This Vitrek instrument is warranted against defects in material and workmanship for a period of two years after the date of purchase. Vitrek agrees to repair or replace any assembly or component (except batteries) found to be defective, under normal use, during the warranty period.
Page 4
All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language in any form with prior written consent from Vitrek. This product’s user guide is copyrighted and contains proprietary information, which is subject to change without notice.
Contents Contents Warranty _______________________________________________ 3 Introduction _____________________________________________ 9 Product Description _________________________________________ 9 Features _________________________________________________ 10 Front Panel Controls and Connections _______________________ 11 Display __________________________________________________ 11 “POWER” Switch _________________________________________ 11 Battery Indicators __________________________________________ 12 Red “BATT CHG” Light _________________________________ 12 Green “OK” Light _______________________________________ 12 “OUTPUT”...
Page 6
2000I & 2000M Instrument’s User Guide, Revision B To Charge the Battery ____________________________________ 19 Charging Status _________________________________________ 20 Charging from the Charging Module ___________________________ 20 Charging Modules _______________________________________ 21 Charging from an External DC Voltage Supply __________________ 22 Continuous Power from External Supply _______________________ 22 Operating the 2000 Instrument _____________________________ 23 Front Panel Operation ______________________________________ 23 Initial Power Application _________________________________ 23...
Page 7
Contents Ensuring Stable Current Connections ________________________ 50 Process Control Industry Applications__________________________ 52 Calibrating in Process Control Environments __________________ 52 Simulating Thermocouple Systems __________________________ 55 Measuring Thermocouple Outputs __________________________ 56 Engineering Laboratory Environment Applications _____________ 57 Calibration and Maintenance ______________________________ 61 Internal Calibration ________________________________________ 61 External Calibration ________________________________________ 62 Equipment Required to Recalibrate _________________________ 62...
Page 8
2000I & 2000M Instrument’s User Guide, Revision B Appendix A– Output Specifications _________________________ 85 DC Voltage Output Specifications _____________________________ 85 DC Current Output Specifications _____________________________ 86 Voltage Measurement Specifications ___________________________ 87 Thermocouple Specifications _________________________________ 87 Appendix B– General Specifications ________________________ 93 Warm-Up Time _________________________________________ 94 Environmental __________________________________________ 94 Isolation _______________________________________________ 94...
Introduction Introduction This chapter describes the 2000I, 2000M, 2000IN and 2000MN DC power source instruments and highlights their features and functionalities. The 2000 instru- ments were introduced in 1989 to support the fields of Process Control, Calibra- tion, and General Equipment Testing. Product Description The 2000I, 2000M, 2000IN and 2000MN products are portable, lightweight DC power source instruments.
2000I & 2000M Instrument’s User Guide, Revision B Note: To ensure full use of this instrument, review this complete user guide at setup. Features Fully bi-polar DC voltage and current output capability. DC output of 22 volts at 10ppm accuracy and 22mA at 40ppm accuracy. ...
Front Panel Controls and Connections Front Panel Controls and Connections This chapter discusses the basic use of the various controls, connectors and indi- cators on the front panel of the 2000I, 2000M, 2000IN and 2000MN DC power source instruments. Before using your instrument refer to the text on Charging the Battery.
2000I & 2000M Instrument’s User Guide, Revision B When the POWER switch is toggled up, in the ON position, the microprocessor and analog circuitry is powered and the instrument may be operated normally. The power source may be either from the battery or the charging module, as ap- plicable.
Front Panel Controls and Connections Keypad These 20 keys are used for all manual entries into the instrument and to initiate any required actions. These keys are in the following groups: Numeric Keys These keys: are used to enter numeric data “0, 1, 2, 3, 4, 5, 6, 7, 8, 9,”...
2000I & 2000M Instrument’s User Guide, Revision B Note that prior to selecting the required units key, an entered multiplier may be changed by pressing the other multiplier or cleared by pressing the multiplier key a second time. Pressing a multiplier key while an actual output level or measured input level is being displayed (i.e., not during numeric entry) changes the display.
Important Note: When connecting a charger other than that provided by Vitrek, ensure the selected charger’s output voltage is: 12V at 1.5A and the center conductor is positive polarity and uses 2.5mm DC plug.
2000I & 2000M Instrument’s User Guide, Revision B Interface Connectors A connector slot is available at the top center in the back panel. A connector in this location is only present when either of the interface options IE-BAT or RS- BAT is fitted in the instrument.
