Page 2
Aiming at the practicality and ease of access to the information contained in the User Manual of our products, DMC® in accordance with IN nº 4/2012 established by ANVISA, provides the documents for download at the following electronic address www.dmcgroup.com.br, in the “USE INSTRUCTIONS” page.
The professional must be qualified to apply the prod- uct techniques. Improper use may lead to irreversible damage. EQUIPMENT FUNCTIONS Therapy Plus emits red and infrared laser light for the following purposes: tissue repair, inflammatory process modulation and analgesia. EQUIPMENT WORKING By means of a small display and three keys, the operator views and performs all settings and features of the equipment.
Page 4
Therapy Plus | User’s Manual Note: DMC provides scientific information about the subject as described in the flyer attached to the product. CONTRAINDICATIONS Laser Therapy: • On pregnant uterus; • Neoplasm in the region to be irradiated; • Clinical lesions without diagnosis;...
SPECIFICATIONS Infrared Laser Features Wavelength 808 nm ± 10 nm Emitter net power 250 mW ± 20% Red Laser (Laser Beam) Features Wavelength 660 nm ± 10 nm Emitter net power 0,5 mW - 2,5 mW Red Laser Features Wavelength 660 nm ±...
Page 6
• Only components mentioned in this manual should be used together with equipment. • This equipment cannot be used with power cable not supplied by DMC because it may provoke an increase of emission or decrease of immunity of equipment.
DISINFECTING”, and insert it into the end of the optical fiber. Handpiece Led When the led of the handpiece is blinking fast, the laser has been activated. LIST OF COMPONENTS Therapy Plus is composed by: PIECES: Handpiece Handpiece support Power cable...
Page 8
Battery ACCESSORIES: 02 Safety glasses 01 Eye protector ILIB bracelet User’s Manual / Warranty term Suitcase for transportation Additional parts or accessories may be purchased from the DMC through the following codes: Code Description 010130131 Power cable 010990317 Power Supply...
Page 9
(G) Optical fiber: it conducts the laser beam. (H) Activation button: it activates or interrupts the laser emission. (I) Battery compartment lid: The image of the withdrawable component of the equipment is shown below with reference to the Manufacturer (DMC) and Therapy Plus the Model (Therapy Plus).
Page 10
Therapy Plus | User’s Manual POSITION OF EQUIPMENT/PATIENT/OPERATOR This equipment should be used in dental, medical or physiotherapy offices and it should be posi- tioned over a bench and connected to power outlet according illustrated below: The spacer inserted in the handpiece must be in contact with the patient and the operator should...
INSTALLATION To install the battery on the handpiece, follow the procedure below: Press the lock pulling the lid outside, to release it from the handpiece as in the figure below: Connect the battery, as in the figure below: Place the battery connector in the compartment as in the figure below: Reinstall the lid of the battery compartment.
Page 12
Therapy Plus | User’s Manual INSPECTION BEFORE USING Before using the equipment always check the integrity of handpiece and power supply cable. They can be neither broken, nor twisted, nor crumpled with protection surface cut or internal conductors exposed. The plastic housing of the handpiece shall not be broken. The optical fiber shall not be damaged.
3. Insert the battery in the compartment and reinstall the lid. The procedure will be illustrated in item “INSTALLATION”. New batteries shall be acquired exclusively from DMC. The Therapy Plus battery contains lithium in its composition and shall not be disposed in common waste. INSULATION OF ELECTRICAL NETWORK In case of emergency or to clean the equipment, power cable plug must be disconnected from outlet.
DMC. DMC doesn’t provide circuit diagrams, component lists, descriptions, instructions for calibration and gauging for unqualified people. If maintenance or other type of technical assistance is performed by unauthorized staff, DMC doesn’t have any responsibility regarding safety of equipment.
Page 15
To decrease these risks, the customer should discard the equipment according to the local legislation. The Therapy Plus battery contains lithium and shall not be disposed in common waste. LASER SUPPLY SYSTEM The equipment is composed by two lasers, one red and one infrared. Laser light is conducted to the end of handpiece by optical fiber.
Page 16
Therapy Plus | User’s Manual CALCULATION OF BEAM DIVERGENCE Laser Emission Applied Standard: 60825-1:2014 / Subclause 3.13 The beam divergence for 600 µm fiber is 0,45 rad ± 0,03 rad. CALCULATION OF MAXIMUM IRRADIANCE Applied Standard: 60825-1:2014 / Subclause 3.43 φ...
Page 17
Manufacturer standards and declaration - Electromagnetic Emissions - for every equipment and system [IEC 60601-1-2 / 2007 – subcl. 6.8.3.201 a) 3)] Therapy Plus is designated to be used in electromagnetic environment specified below. Therapy Plus’ custo- mer or user shall ensure it will be used in such environment.
Page 18
Manufacturer standards and declaration - Electromagnetic Immunity - for every Therapy Plus [IEC 60601-1-2 / 2007 – subcl. 6.8.3.201 a) 6)] Therapy Plus is designated to be used in electromagnetic environment specified below. Therapy Plus’ custo- mer or user shall ensure it will be used in such environment.
Page 19
Manufacturer standards and declaration - Electromagnetic Immunity - Equipment which doesn’t give SUPPORT FOR LIFE [IEC 60601-1-2 / 2007 – subcl. 6.8 Therapy Plus is designated to be used in electromagnetic environment specified below. Therapy Plus’ customer or user shall ensure it will be used in such environment.
Minimal distances recommended for portable and mobile RF communication equipment and Therapy Plus. Therapy Plus is designated to be used in electromagnetic environment in which radiated RF disturbances are controlled. Therapy Plus’ customer or user may help avoiding electromagnetic interference keeping a minimal distance between por- table or mobile RF communication equipment (transmitters) and Therapy Plus, as recommended below, according to the maximum output power of communication equipment.
Page 21
Applied part type B. It identifies the positive and the negative connections (polarity) of a direct cur- rent power supply or the positive and the negative connections of a handpiece of the equipment, which will be connected to the direct current power supply. Continuous current Protected against 12.5-mm-size solid unknown objects and not protected I P 20...
E. The most common defects are physical shocks applied to the equipment, and in these cases the warranty is cancelled. F. DMC is not responsible for personal or material damages originating from undue use of equip- ment, and user should provide safety measurements in order to avoid such occurrences.
Page 23
DMC IMPORTAÇÃO E EXPORTAÇÃO DE EQUIPAMENTOS LTDA. Rua Sebastião de Moraes, 831 | Jardim Alvorada | São Carlos/SP CEP 13562-030 | CNPJ 02.827.605/0001-86 ANVISA Registration: 80030810157 Tech. Responsible: Renaldo Massini Jr. CREA 0601706815 Technical Name: Laser therapy system...
Page 24
For more information about our technology and how to use it, visit: www.nupen.com.br Technical Assistance DMC IMP. E EXP. DE EQUIP. LTDA. +55 (16) 2107-2323 www.dmcgroup.com.br...