INDUSTRIAL PRO HYDROGEN SULFIDE DETECTOR PRODUCT MANUAL

WARRANTY:

The Inspector has a one year warranty on all of its parts. It is recommended that you calibrate the Inspector every six
months. If you do not calibrate it then there is a good chance it will not provide an accurate measurement. If a calibration
is never performed an EOL (End of Life) icon will display on the LCD after two years of use. A successful calibration
postpones an EOL for 180 days.

NORMAL OPERATING CONDITIONS:

Temperature: -20oC to +50oC (-4 F to +122 F)
Humidity: 20-90% R.H (0-99% intermittent)

PRACTICAL CONSIDERATIONS:

Short term measurements (i.e. <1hr) in any humidity (0-99% R.H.) are fine as long as condensation does not block the
sensor inlet. A longer term exposure to extreme humidity outside the normal range can cause damage to the sensor.
Damage occurs much faster at higher temperatures. The LCD display and sensor will not work properly at lower
temperatures. NOTE: Higher temperatures can damage the device.
PLEASE NOTE:
Hydrogen Sulfide is a toxic gas that is produced from sour gas, bacteria, waste breakdown processes and other various
waste sources. The H2S Inspector is a professional tool that allows you to measure Hydrogen Sulfide levels.
Thank You for choosing The Sensorcon Hydrogen Sulfide Inspector! We want to be your go to company for all your
professional needs! Please feel free to contact us with any comments, questions or concerns and we'll do our best to
make sure you're 100% satisfied.
Industri^l
Industri^l
Front View
150 Airport Drive Buffalo, NY 14225

CERTIFICATION:

ETL CLASSIFIED
THIS DEVICE CONFORMS TO UL STD 913;
CERTIFIED TO CSA STD C22.2 NO 157
Intertek / ETL control number 4004813
PRO
Hydrogen Sulfide
PRO
Side View
Office: (716) 566-2728
Fax: (716) 276-3047
CONTACT INFORMATION:
150 N. Airport Drive Buffalo, NY 14225
Office: (716) 566-2728 Fax: (716) 276-3047
sales@sensorcon.com
www.sensorcon.com
CUSTOMER SERVICE HOURS OF OPERATION:
Monday through Friday: 8am to 4pm
Please Note: East Coast Time Zone
sales@sensorcon.com
Rear View
www.sensorcon.com
Page 1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Sensorcon Inspector Industrial Pro and is the answer not in the manual?

Questions and answers

Summary of Contents for molex Sensorcon Inspector Industrial Pro

  • Page 1 INDUSTRIAL PRO HYDROGEN SULFIDE DETECTOR PRODUCT MANUAL WARRANTY: The Inspector has a one year warranty on all of its parts. It is recommended that you calibrate the Inspector every six months. If you do not calibrate it then there is a good chance it will not provide an accurate measurement. If a calibration is never performed an EOL (End of Life) icon will display on the LCD after two years of use.
  • Page 2 Industri^l Displays the H2S concentration from 0 to 400ppm.(Other figures below) Maximum H2S level that is measured. Hydrogen Sulfide LCD DISPLAY LAYOUT 15 ppm line is exceeded. 10 CARBON MONOXIDE INSPECTOR QUICKSTART GUIDE 150 Airport Drive Buffalo, NY 14225 Office: (716) 566-2728 Fax: (716) 276-3047 sales@sensorcon.com www.sensorcon.com...
  • Page 3 Industri^l Hydrogen Sulfide Sensor Inlet HOW THE INSPECTOR WORKS: An electrochemical sensor behind the sensor inlet shown in the image above will create an electrical signal when hydrogen sulfide (H2S) is sensed. The gas concentration is displayed in parts per million (PPM). Please Note: The Sensorcon Inspector is a professional tool.

Table of Contents