Page 17
Rear Panel Connections and Controls Note that the battery life is considerably shortened when the IE-BAT Interface is enabled, thus it is highly recommended that this switch be maintained in the OFF (down) position whenever this interface is not being used. The IE-BAT Interface is only usable when this switch is in the ON position.
Page 18
2000I & 2000M Instrument’s User Guide, Revision B...
12 hours prior to its initial use. Use the charging unit provided Vitrek to charge the internal battery and to operate this instrument. Ensure that the included charging unit is compatible with your local power ratings and socket connection.
2000I & 2000M Instrument’s User Guide, Revision B Charging Status For battery charge status, refer to the front panel’s red BATT CHG light. If the BATT CHG light extinguishes then the battery is fully charged. If the BATT CHG light does not extinguish after several hours of charg- ing, turn the unit OFF and back ON (2000I and 2000M Only).
Charging the Battery Warning: Prior to recharging the internal battery disconnect the 2000 instru- ment from any intrinsically-safe process control system. Charging while con- nected could adversely affect the safety of the process control system. Charging Modules One of these modules is included with the main unit (as specified at the time of order).
2000 instrument. It is recommended that for the 2000I and 2000M products the unit is allowed to discharge completely to reduce the memo-ry effect of the Lead Acid battery. Vitrek recommends using an external DC supply for process control applications.
Operating the 2000 Instrument Operating the 2000 Instrument This chapter discusses operating the 2000I, 2000M, 2000IN and 2000MN DC power source instruments. Throughout this chapter, the instrument is assumed to be a 2000M or 2000MN, which has full capabilities, described in this User Guide. If data is entered for which the option is not fitted, then a message displays for two seconds, and the entry is discarded.
2000I & 2000M Instrument’s User Guide, Revision B When the POWER switch is placed in the OFF position, the terminals remain in the condition prevailing. Vitrek recommends selecting the “Stand-by” state prior to switching the POWER switch OFF, to afford the highest level of protection both to any circuitry connected to the instrument and to protect the instrument from inadvertent application of external signals.
Page 25
Operating the 2000 Instrument The displayed output level will be present when the instrument returns to “Oper- ate” status. Displaying Mode Options The menu-selectable modes display the opposite selection that is presently active in the instrument. For instance: When “Select Operate” displays, the instrument is actually in “Standby”. ...
2000I & 2000M Instrument’s User Guide, Revision B The display then returns to displaying the selected output level and mode. The output terminals will either be open circuit–if “Standby” was selected, or at the requested output level–if “Operate” was selected. Caution: Prior to selecting the “Operate”...
Operating the 2000 Instrument 2. Enter or reenter the entire numeric. If the CLR key is pressed at this time, then the numeric entry is aborted, the display returns to the previous display. 3. If an incorrect multiplier key is pressed, then simply press the correct multip- lier to overwrite the previously selected incorrect one.
2000I & 2000M Instrument’s User Guide, Revision B Note that the output level is displayed using the same multiplier as used during the entry of the output level, thus entries of: 1V, 1000mV 1000000uV Each achieve an output level of: , but the display format will differ.
Operating the 2000 Instrument 1. For example: if the display is: +.1000000 V 2. Press the key to reformat the display to: µ +100000. µV Direct Entry of Temperature Data The temperature of the hot junction of the thermocouple system to be simulated may be directly entered similarly to entering output voltage or current.
2000I & 2000M Instrument’s User Guide, Revision B 2. Press the +/- key to change the simulated temperature to: -100.00ºC R Note that unlike voltage or current output levels, the limits of available simulation temperatures are not identical in both polarities (in general). If the attempted tem- perature exceeds these limits, then the simulation temperature will be set to the limit at the requested polarity and a warning message will be displayed for ap- proximately two seconds.
Operating the 2000 Instrument The amount by which the output has been deviated from the initial value is shown in the display. The level present when the “Adjust” mode was initiated is referred to as the no- minal output in this document. Examples of display formats during “Adjust” mode are shown below, with the increment digit underlined for clarity.
2000I & 2000M Instrument’s User Guide, Revision B (down arrow) key. The output level or simulated temperature is incremented or reduced by the value of the selected (flashing) digit with full under/ and over- flow carry to the remaining digits (including carry through zero). Changing Adjustment Value As explained in the preceding paragraph, the output level (or simulated tempera- ture) is adjusted by the value of the flashing digit in the display.
