Thank you very much for purchasing pH / ORP meter PE-8E for industrial use. This manual is for PE-8E, which is dual use type of pH / ORP . You are requested to read this manual carefully to study various functions and operations.
A location where is not exposed to direct sunshine, to high radiant heat. Less humid ③ : Do not install PE-8E at the places of high humidity or under the conditions that the water and chemicals may splash on. These may be causes of the damages to insulation or of rust.
Lower limit action Internal Switches are composed of DIG.Switch and DIP.Switch. When you use PE-8E with other setting from the above, proceed the setting as follows ; 2.3.1 Definition of DIG. Switch 16 variations from 0 to F at setting are available by DIG. Switch.
Page 6
Setting contents Change-over the resistance of 6.8 k Ω at25℃ Ω at 25℃ measuring temperature Change-over ALM1, relay control Upper limit Lower limit action action ( ) action ) Change-over ALM2, relay control Upper limit Lower limit action action action (...
Page 7
1 Automatic calibration Automatically calibrates the electrode sensitibity by standard solutions. 2-Points Automatic calibration :Types of standard solutions : pH 4 , 7 , 9 JIS Combinations of standard solutions pH 4 , 7 or pH 7, 9 ★ Automatic discrimination of Standard solution types Automatically discriminates the types of standard solutions...
2.3.3 Confirmation of the setting contents This PE-8E displays main parameters of DIG.Switch and DIP.Switch when powered ON, and then enters into measuring. Please check the displays when you re-set the contents after powerd Powered ON › First Step ›...
Please pay attention in wiring load to it that COM-B contact is ON when instrument power is OFF. This PE-8E has 2 channels of alarm contact output; ALM1 and ALM2 1c contact output Control action of both ALM1 and ALM2 can be set upper limit action or lower limit action by DIP.
WIRING Electrode cable has the terminals as under-mentioned ; Make sure to fix the fixed resistance between T and T. This PE-8E is applicable to two types of temperature compensation electrode. 6.8 k at 25 Ω at 25℃ or 500Ω...
20mA corresponds to the measuring range. As the input resistance to the receivers is limited to 500 at maximum, select the Ω suitable receivers recorders, meter relays, etc which are adjusted to this PE-8E. 20mA less 〜 Ω or +...
4. Explanation of each parts Display: Measured Value Set Value or Parameter MEAS DISP Operation lamps MEAS ▲ ▼ Operation keys PE‑8 E S X If you open this cover, you find ENSONI JAP AN DIG.Switch and DIP. Switch. 4.1 Operation keys It returns to the normal operation measuring and MEAS lamps MEAS lights up when you push this key during no measuring status...
4.2 Operation lamps MEAS Measuring Lamp, which lights up under ordinary operations. Under this conditions, this PE-8E refers measured values and set values, ○ then decides ON/OFF of control output ALM 1 , ALM 2 . The lamp stays off during calibrations.
This PE-8E is capable for automatic and manual calibrations. ③ It is changeable by DIP Switches. Refer to the page: During calibrations by standard solutions, control actions ALM 1 , ALM 2 are all OFF. ④ During calibrations by standard solutions, transmission output turns into the mode set by DIP ⑤...
Page 15
Push CAL key. ③ Stability of electrode potential is discriminated, then calibrates it to the pH value of standard solution at that temperature of discrimination. MEAS DISP Push CAL key again even if you calibrate with the same standard solutions. It is possible to calibrate many times with the same standard solutions, but the only final data is effective.
Push ENT key last , otherwise calibration data would not be renewable. Push ENT key. ⑥ Set Value Display shows 『 』 , then automatically turns into measuring mode. MEAS DISP Now, calibration is finished. ⑦ 5.2.2 Function of manual calibration It is actually used when you do not want full autmatic calibrations by standard solution, or you need any other combinations of standard solutions except pH 4 , 7 , and 9.
5.2.3 Manual calibration procedure When you stop calibrations halfway , push MEAS key. it returns into measuring mode. In this case, calibration data is not renewed. ① Immerse electrode in the first point standard solution . ② Push CAL key for 1 second or over. MEAS lamp goes off , then CAL lamp lights up.
Page 18
shows on Set Value Display, ⑤ 『 』 which means " the second-point calibrations started ". MEAS DISP Immerse electrode in the second point standard solution . ⑥ ⑦ Push ENT key after measured value came to stable. The measured value is copied on Set Value Display, then starts flashing.
MEAS DISP 5.3. Calibration For ORP Meter Use Cautions at during calibrations Following works shall be unnecessory in pH Meter Use. This work is for a check if the ORP electrode would be sensible enough. No corrections of electrode sensitivity are carried out at meter body side as pH electrode is. 5.3.1 Sensitivity check by ORP standard materials Prepare ORP standard materials.
5.4 Parameter Display This PE-8E is capable of showing and setting each parameter on Set Value Display concerning ALM 1 , ALM 2 , measured value, and transmission output. Even during operations, you can check these parameters. Push SEL key, each parameter shows up in order as follows.
Parameter concerning Measured Value ☆ ☆ Shift Value ・・・・ Displays total quantity added / subtracted with a certain quantity. When shift quantity is 0 , measured value shows on display untouched. When entered into the renewal mode of shift quantity , measured value shows on Measured Value Display.
6.Error codes This PE-8E is equipped with display function of each error code. Each error code is divided two depending on application of pH Meter-Use and ORP Meter-Use. Each error code is shown on Set Value Display. Measured Value flashing Error Codes shown on 6.1 Error codes of pH / ORP meter...
6.2 Countermeasures to error codes Take following countermeasures when error codes show on display. ERROR CODES CAUSES COUNTERMEASURES The electrode is not Whether the level of sample ・ ・ immersed in the sample liquid fluctuates,make sure to oF / uF liquid .
Page 24
ERROR CODES CAUSES COUNTERMEASURES E-14 ・ pH standard solution ・ Calibrate it with new standard Abnormal Standard deteriorated. solution. Solutions Calibrations with any other Calibrate it with new standard ・ standard solutions but pH solution of pH 4, 7 ,and 9 4 ,7 , and 9 JIS Manual calibrations with Calibrate it manually with...
7 Spare Parts SPARE PARTS ODER SPECIFICATION Cable For GR Combination Electrode Special For GRT Combination Electrode Holder Submersible Type Holder Length : 1 m Flow Type Connector Box CT-A,B Special Fitting NT-B AES Resin Stainless Steel SUS304 ACCESORIES Calibration Set Standard Solution , pH 7 500ml )・・・・...
Specifications PE-8E MODEL pH 0 resolving power : 0.01 pH 〜 MEASURING RANGE resolving power : 1 mV ± 1400mV 0.05 , ORP 5 mV ± ± REPEATABILITY DC 4 20 mA Isolated Type for both of input / output 〜...