Titer/Calibration factor 3.6.9 Rinsing time 3.6.10 Reagent Volume 3.6.11 Analog Output setting (mode) SWITCHING FUNCTIONS OF THE AQUACON PH-CL External Switching Display in the standby mode Analysis state relay Limit Value Relay K2 and K3 Alarm Limit Relay May 2011...
Page 4
AQUACON PH-CL (PC Cabinet) Technical Information Page 2 Current Output START-UP AND MAINTENANCE INSTRUCTIONS Connection of sample water Connection of reagent, bleeding of pumps and tubes Replenishment of Reagent Changing of Tube box Recalibration of the Process analyzer Maintenance interval...
No remodelling ! Never remodel an analyzer. Otherwise, a serious accident may result. IOTRONIC will not be responsible for any accident or damage of any kind which is caused by the user remodelling the analyzer. Operating site must be free of water and humidity The analytical instrument is not designed to be water-proof or dust-proof.
Page 4 Principle of Operation The process photometer AQUACON PH-CL performs a discontinuous photometric determination of the concentration of free chlorine and of the pH value. The measurements are performed one after the other. The entire operation, which includes filling and rinsing of the measuring cuvette, is controlled by a microprocessor.
Page 7
AQUACON PH-CL (PC Cabinet) Technical Information Page 5 View of the process photometer AQUACON PH-CL AQUACON FE10 Automatic Operation OPERATING MODE / PARAMETER DISPLAY Manual Operation Operating Mode: Mode is selected for the following drives by pressing the appropiate Standby...
Page 8
AQUACON PH-CL (PC Cabinet) Technical Information Page 6 AQUACON PH-CL Display and Keyboard Explanation of Symbols Symbol Function Select Automatic Mode, start an analysis Reagent pump 1 on/off Reagent pump 2 on/off Reagent pump 3 on/off Sample inlet valve open/close...
Page 7 Analytical Part The transmitter of the process photometer AQUACON PH-CL is contained in a compact housing and includes the microprocessor which recognizes the absorbance (extinction), displays the final calculated result on a touch screen and controls the operation of the process photometer with respect to sampling, rinsing of the measurement chamber, dosing of the reagent and surveillance of the photodetection system.
(ERROR E-4) is displayed on the touchscreen, subsequent relay activation is initiated by the process photometer AQUACON PH-CL and the analyzer stops. The optical transparency of the measurement cuvette is measured at the start of each analysis. An automatic zero adjustment takes place before each measurement to compensate colored or turbid samples or discoloration/contamination of the cuvette windows.
AQUACON PH-CL (PC Cabinet) Technical Information Page 9 Automatic Operation The process photometer will automatically carry out an analysis once the ”Start” button is pressed. On the return of power, following a power failure, the process photometer will automatically enter the automatic mode and starts a measurement.
AQUACON PH-CL (PC Cabinet) Technical Information Page 10 Manual Operation Manual operation of the process photometer is useful for maintenance, commissioning of the analyzer and when correction of alarm conditions during normal operation occur. Manual operation of the process photometer is effected by pressing the button ”Hand” on the touchscreen.
AQUACON PH-CL (PC Cabinet) Technical Information Page 11 Display/Set Parameters The parameter mode can be activated from automatic mode by touching the button ”Param” or from manual mode by touching the symbol In parameter mode, a new screen will be displayed which shows the parameters Limit value 1 pH, Limit value 2 pH, Limit function pH, Limit value 1 Cl2, Limit value 2 Cl2 and Limit function Cl2.
AQUACON PH-CL (PC Cabinet) Technical Information Page 12 To change the parameters, the corresponding values must be touched. Then, a parameter specific screen opens where the new parameter values can be typed by touching the figures (except parameters Output mode, Limit function pH, Limit function Cl2).
AQUACON PH-CL (PC Cabinet) Technical Information Page 13 3.6.3 Limit Function pH The parameter limit function pH defines the function of the limit value relay K2 for the pH analysis. It can be set to min function or to max function.
Technical Information Page 14 3.6.8 Titer/Calibration factor The process photometer AQUACON PH-CL may be re-calibrated by entering a titer/calibration factor (see chapter 5.5). A value between 0.00 and 9.99 may be chosen. 3.6.9 Rinsing time The measuring chamber is rinsed for 15 seconds before an analysis takes place. The preset rinsing time may be altered by the end-user.
Page 15 Switching Functions of the AQUACON PH-CL CAUTION: Before opening the cover below the process photometer display, the AQUACON PH-CL must be disconnected from the mains supply. External Switching The process photometer PH-CL has a potential-free input contact which may be used to control the instrument from a remote position/device.
”ERROR E-1, ERROR E-2, ERROR E-3” or ”ERROR E-4” (see also chapter 6). Current Output The process photometer AQUACON PH-CL delivers a current output, 0 or 4 to 20 mA, which is proportional to the measurement range selected by the end-user (see chapter 9.1). This current output may be used as input to remote recording devices, for data processing, controlling functions etc.
AQUACON PH-CL (PC Cabinet) Technical Information Page 17 Start-up and Maintenance Instructions CAUTION: As soon as the analyzer is connected to the main supply, automatic operation commences. This is required for automatic restart of the analyzer in the event of a power failure. At start-up, it is necessary to access the Manual Operating mode by touching the ”Hand”...
AQUACON PH-CL (PC Cabinet) Technical Information Page 18 Changing of Tube box Replacement of the complete tube-box by pressing the brackets at the sides,take off the old box and give up a new box. Important: All tube boxes must be replaced after consumption of 5 bottles reagent or latest after 6 months.
AQUACON PH-CL (PC Cabinet) Technical Information Page 19 Alarm Messages Message in Description ACTION display Error in the optical system; the ERROR E -1 Check sample inlet valve, clean analyzer operation is interrupted: the measuring chamber, call - contaminated cuvette windows maintenance service.
AQUACON PH-CL (PC Cabinet) Technical Information Page 20 Specifications Parameter pH value and free chlorine Description Automatic microprocessor controlled analyser for the photometric determination of pH value and free chlorine in water AQUACON PH-CL Typical Applications Control and supervision of chlorine plants and swimming pools...
AQUACON PH-CL (PC Cabinet) Technical Information Page 21 Connections In – and outputs are available at circular connectors at the top of the device. CAUTION: technical data ! power supply alarm external start 0/4-20 mA recorder K3 option (Male) K1 / K2 / alarm...
AQUACON PH-CL (PC Cabinet) Technical Information Page 22 Installation WARNING Only Qualified Personnel ! Only Specified Power Supply ! See required power supply as noted on instrument. Observe local regulations! Install a power supply that can be disconnected (plug, switch).
AQUACON PH-CL (PC Cabinet) Technical Information Page 24 10.2 Spare part and consumable item list Consumable item Term Quantity Order No Reagent 1 for AQUACON PH-CL PH80 500 ml 101 3180 01 Reagent 2 for AQUACON PH-CL CL-R1001 250 ml 101 2725 01...
AQUACON PH-CL (PC Cabinet) Technical Information Page 25 Troubleshooting Guide Problem Reason • The contact bridge for external switch function (see 4.1) is not installed or On pressing the “automatic operation” key switch, the analyzer has no contact switch. does not enter automatic analysis “H-E”...