Operating the 2000 Instrument If neither “Standby” or “Operate” are the desired action at this point, con- tinue to press any of the arrow keys, or the MNU key until the “Set De- faults” action is displayed: < Set Defaults > ...
2000I & 2000M Instrument’s User Guide, Revision B The “Range Lock” mode can be selected by stepping through the menu, similar to the operation of all earlier firmware versions. The three possible locked ranges, 20V, 2V, and 20mA, will appear in sequence. With the desired range on the dis- play, pressing any one of the Units/Enter keys will lock the instrument in the shown range.
Page 35
Operating the 2000 Instrument 1. Press the MNU key, the display will then prompt you for selection of either the “Standby” or “Operate” state as described previously in this chapter. 2. Since neither “Standby” nor “Operate” are desired, press an arrow key until the “Calculated Output”...
Page 36
2000I & 2000M Instrument’s User Guide, Revision B Note that the units for this data are automatically those previously selected for the Output A data. 9. The display then prompts you to enter User Data B and displays the presently stored data.
Operating the 2000 Instrument Selecting Measurement Mode Use this mode to measure voltage or temperature of another device. With the instrument in Measure mode, the output connectors automatically reconfigure as input terminals. To select Measurement mode: 1. Press the MNU key to access “Operate” or “Standby” modes. 2.
2000I & 2000M Instrument’s User Guide, Revision B Rear Panel Operation All the connections available on the 2000 instruments rear panel are dependent on which options are fitted. Refer back to the Rear Panel Connections and Controls. Note: In the 2000M instrument, the thermocouple simulation output, or thermo- couple measurement input, should be taken from the connector mounted on the rear panel.
Page 39
Operating the 2000 Instrument Notes: Selecting and deselecting thermocouple simulation is simply achieved by: 1) Sending a temperature to the instrument, for example: Send the com- mand string to select the thermocouple simulation mode “+22.89C” with a simulated hot junction of +22.89ºC 2) Sending a voltage or current to the unit: for example: The command string “...
Page 40
2000I & 2000M Instrument’s User Guide, Revision B Neither “Standby” nor “Operate” are desired actions, therefore continue to press any of the arrow keys repeatedly until the “Thermocouple” set up ac- tion is displayed. < Thermocouple > Press any of the Units/Enter keys (V, A, ºC or ºF to select “Thermocouple”.
Operating the 2000 Instrument For Thermocouple specifications refer to Appendix A—Power Output Capacity Specifications. Deselecting Thermocouple Simulation Mode To deselect the thermocouple simulation mode, select for a voltage or current output by pressing the Units/Entry keys (V or A). This method offers the ability to view the actual output level being used for the Thermocouple Simulation.
Page 42
2000I & 2000M Instrument’s User Guide, Revision B...
Using the Memory Options Using the Memory Options The option called MEM is a 10-step memory, which allows you to program up to 10 output levels. These levels may be recalled speedily from either the front panel or by the rear panel mounted contact closure input. The ten memory steps are automatically stored in electronic nonvolatile mem- ory.
Page 44
2000I & 2000M Instrument’s User Guide, Revision B Editing Memory Steps 1. Press the MNU key twice. The display now shows the choice: < Edit Memory > 2. Press any of the ENTER keys to select this choice. The display now shows the prompt for the memory step number to be edited, for example: <...
Using the Memory Options pressing any of the Units/Enter keys. The display is then returned to the “Press Memory” prompt, as shown in step 2 above. Recalling Memory Steps Front panel memory steps may be sequentially recalled by following the proce- dure shown below.
2000I & 2000M Instrument’s User Guide, Revision B with each instrument as standard and when MEM is ordered. (Vitrek part number C109001001000001). General Memory Considerations Memory of Instument Status The contents of a memory step only affect the selection of the “Operate” or “Standby”...
2000I & 2000M Instrument’s User Guide, Revision B Applications This chapter describes specialized applications for using the 2000I, 2000M, 2000IN and 2000MN DC power source instruments and some ways to ensure the quality of the power source. The front panel terminals are used to: ...
Use very heavy gauge wire (16 AWG or larger), achieving the same effect by pulling the terminal to the same temperature as that of the wire. • Avoid using standard, commercial banana jacks. Vitrek en-sures the SP48 PL36 options are manufactured from highly pure copper and gold plating, ensuring they exhibit low thermal effects.
