MP30 and MP35 Specifications
- 1 -
IMPORTANT SAFETY NOTICE
BIOPAC Systems, Inc. instrumentation is designed for educational and research
oriented life science investigations. BIOPAC Systems, Inc. does not condone the
use of its instruments for clinical medical applications.
Instruments, components, and accessories provided by BIOPAC Systems, Inc.
are not intended for the cure, mitigation, treatment, or prevention of disease.
The MP35/30 is an electrically isolated data acquisition unit, designed for
biophysical measurements.
Exercise extreme caution when applying electrodes and taking bioelectric
measurements while using the Biopac Student Lab with other external
equipment that also uses electrodes or transducers that may make electrical
contact with the Subject.
Always assume that currents can flow between any electrodes or electrical
contact points. In case of equipment failure, it is very important that significant
currents are not allowed to pass through the heart.
If electrocautery or defibrillation equipment is used, it is recommended that the
BIOPAC instrumentation be disconnected from the Subject.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MP30 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BIOPAC Systems, Inc. MP30

  • Page 1 MP30 and MP35 Specifications - 1 - IMPORTANT SAFETY NOTICE BIOPAC Systems, Inc. instrumentation is designed for educational and research oriented life science investigations. BIOPAC Systems, Inc. does not condone the use of its instruments for clinical medical applications. Instruments, components, and accessories provided by BIOPAC Systems, Inc.
  • Page 2 - 2 - BIOPAC Systems, Inc. MP35/30 ACQUISITION UNIT The MP35/30 data acquisition unit is the heart of the Biopac Student Lab PRO System. The MP35/30 has an internal microprocessor to control data acquisition and communication with the computer. The MP35/30 unit takes incoming signals and converts them into digital signals that can be processed with your computer.
  • Page 3: Front Panel

    - 3 - MP30 Hardware Guide SIMPLE SENSOR CONNECTORS Regardless of the type of device connected, every sensor or I/O device connects to the MP35/30 using a “Simple Sensor” connector. Simple Sensor connectors are designed to plugs only one way into the MP35/30, so you don’t have to worry about plugging things in upside down or into the wrong socket.
  • Page 4: Back Panel

    The back panel of the MP30 has an analog output port, a serial port, a DC input, a fuse holder, and a power switch, and the unit’s serial number.
  • Page 5: Fuse Holder

    Fuse holder The fuse holder contains a fast-blow fuse that helps protect the MP35/30 from shorts on its power, analog, and digital I/O lines. The MP35 uses a 1.0 amp fast-blow fuse and the MP30 uses a 2.0 amp fast-blow fuse.
  • Page 6: Specifications

    - 6 - BIOPAC Systems, Inc. SPECIFICATIONS Specification MP35 Unit MP30 Unit Front Panel ELECTRODE CHECKER 0-100 KΩ 0-100 KΩ Resistance Range (Vin+ and Vin- to GND) ANALOG INPUTS 4 isolated (front panel CH 1–CH 4), 4 (front panel CH 1–CH 4)
  • Page 7: Serial Interface

    - 7 - MP30 Hardware Guide Specification MP35 Unit MP30 Unit SERIAL INTERFACE Transmission type RS422-clocked asynchronous Transmission rate Type 2.0 full speed 524,000 bits per second (KBPS) HEADPHONE (MP35 only) Drives low-impedance standard N/A – MP35 only stereo headphones...
  • Page 8 Buffered DC output Z out = 50 Ω Z out = 50 Ω V out range MP35: (0 to 4.096 V) MP30: (0 to 5 V) +5.0 V (100 mA max) +7.5 V (100 mA max) Buffered digital output Unbuffered DC output Z out = 1 kΩ...
  • Page 9 - 9 - MP30 Hardware Guide MP35 Only MP UNIT PIN OUTS continued Digital Output 1 I/O Port — MP35 Back Panel Digital Output 2 DSUB 25 (male) Digital Output 3 Digital Output 4 10 11 12 13 GND Unisolated...