Summary of Contents for Magnetrol Solitel
Page 1
Installation and Operating Manual Solitel Vibrating Rod evel Switch...
Page 2
Safety Messages Quality ssurance The Solitel system is rated by the IEC for use in Category The quality assurance system in place at MAGNETROL II, Pollution Degree 2 installations. Follow all standard guarantees the highest level of quality throughout the industry procedures for servicing electrical and computer company.
Refer to Calibration, Section 2.5. IMPORTANT: Handle the instrument with great care, especially the probe. Any impact on the probe can damage the vibration system. Figure 1 56-601 Solitel Vibrating Rod Level Switch...
The SOLITEL switch is mounted to the vessel using a ⁄ " NPT threaded connection. • Ensure that the tip of the probe is located at the desired switching point giving consideration to the slope angle of the material. •...
Page 5
This allows detection of power failure as an alarm status. 7. Wiring is complete. Replace the housing cover. 8. Apply power to the unit. NO CO NC Relay L1(+) No Connection L2/N(-) Power Figure 6 56-601 Solitel Vibrating Rod Level Switch...
Page 6
NOTE: If the adjustment is too sensitive, the switch may detect even slight residue of material rather than level. Prior to installation, it is recommended that the unit be calibrated using a sample of the media to be measured. 56-601 Solitel Vibrating Rod Level Switch...
Reference Information Troubleshooting SOLITEL is a simple and reliable device. However, care must be taken to avoid impacting or damaging the vibrat- ing probe. In the event the SOLITEL stops vibrating, the following procedure will identify if the problem is with the electron- ic module or with the sensor: Disconnect the wires going to the sensor.
316 stainless steel Insertion length 8.25 inches (209 mm) 10 to 100 inches (25 to 254 cm) Maximum mechanical load on probe Stranded probe 63 Ft/lbs. (85 Nm) Extended rigid probe 63 Ft/lbs. (85 Nm) 56-601 Solitel Vibrating Rod Level Switch...
Page 10
Conduit connection plugged (38) (38) " NPT " process connection process connection 1.73 (44) Dia. Insertion length Insertion length 1.73 8.25 (44) Dia (209) 6.34 (161) 6.34 (161) Standard Length Extended Rigid Probe 56-601 Solitel Vibrating Rod Level Switch...
Page 11
Design type code 1 or 2 with probe type code R: Code 000 fixed length of 8.25" (21 cm) Design type code 1 or 2 with probe type code R 10 to 100 inches — Example: 10 inches = code 010 56-601 Solitel Vibrating Rod Level Switch...
Page 12
Service Policy Return aterial Procedure Owners of Magnetrol controls may request the return of So that we may efficiently process any materials that are a control or any part of a control for complete rebuilding returned, it is essential that a “Return Material or replacement.