2000I & 2000M Instrument’s User Guide, Revision B In most circumstances, the use of a tightly twisted pair of 28AWG wires is recommended, as this both reduces the effects of thermal E.M.F. voltages and achieves the close coupling required to reject interference. ...
Page 51
Applications When dealing with low-level DC currents at high compliances, the effects of leakage currents become dominant. To limit leakage use the correct in- sulation throughout the cabling, and keep all connectors and front panels free of grease and dirt. Note that most low-level leakage currents are non-ohmic in nature and tend to be a constant current leakage for com- pliances above a few hundred millivolts.
2000I & 2000M Instrument’s User Guide, Revision B Use the Vitrek option SP48 shielded lead set. To reduce current interference to acceptable levels: Use twisted pairs of 26AWG (or smaller) wire with Teflon® insula- tion. Use Polyethylene insulated shielded cables or short lengths of bare wire.
Page 53
Applications Using the Adjust Mode to Check Calibration For this application the “Adjust” mode of operation is used to measure the error of the equipment being tested. To test, simply set the calibrator to the required output level and then adjust the output level until the equipment being calibrated reads the correct value.
Page 54
2000I & 2000M Instrument’s User Guide, Revision B Where is the tolerance of the input impedance (in ppm, 1% = 10000ppm). Where Error is the increase in the limits of error (in ppm). Thus, if the combined input impedance is 1M •...
Applications the required correction, in the units used by the transducer. Note that this correc- tion may be of either polarity. Note: The entered data are non-volatile, thus they may be entered at your base location, and then be used at a remote site without the need to reenter these data. Refer to Using the Memory Options.
2000I & 2000M Instrument’s User Guide, Revision B The 2000M and 2000MN instruments simulate the thermocouple system by com- puting the hot and reference junction-E.M.F. voltages from the N.I.S.T. tables with linear interpolation. The output level is set to the difference between these two levels (V ).
Applications Engineering Laboratory Environment Applications The 2000 instruments have many uses in the Calibration and Engineering Labora- tory environments. Short-Term Transfer Standards The 2000 instruments can be used for medium accuracy transfers in the calibra- tion laboratory based on their: ...
2000I & 2000M Instrument’s User Guide, Revision B , thus this The 2000 instruments have an output impedance greater than 10 affect is not of importance for unknown values below 1M. The following table gives the uncertainty of the transfer for various values of standard and unknown resistors.
Page 59
Applications Other Laboratory Applications: Variable voltage or current reference source: The 2000 instruments may be used as voltage and current reference sources in most instrumentation circuits, enabling the measurement of reference sen- sitivity and related effects. The low noise of battery power injects little noise into the circuitry.
Page 60
2000I & 2000M Instrument’s User Guide, Revision B...
Calibration and Maintenance Calibration and Maintenance This chapter describes the procedures for internal calibration, external calibration, periodic maintenance and internal maintenance of the 2000I, 2000M, 2000IN and 2000MN DC power source instruments. Internal Calibration The internal calibration procedure ensures the 2000 instrument’s low-level DC voltage and current specifications are maintained.
The period at which these calibrations are performed is dependant on your accu- racy requirements, as shown in the accuracy tables in Appendix A-Power Output Capacities Vitrek recommends calibration intervals of 90 days, 6 months or 1 year to meet your in house accuracy needs. For example: in some applications, such as, general engineering laboratories and process control envi-ronments, the 1 year accuracies may be excessive.
The 2000 instruments require current measuring equipment with the accuracy requirements shown in Table 3, for a margin of 2:1 or better. Vitrek recommends using a high-accuracy digital DC ammeter, such as, the Hew- lett-Packard 3458A, or a DC voltage measuring device and laboratory standard resistors to achieve the current measurement accuracies.
2000I & 2000M Instrument’s User Guide, Revision B instrument. Any zero offsets in the measuring system can be compensated for by calibrating the zero current of the system using an open circuit measurement. Initiating External Calibration To initiate an external calibration you may be required to enter a password and/or the common (global) password.
Calibration and Maintenance Note: To abort the external calibration selection at any time, press the CLR key. You will return the instrument to the output level and status that was present when the MNU key was initially pressed. 6. In no password is requested or the correct password is entered, press any of the arrow keys to continue.
Page 66
2000I & 2000M Instrument’s User Guide, Revision B Front Panel Key Adjustments The left and right arrow keys select the adjustment digit, the up and down arrow keys adjust the selected digit. To abort any external calibration step press the CLR key. To abort the calibration entirely press the MNU key. To nor- mally end a calibration step—thus entering the adjusted calibration factor, press any of the Units/Enter keys.
