Epilepsy and seizures Local Contraindications to direct current: - The first 3 months of pregnancy...
An insertion into a hair follicle must then be per- The Apilus xCell will graphically display the degree formed. The current intensity gradually increases of hydration on the screen. In addition, it also clas- as you press on the pedal.
Page 34
Create a new File Existing files Client Folder Print client files Appointment Calendar 1 de 1 Davis, James File Search Adams, Laura (418-555-1234) Modify a client file and a medical record Statistics Print a client file Open a client file To activate a condition in the medical profile, press the X next to the...
Allows you to create a new client record. During 1, section titled: configuration). Next, you will need the creation of a client file, the Apilus xCell will ask to open a Client file (see Chapter 2 and 4, section if you would like to complete a medical profile at titled: client file) so that the Apilus xCell device will this time.
EARNINGS REPORT ned on the device). This code will be recorded on Your Apilus xCell device can print X and Z earnings each chip card sold and will ensure that the cards reports. This function is very useful to help you will be only valid for your Clinic/Salon.
Page 38
Statistics of the card inserted Overall statistics for all cards into the card reader Current card Total cards Insertions count Insertions count Visit count Visit count Average time Average time 0 h 00 m 0 h 00 m Total time Total time 0 h 00 m 0 h 00 m...
TRANSFERS AND UPDATES slot located on the right side of the device’s front panel. The Apilus xCell will ask you to associate the Thanks to its USB port, the Apilus xCell device card to an existing customer file and to choose the allows for program updates, to backup the device’s...
MAINTENANCE Accessories The bar electrode or plate electrode should be clea- Your Apilus xCell epilator requires minimal main- ned after every use with an alcohol wipe. The tip of tenance. To protect the original finish, clean it up the brass probeholder can be cleaned with a small with a soft cloth and a mild detergent.
Use in a professional quiet, clean and WARNING: device that uses advanced computer technology. dry environment. Avoid plugging the Apilus xCell into an electrical This device is not sterile and does not WARNING: circuit that is loaded with other high energy consu-...
Page 44
• Do not place this device in a dusty location. The most effective way to prevent your Apilus xCell • Never replace the fuses with others that have a device from interfering with the operation of other different amperage.
Page 45
Legend of Symbols SYMBOLS - REFERENCES - DESCRIPTIONS/DEFINITIONS Corrugated Packaging Alliance Recyclable packaging EN 980-4.11 Manufacturer European conformity marking according to directives 93/42/ European Council Directive 93/68/EEC EEC, 94/62/EC, 2002/96/EC et 2011/65/EU The equipment is subject to a selective collection at the end of European Council Directive 2002/96/EC its life IEC 60417-5004...
CUSTOMER SERVICE Dectro International Your Apilus xCell carries a limited warranty against 1000, boulevard du Parc-Technologique manufacturing and material defects. Please consult Québec (Québec), Canada G1P 4S3 the warranty enclosed with the unit for more details. Telephone: 418.650.0303 A troubleshooting guide is available to help resolve Toll free: 1 800.463.5566...
WARRANTY 8. Shipping and delivery are not covered under this warranty. In the case of manufacturing defects, 1. Your warranty covers the device’s electronic the owner must ship the device, prepaid to their components for a period of thirty-six (36) months nearest authorized service center.
If a problem occurs, you can try to resolve it by following these recommendations. If • You should verify that the probe being used is the problem cannot be resolved, your Apilus xCell working properly by performing the accessory should be entrusted to an authorized service cen- test or by using a new probe.
PicoBlend, OmniBlend, MultiBlend, SynchroBlend*, Thermolysis: EvoluBlend* (*included in the Pro version, optional PicoFlash, MeloFlash, MultiPlex, Synchro* for the Pure version) (*included in the xCell Pro version, optional for the Galvanic current: xCell Pure version) adjustable from 0.00 to 0.99mA, in 0.01mA incre- ments 27.12 MHz...
Page 50
Probe sizes: 5 Power consumption: 110 VA Number of user programs: Operating Conditions: xCell Pur version: 60, 90 with Synchro option xCell Proversion: 900 Temperature: 0°C to +40°C, Relative humidity: 20% to 90%, Current activation: Atmospheric pressure: 70 kPa to 106 kPa...
GUIDE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY The Apilus xCell is for use in the electromagnetic environment as described below. The customer or the user of the Apilus xCell must ensure that it is used in such an environment. Test Condtion Immunity Test...
Page 52
GUIDE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY The Apilus xCell is for use in the electromagnetic environment as described below. The customer or the user of the Apilus xCell must ensure that it is used in such an environment. Test Condtion Immunity Test...
Page 53
GUIDE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY The Apilus xCell is for use in the electromagnetic environment as described below. The customer or the user of the Apilus xCell must ensure that it is used in such an environment Emission Tests Compliance...