Page 67
Calibration and Maintenance This output level may be measured directly by using: A high precision DMM, or; A nullmeter to monitor the deviation between this instruments' output level and that of a precision calibrated 10mV source, or; The preferred method of using a nullmeter to monitor the deviation be- tween this instrument’s output level and the output of a high precision 100:1 attenuator whose input is obtained from a precision calibrated 1V source.
Page 68
2000I & 2000M Instrument’s User Guide, Revision B Nullmeter 2000 10V DC Source 10:1 Divider This output level may be measured directly by using: A high precision DMM, or; A nullmeter to monitor the deviation between this instrument’s output level and that of a precision calibrated 1V source, or;...
Page 69
Calibration and Maintenance This output level may be measured directly by using: A high precision DMM, or; The preferred method of using a nullmeter to monitor the deviation be- tween this instrument’s output level and that of a precision calibrated 10V source.
Page 70
2000I & 2000M Instrument’s User Guide, Revision B This output level may be measured directly by using: A calibration standard 10K resistor and a high precision DMM, or; A nullmeter to monitor the voltage drop across the standard resistor. 11.
Periodic Maintenance The 2000I(N) and 2000M(N) DC power source instruments require little periodic maintenance; however, Vitrek makes the following recommenda-tions: 1. Allow the internal battery to fully discharge, and then fully recharge at least once a year. This will ensure that the battery retains its storage capacity.
2000I & 2000M Instrument’s User Guide, Revision B cessively dirty or damaged, contact Vitrek Customer Service (or one of its agents) for details on how to rectify this situation. 3. If the instrument is to be stored for extended periods of time, fully charge before storing.
The BATT CHG light does not extinguish after 15 hours of charging. Caution: The internal battery should only be replaced by an exact replacement. Obtain a replacement battery from Vitrek. Warnings: 1) Sealed lead-acid and Nickel Metal Hydride batteries are normally stored in a charged condition, take extreme care when handling the battery.
Page 74
2000I & 2000M Instrument’s User Guide, Revision B To replace a battery: Open the instrument and remove the Microprocessor board as described in the instruction: To obtain access to the internal circuitry. Additionally re- move the Analog board and loosen the Power Supply board, as required, to assist in replacing the battery.
Interface Operation Interface Operation This chapter discusses the methods used to interface the 2000I and 2000M DC power source instruments using the optional interface connectors RS-BAT and IE-BAT. In general, operation of these instruments via an interface involves the writing and reading of ASCII data to or from the instrument. In both interfaces, the same characters are used for each command.
2000I & 2000M Instrument’s User Guide, Revision B Selecting RS232 To select the RS232 data format: Repeatedly press the MNU key until the “SET RS232” selection displays. Press any of the Units/Enter keys (V, A, ºC or ºF) to select this action. The unit will then display the presently selected baud rate for the RS232 inter- face.
Interface Operation Remote/Local Operation The IEEE488 interface defines separate “REMOTE” and “LOCAL” states. While in the LOCAL state, which is following the initial application of power, the in- strument will only respond to commands from the instrument’s front panel. Plac- ing the instrument in the “REMOTE”...
2000I & 2000M Instrument’s User Guide, Revision B Bus Commands As previously mentioned, the IEEE488 interface can perform many tasks that the RS232 does not; those used by the 2000 instrument are as follows: 1. Remote/Local Selection. This has been previously described. 2.
Interface Operation Reading Data Via an Interface With either interface, you can read the data presently being displayed on the front panel of the 2000 instrument. With the 2000-RS-BAT interface this is achieved by transmitting the @ character to the 2000 instrument, with the IE-BAT inter- face this is achieved by simply reading the instrument.
2000I & 2000M Instrument’s User Guide, Revision B Note: Command types other than those listed in Tables 4, 5 or 6 are always dis- carded and ignored. Table 4: Command Characters Front Panel Characters Front Command Charac- Received Panel Key ter Received .
Interface Operation Char/Seq Description of Action Selects the “Operate” state. Commands the interface to assert the SRQ (service request) line when a previously received command has been fully actioned. Ap- plies to UE2000 only. Commands a power-on reset sequence. Selects the “Standby” state. Sets the Thermocouple type to that described by the character im- mediately following, e.g., “TK”...
Unlocks the 20mA Range Command String Usage Vitrek assumes the user is conversant with the send and receive data commands when using the actual computer interface and with the language used to control the 2000 instruments. Vitrek also assumes all com-mand strings (those shown enclosed in quotation marks) are followed by the car-riage return and linefeed characters.
Interface Operation Adjusting Output Level This may be achieved by shortcut commands or by simulating keyboard com- mands by using command characters. To select the “Adjust” mode of operation the instrument may send the character string D or the character string MMV. In the later character string, note that these characters simulate the key presses re- quired to perform the desired task: MNU, MNU and V, whereas the first com- mand string uses the shortcut form.
2000I & 2000M Instrument’s User Guide, Revision B be obtained (for both interfaces) by reading the 2000 instrument display data as described in the Maintenance and Calibration chapter. The accuracy specifications presented in the Appendix A tables assume the use of this function at least every 30 days, or for a temperature change of greater than 5ºC from that at the previous internal calibration.
Appendix A—Output Specifications Appendix A– Output Specifications The DC voltage and current output capabilities of the 2000I,2000M, 2000IN and the 2000MN instruments are provided here. Additionally, specifications for the 2000M instrument’s of thermocouple measurement and simulation option are included here. DC Voltage Output Specifications The accuracy figures given in Table 7 are valid throughout the temperature range of 18º...
2000I & 2000M Instrument’s User Guide, Revision B Note: To maintain accuracy whenever outside this temperature range, apply the relevant temperature coefficient adder. Settling Time: ---------- Less than 2 seconds to within 10ppm of change + 0.25µV Drive Capability: ------ 15mA max. (50 output impedance below 220mV, 0.1 above 220mV) Overload: --------------- Output fully protected against short-circuits Long-Term Effects: --- Add 25ppm of output to 1 year specification for 3 year...
Appendix A—Output Specifications Long-Term Effects: --- Add 60ppm of output to 1 year specification for three year accuracy Voltage Measurement Specifications Voltage measurement capabilities are present in the 2000M instrument. To meas- ure voltage, access the instrument’s “Measurement” mode. In Voltage Measurement mode (-0.15 to +0.15V): ...
2000I & 2000M Instrument’s User Guide, Revision B Table 9. Thermocouple Simulation and Measurement Specifications Simulation Measurement Measurement Range Type Resolution Resolution 0.01ºC 0.15ºC +200 to +300ºC 0.01ºC 0.08ºC +300 to +1820ºC 0.01ºC 0.15ºC -270 to -260ºC 0.01ºC 0.05ºC -260 to -70ºC 0.01ºC 0.01ºC -70 to +1000ºC...
Simulation using Passive CJC Simulation using Active CJC Range C Type Stability Relative Accuracy Absolute Accuracy Stability Relative Accuracy Absolute Accuracy 24 hr 90 day 1 yr. 90 day 1 yr. 24 hr 90 day 1 yr. 90 day 1 yr. 200 to 300 0.06 0.11...
Page 90
Measurement using Passive CJC Measurement using Active CJC Range C Type Stability Relative Accuracy Absolute Accuracy Stability Relative Accuracy Absolute Accuracy 24 hr 90 day 1 yr. 90 day 1 yr. 24 hr 90 day 1 yr. 90 day 1 yr. 200 to 300 0.11 0.19...
Page 92
2000I & 2000M Instrument’s User Guide, Revision B...
Sealed Lead Acid–Maintenance Free (2000I and 2000M) Nickel Metal Hydride, STD 3800mAH, Main-tenance Free, Additional batteries available, Contact Vitrek customer support for availability. Battery Charging Signal: Red LED illuminates on the front panel when internal battery during charging. Battery Voltage Level: Green LED illuminates on the front panel when the in- ternal battery has sufficient voltage to power the instru- ment (2000I and 2000M).
2000I & 2000M Instrument’s User Guide, Revision B Note: The internal circuitry is automatically powered down when the battery voltage is low. Note that when this condition is achieved the only way to repower the instrument is to plug it into the external charger. This will reset the internal circuitry and reconnect the internal battery circuitry.
Page 96
2000I & 2000M Instrument’s User Guide, Revision B The common password for enabling External Calibration is: 10288 Place instrument in a known state, such as, current or voltage before entering the password. Note that this page can be removed, as required, to protect this data from unauthorized users.
Need help?
Do you have a question about the Xitron 2000 Series and is the answer not in the manual?
Questions and answers